Using Prenatal Blood Samples to Evaluate COVID-19 Rapid Serologic Tests Specificity.

Jackeline Alger1,2,3,4, Maria Luisa Cafferata5,6, Tito Alvarado1,3, Alvaro Ciganda6, Arturo Corrales1,7, Hans Desale8, Arnaud Drouin9, Dahlene Fusco9, Jorge Garcia2,3, Luz Gibbons5, Emily Harville10, Wendy Lopez2,3, Ivette Lorenzana1,11, Fausto Muñoz-Lara1,12,13, Elsa Palou1,13, Eduardo Retes1,14, Manuel Sierra1,3,15,16, Candela Stella5, Xu Xiong10, Lysien I Zambrano1,17, Pierre Buekens18,19.   

Abstract

INTRODUCTION: Background cross-reactivity with other coronaviruses may reduce the specificity of COVID-19 rapid serologic tests. The vast majority of women attend prenatal care, which is a unique source of population-based blood samples appropriate for validation studies. We used stored 2018 serum samples from an existing pregnancy cohort study to evaluate the specificity of COVID-19 serologic rapid diagnostic tests.
METHODS: We randomly selected 120 stored serum samples from pregnant women enrolled in a cohort in 2018 in Tegucigalpa, Honduras, at least 1 year before the COVID-19 pandemic. We used stored serum to evaluate four lateral flow rapid diagnostic tests, following manufacturers' instructions. Pictures were taken for all tests and read by two blinded trained evaluators.
RESULTS: We evaluated 120, 80, 90, and 90 samples, respectively. Specificity for both IgM and IgG was 100% for the first two tests (95% confidence intervals [CI] 97.0-100 and 95.5-100, respectively). The third test had a specificity of 98.9% (95% CI 94.0-100) for IgM and 94.4% (95% CI 87.5-98.2) for IgG. The fourth test had a specificity of 88.9% (95% CI 80.5-94.5) for IgM and 100% (95% CI 96.0-100) for IgG. DISCUSSION: COVID-19 serologic rapid tests are of variable specificity. Blood specimens from sentinel prenatal clinics provide an opportunity to validate serologic tests with population-based samples.

Entities:  

Keywords:  COVID-19 diagnostic testing; Pregnancy; Prenatal care; Serologic tests

Mesh:

Year:  2020        PMID: 32671537      PMCID: PMC7363497          DOI: 10.1007/s10995-020-02981-9

Source DB:  PubMed          Journal:  Matern Child Health J        ISSN: 1092-7875


Significance

What is already known on this subject? The validity of COVID-19 serologic rapid tests is not established, although they are becoming widely available. Sentinel prenatal clinics provide a unique opportunity to collect population-based blood samples for validation studies. What this study adds? We used stored serum samples collected in a sentinel prenatal clinic before the COVID-19 pandemic in Tegucigalpa, Honduras. We evaluated four rapid serologic tests and found specificities varying from 88.9 to 100%. The high number of false positives for some tests would make them less suitable for field studies.

Introduction

There is an urgent need to obtain serologic data for SARS-CoV-2, the virus causing COVID-19, for populations worldwide. Rapid diagnostic tests (RDTs) could play a key role in serologic surveys, but their validity is often not well documented for COVID-19 tests (Sethuraman et al. 2020). Blood routinely collected during prenatal care is a unique source of population-based samples which could be used to perform COVID-19 serologic studies (Buekens et al. 2020). UNICEF estimates that 87.4% of pregnant women in the world had at least one prenatal visit in 2014–2019 (UNICEF 2019). Data from sentinel prenatal clinics are used to generate population-based estimates of infectious diseases seroprevalence. For example, HIV/AIDS seroprevalence is often estimated from prenatal care data (Eaton et al. 2014). Additionally, prenatal care offers a unique opportunity to collect data from a population of both asymptomatic and symptomatic women. Most women receive prenatal care to some extent, providing a better estimate of population seroprevalence than samples based on blood banks, hospital biobanks, or volunteers. The specificity of COVID-19 serologic tests has been questioned because of possible background cross-reactivity with other coronaviruses. This has been an issue for serological testing of SARS-CoV and MERS (Meyer et al. 2014). Specificity corresponds to the proportion of true negative among healthy subjects. We used stored serum samples collected before the pandemic from the Zika in Pregnancy in Honduras (ZIPH) cohort study to evaluate the specificity of COVID-19 serologic RDTs available for field studies in Honduras. The first COVID-19 case in Honduras was reported on March 10, 2020 (Gobierno de Honduras 2020). Nine weeks later, the country has reported 2646 COVID-19 cases and a low testing rate of 572 tests per million inhabitants (Worldometer 2020). Soon after the start of the COVID-19 pandemic in Honduras, the government independent group named Plataforma Todos Contra el COVID-19 (all against COVID-19 platform) was formed. One of the working aims of this group has been to guide and scale up the laboratory testing for confirmation of individual cases and for epidemiologic surveillance studies (Plataforma Todos Contra el COVID-19 2020).

