Literature DB >> 32666630

Sequential His bundle and left ventricular pacing for cardiac resynchronization.

Amrish Deshmukh1, Sudhakar Sattur2, Tim Bechtol3, Luuk I B Heckman4, Frits W Prinzen4, Pramod Deshmukh2.   

Abstract

INTRODUCTION: Fusion of left ventricular pacing with intrinsic conduction provides superior resynchronization compared to biventricular pacing. His bundle pacing (HBP) preserves intrinsic conduction and allows for constant fusion with left ventricular pacing. This study evaluated sequential His bundle and left ventricular pacing for cardiac resynchronization therapy (CRT).
METHODS: In patients referred for CRT, sequential His bundle and left ventricular pacing was performed when HBP did not correct the QRS. At implant, QRS duration and area were compared between biventricular pacing and His bundle and left ventricular pacing. Devices were programmed for His and left ventricular pacing. Functional status and echocardiography were evaluated in follow up.
RESULTS: Twenty-one patients, seven female, 70.7 ± 9.9 years, 57% with nonischemic cardiomyopathy were included. Baseline QRS duration was 170 ± 21 ms and was 157 ± 16 ms with HBP. Biventricular pacing resulted in a QRS duration of 141 ± 15 ms and decreased to 110 ± 14 ms with His bundle and left ventricular pacing (p < .0005). His bundle and left ventricular pacing resulted in a smaller paced QRS area (38.5 ± 22.6 µVs) compared to biventricular pacing (67.5 ± 24.0 µVs) and baseline (78.1 ± 28.1 µVs; p < .0005). Left ventricular ejection fraction increased from 27.6 ± 6.4% to 41.1 ± 12.5 (at 25 mean months, p = .001) and functional class improved from 3.1 ± 0.5 to 2.1 ± 0.8 (at mean 32 months, p < .001).
CONCLUSIONS: Sequential His bundle and left ventricular pacing results in superior electrical synchrony in patients with indication for CRT when HBP does not correct the QRS and resulted in promising clinical and echocardiographic response rates.
© 2020 Wiley Periodicals LLC.

Entities:  

Keywords:  His bundle pacing; QRS area; cardiac resynchronization

Mesh:

Year:  2020        PMID: 32666630     DOI: 10.1111/jce.14674

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  6 in total

Review 1.  Evaluation of electrocardiogram and echocardiographic characteristics of pre-and post-operation of His bundle pacing: A comprehensive review and meta-analysis.

Authors:  Mingzhu Li; Fei Ren; Jing Tian; Kai Yang; Jie Zhang; Hejian Song; Delu Yin; Steven Cui
Journal:  Anatol J Cardiol       Date:  2021-12       Impact factor: 1.596

Review 2.  Physiologic Pacing Targeting the His Bundle and Left Bundle Branch: a Review of the Literature.

Authors:  Seth D Scheetz; Gaurav A Upadhyay
Journal:  Curr Cardiol Rep       Date:  2022-06-09       Impact factor: 3.955

Review 3.  Conduction System Pacing for Cardiac Resynchronisation.

Authors:  Parikshit S Sharma; Pugazhendhi Vijayaraman
Journal:  Arrhythm Electrophysiol Rev       Date:  2021-04

4.  Amulet™ Shines and Protects; Pacing Battle Intensifies with "More Leads or No Leads"?

Authors:  Christopher R Ellis; Nicholas King
Journal:  J Innov Card Rhythm Manag       Date:  2022-01-15

Review 5.  Alternative pacing strategies for optimal cardiac resynchronization therapy.

Authors:  Juan Hua; Qiling Kong; Qi Chen
Journal:  Front Cardiovasc Med       Date:  2022-09-27

Review 6.  The electrocardiogram characteristics and pacing parameters of permanent left bundle branch pacing: a systematic review and meta-analysis.

Authors:  Jia Gao; Bing-Hang Zhang; Nan Zhang; Meng Sun; Rui Wang
Journal:  J Interv Card Electrophysiol       Date:  2021-06-26       Impact factor: 1.900

  6 in total