Methods

The current study leveraged stored serum samples from an existing prospective pregnancy cohort. The ZIPH study enrolls women at their first prenatal visit at the Alonso Suazo Health Center (Tegucigalpa, Honduras) and follows them up until delivery (Buekens et al. 2016). The health center is a large urban prenatal clinic with more than 1,100 new prenatal visits a year. We enrolled 3991 women from July 2016 to March 2020. Data management is centralized at the Institute for Clinical Effectiveness and Health Policy (Instituto de Efectividad Clínica y Sanitaria, IECS), Buenos Aires, Argentina. The Data Center selected at random 120 women among 977 women who were enrolled in 2018 and had authorized their blood samples to be stored for 10 years for additional studies. All samples were collected more than 1 year before the first case of COVID-19 was reported in Honduras. The sample size was based on the number of RDTs available for evaluation. The Data Center prepared a list of study numbers (with control digits), which was emailed to Honduras where study labels were printed and pasted on data forms and on rapid tests immediately before use (Fig. 1). Data were collected on paper forms and entered in REDCap (Harris et al. 2009). Additionally, scans of each data form were taken and were sent (encrypted) to IECS. This system allows for a digital backup of all study data forms.
Fig. 1

a–c Digital picture of COVID-19 rapid serological test (RDT#1) with negative result (a) and RDT#3 with IgG (b) and RDT#4 with IgM (c) positive results

a–c Digital picture of COVID-19 rapid serological test (RDT#1) with negative result (a) and RDT#3 with IgG (b) and RDT#4 with IgM (c) positive results Four lateral flow RDT were obtained by the Plataforma Todos Contra el COVID-19 through the Honduran Sanitary Regulatory Agency (Agencia Regulatoria Sanitaria, ARSA,) or directly as a donation. We used the following lateral flow RDTs: RDT#1: Hightop COVID-19 IgM/IgG Ab Rapid Test Kit (Qingdao Hantang Biological Technology Co., Ltd., Qingdao, China); RDT#2: COVID-19 IgG/IgM Rapid Test Kit (Nantong Egens Biotechnology Co. Ltd., Nantong, China); RDT#3: Orient Gene COVID-19 IgG/IgM Rapid Test (Zhejiang Orient Gene Biotech Co. Ltd, Huzhou, China). RDT#4: Standard Q COVID-19 IgM/IgG Duo Test (SD Biosensor, Gyeonggi-do, Republic of Korea). Laboratory procedures were performed at the Departamento de Laboratorio Clínico, Hospital Escuela, Tegucigalpa, Honduras. One aliquot of frozen serum was thawed and kept at 4 ℃ for up to 3 days. Samples were brought to room temperature before testing. RDT packages were opened immediately before use. RDTs were performed following the manufacturers’ instructions. Rapid tests were read at 10 to 20 min according to the respective RDT instructions, and digital pictures were immediately taken using a cell phone camera. Pictures were taken under constant lighting conditions using two light sources. We used a 3 × magnification to allow the picture to be taken from enough distance to avoid shadows. The pictures were uploaded to an encrypted cloud site and were immediately available to the Data Center. The pictures were then uploaded to a secure website and were read by two blinded trained evaluators and compared to the initial reading in the laboratory. Discrepancies between the initial reading and readings from the blinded evaluators, or discrepancies between blinded evaluators, were resolved by an independent third blinded trained evaluator. This is similar to the approach we used earlier for Chagas RDTs (Buekens et al. 2013, 2018). We used OpenEpi version 3.01 to calculate 95% Fisher exact confidence intervals (CI). This study was approved by Tulane University Institutional Review Board (893652; Amendment approval April 22, 2020), and by the Facultad de Ciencias Médicas, Universidad Nacional Autónoma de Honduras ethics committee (CEIB-079-2016; Amendment approval April 24, 2020). All women provided written informed consent at enrollment and authorized their blood samples to be stored for up to 10 years.

Results

Table 1 shows the characteristics of the selected women enrolled in the ZIPH cohort.
Table 1

Maternal characteristics

N = 120
n/N%
Maternal age
 Between 15 and 1931/12025.8
 Between 20 and 3480/12066.7
 35 or higher9/1207.5
Years of Education*9(6–12)
Gestational age at first visit (weeks)
 Between 1 and 1369/9374.2
 Between 14 and 2719/9320.4
 28 or higher5/935.4
Symptoms during pregnancy
 Eruption0/1160.0
 Fever0/1160.0
 Arthralgia0/1160.0
 Arthritis0/1160.0
 Conjunctivitis0/1160.0
 At least one symptom0/1160.0

*Median (interquartile range IQ)

Maternal characteristics *Median (interquartile range IQ) We tested 120 RDTs #1, 80 RDTs #2, and 90 RDTs #3 and #4 each. All tests had a clearly visible control band for all samples, except one for RDT#4. Pictures were taken for all tests (Fig. 1). Both blinded evaluators agreed with the initial reading in 100% (120/120) of RDT#1 s, 96.3% (77/80) of RDTs#2, 95.6% (86/90) of RDTs#3, and 93.3% (84/90) of RDT#4. The third independent blinded evaluator made the final decision in 13 cases. Table 2 shows that the specificity for both IgM and IgG was 100% for both RDT #1 and RDT #2, with 95% CIs lower limits of 97.0% and 95.5%, respectively. Specificity for RDT #3 was 98.9% for IgM and 94.4% for IgG. Specificity for RDT#4 was 88.9% for IgM and 100% for IgG with a 95% CI lower limit of 96.0%.
Table 2

Serologic rapid diagnostic tests (RDTs): positive (+) tests for IgM and IgG and specificity (% and 95% confidence intervals [CI])

IgM+ IgM specificity% (95% CI)IgG+ IgG specificity% (95% CI)
RDT #10/120100 (97.0–100)0/120100 (97.0–100)
RDT #20/80100 (95.5–100)0/80100 (95.5–100)
RDT #31/9098.9 (94.0–100)5/9094.4 (87.5–98.2)
RDT #410/9088.9 (80.5–94.5)0/89100 (96.0–100)
Serologic rapid diagnostic tests (RDTs): positive (+) tests for IgM and IgG and specificity (% and 95% confidence intervals [CI])

Discussion

Our results suggest that the COVID-19 serologic RDTs available in Honduras are of variable specificity when used in a general population. Specificity should have been 100% for all tests in these pre-COVID-19 samples. RDT #1 and #2 was 100% specific for both IgM and IgG, but specificity was as low as 88.9% for IgM for RDT#4 and 94.4% for IgG for RDT #3. None of the RDTs we tested are approved by the United States Food and Drug Administration (FDA), but all of them have a CE Mark for use in the European Union (Weissleder et al. 2020). The specificities we found are comparable to the ones reported in the recent literature. A study of 10 serologic rapid tests found specificities varying from 91.6 to 100% among 108 blood donor plasma specimens collected in the United States before July 2018 (COVID-19 Testing Project 2020). A review of mostly unpublished data about nine serologic RDTs found specificities of 98.7% to 100%, including RDT #1, for which reported specificities were of 96.0% for IgM and 97.5% for IgG (Zainol Rashid et al. 2020). Hoffman et al. (2020) evaluated RDT #3 and found higher specificities, 100% for IgM and 99.2% for IgG. A meta-analysis of regulatory data from serologic RDTs approved in Brazil found a pooled specificity of 97% for IgM and 98% for IgG (Castro et al. 2020). A study evaluating a RDT in China included 128 “clinical negative samples” and found one sample positive for IgG and 10 positive for IgM (Li et al. 2020). Of note, “clinical negative samples” in Li. et al. (2020) were from unconfirmed suspected COVID-19 cases, which could explain the high frequency of positive IgM. Another study of a COVID-19 serologic RDT found no positive among 26 healthy blood donors, with a specificity of 100% (Shen et al. 2020). Data from SARS-CoV and MERS suggested that background cross-reactivity with other human coronaviruses was an issue (Meyer et al. 2014). However, this was mostly the case for tests detecting antibodies against the whole virus rather than the ones detecting antibodies against specific recombinant antigens. More than 90% of adults have been reported to have antibodies against the four human coronaviruses (Gorse et al. 2010). Coronavirus infection among children with influenza-like illness occurs in Central America as in other countries of the Americas (Taylor et al. 2017). An unexpected finding from a study from Vietnam showed that human coronavirus infection was less frequent among children with influenza-like illness when housing was overcrowded (Nguyen et al. 2016). The 2012 Honduras Demographic and Health Survey (DHS) shows that in Tegucigalpa, 38% of the households have > 4 people (Secretaría de Salud de Honduras, Instituto Nacional de Estadística e ICF International 2013). It is thus possible that circulation of other human coronaviruses is relatively low in our setting. Rapid tests used for seroprevalence surveys need to be highly specific. The low specificity for IgG of RDT #3 is a concern, as it would falsely classify more than 5% of those tested as positive. The other three RDTs were highly specific for IgG and would be recommended if further studies would show a high sensitivity. We conclude that among those tested, COVID-19 serologic RDTs specificity varies from a low 88.9 to 100% in unaffected populations from Tegucigalpa, Honduras. If further studies confirm their sensitivity in that setting, highly specific tests would be the instrument of choice for serological surveys. The validity of all RDTs should be carefully evaluated, and blood specimens from sentinel prenatal clinics provide an opportunity to test them with population-based samples.
  17 in total

1.  Clinical evaluation of a rapid colloidal gold immunochromatography assay for SARS-Cov-2 IgM/IgG.

Authors:  Bo Shen; Yufen Zheng; Xiaoyan Zhang; Weituo Zhang; Donglian Wang; Jie Jin; Rong Lin; Ying Zhang; Guangjun Zhu; Hongguo Zhu; Jun Li; Jiaqin Xu; Xianhong Ding; Shiyong Chen; Ruyue Lu; Zebao He; Haihong Zhao; Lingjun Ying; Chao Zhang; Dongqing Lv; Baofu Chen; Jiya Chen; Jiansheng Zhu; Bingjie Hu; Chenliang Hong; Xiangyu Xu; Jiaxi Chen; Chong Liu; Kai Zhou; Jing Li; Guoling Zhao; Weixiang Shen; Chunfeng Chen; Chunyan Shao; Xiaoying Shen; Jingjing Song; Zhipeng Wang; Ying Meng; Chao Wang; Junsong Han; Aojun Chen; Daru Lu; Biyun Qian; Haixiao Chen; Hengjun Gao
Journal:  Am J Transl Res       Date:  2020-04-15       Impact factor: 4.060

2.  Research electronic data capture (REDCap)--a metadata-driven methodology and workflow process for providing translational research informatics support.

Authors:  Paul A Harris; Robert Taylor; Robert Thielke; Jonathon Payne; Nathaniel Gonzalez; Jose G Conde
Journal:  J Biomed Inform       Date:  2008-09-30       Impact factor: 6.317

3.  Interpreting Diagnostic Tests for SARS-CoV-2.

Authors:  Nandini Sethuraman; Sundararaj Stanleyraj Jeremiah; Akihide Ryo
Journal:  JAMA       Date:  2020-06-09       Impact factor: 56.272

Review 4.  Diagnostic performance of COVID-19 serology assays.

Authors:  Z Zainol Rashid; S N Othman; M N Abdul Samat; U K Ali; K K Wong
Journal:  Malays J Pathol       Date:  2020-04       Impact factor: 0.656

5.  Prevalence of antibodies to four human coronaviruses is lower in nasal secretions than in serum.

Authors:  Geoffrey J Gorse; Gira B Patel; Joseph N Vitale; Theresa Z O'Connor
Journal:  Clin Vaccine Immunol       Date:  2010-10-13

6.  Congenital Transmission of Trypanosoma cruzi in Argentina, Honduras, and Mexico: An Observational Prospective Study.

Authors:  Pierre Buekens; María Luisa Cafferata; Jackeline Alger; Fernando Althabe; José M Belizán; Norma Bustamante; Yves Carlier; Alvaro Ciganda; Jaime H Del Cid; Eric Dumonteil; Rubí Gamboa-León; Jorge A García; Luz Gibbons; Olga Graiff; Jesús Gurubel Maldonado; Claudia Herrera; Elizabeth Howard; Laura Susana Lara; Benjamín López; María Luisa Matute; María Jesús Ramírez-Sierra; María Cecilia Robles; Sergio Sosa-Estani; Carine Truyens; Christian Valladares; Dawn M Wesson; Concepción Zúniga
Journal:  Am J Trop Med Hyg       Date:  2017-11-30       Impact factor: 2.345

7.  Congenital transmission of Trypanosoma cruzi in Argentina, Honduras, and Mexico: study protocol.

Authors:  Pierre Buekens; Maria-Luisa Cafferata; Jackeline Alger; Fernando Althabe; José M Belizán; Yves Carlier; Alvaro Ciganda; Eric Dumonteil; Rubi Gamboa-Leon; Elizabeth Howard; Maria Luisa Matute; Sergio Sosa-Estani; Carine Truyens; Dawn Wesson; Concepcion Zuniga
Journal:  Reprod Health       Date:  2013-10-11       Impact factor: 3.223

8.  Development and clinical application of a rapid IgM-IgG combined antibody test for SARS-CoV-2 infection diagnosis.

Authors:  Zhengtu Li; Yongxiang Yi; Xiaomei Luo; Nian Xiong; Yang Liu; Shaoqiang Li; Ruilin Sun; Yanqun Wang; Bicheng Hu; Wei Chen; Yongchen Zhang; Jing Wang; Baofu Huang; Ye Lin; Jiasheng Yang; Wensheng Cai; Xuefeng Wang; Jing Cheng; Zhiqiang Chen; Kangjun Sun; Weimin Pan; Zhifei Zhan; Liyan Chen; Feng Ye
Journal:  J Med Virol       Date:  2020-04-13       Impact factor: 2.327

9.  Respiratory viruses and influenza-like illness: Epidemiology and outcomes in children aged 6 months to 10 years in a multi-country population sample.

Authors:  Sylvia Taylor; Pio Lopez; Lily Weckx; Charissa Borja-Tabora; Rolando Ulloa-Gutierrez; Eduardo Lazcano-Ponce; Angkool Kerdpanich; Miguel Angel Rodriguez Weber; Abiel Mascareñas de Los Santos; Juan-Carlos Tinoco; Marco Aurelio P Safadi; Fong Seng Lim; Marcela Hernandez-de Mezerville; Idis Faingezicht; Aurelio Cruz-Valdez; Yang Feng; Ping Li; Serge Durviaux; Gerco Haars; Sumita Roy-Ghanta; David W Vaughn; Terry Nolan
Journal:  J Infect       Date:  2016-09-22       Impact factor: 6.072

10.  COVID-19: a meta-analysis of diagnostic test accuracy of commercial assays registered in Brazil.

Authors:  Rodolfo Castro; Paula M Luz; Mayumi D Wakimoto; Valdilea G Veloso; Beatriz Grinsztejn; Hugo Perazzo
Journal:  Braz J Infect Dis       Date:  2020-04-18       Impact factor: 3.257

View more
  2 in total

1.  SARS-CoV-2 has been circulating in northeastern Brazil since February 2020: evidence for antibody detection in asymptomatic patients.

Authors:  Ricardo Queiroz Gurgel; Laís Catarine de Sá; Daniela Raguer Valadão Souza; Aline Fagundes Martins; Igor Leonardo Santos Matos; Alexandra Giovanna Aragão Lima; Sarah Cristina Fontes Vieira; José Melquiades de Rezende Neto; Luis Eduardo Cuevas; Lysandro Pinto Borges
Journal:  J Infect       Date:  2020-12-01       Impact factor: 6.072

Review 2.  Implementing COVID-19 (SARS-CoV-2) Rapid Diagnostic Tests in Sub-Saharan Africa: A Review.

Authors:  Jan Jacobs; Vera Kühne; Octavie Lunguya; Dissou Affolabi; Liselotte Hardy; Olivier Vandenberg
Journal:  Front Med (Lausanne)       Date:  2020-10-30
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.