Effect of T-Piece Resuscitator versus Self Inflating Bag on Arterial CO2 Levels in Preterm Neonates.

Amrit Jeevan1, Haribalakrishna Balasubramanian2, Nandkishor Kabra1.   

Abstract

OBJECTIVE: To compare the effect of self-inflating bag (SIB) vs. T-piece resuscitator (TPR) on PaCO2 levels, when used for brief manual ventilation during administration of first dose of surfactant.
METHODS: Preterm neonates were randomized to receive positive pressure ventilation with either self-inflating bag or T-piece resuscitator during administration of first dose of surfactant. Arterial blood gases were obtained at baseline and 1 h after the intervention. Primary outcome was the mean change in PaCO2 levels 1 h after the intervention.
RESULTS: Eighty neonates were enrolled (40 in each group). The drop in PaCO2 at 1 h was significantly greater in the self inflating bag group as compared to the T-piece resuscitator group [8.96 ± 9.06 mmHg vs. 1.37 ± 9.06 mmHg, Mean difference = 7.58 mmHg, (95% CI: 3.78 to 11.4); P < 0.01]. The PaCO2 change was also statistically significant in the subgroup of infants that required only non-invasive ventilation. The need for second dose of surfactant was higher in the self-inflating bag group [77% vs. 55%, RR - 1.41 (95% CI: 1.02 to 1.94); P = 0.03].
CONCLUSIONS: T-piece resuscitator results in smaller reduction in PaCO2 levels compared to the self inflating bag, when used for brief manual ventilation during surfactant administration.

Entities:  

Keywords:  Arterial blood gas; Preterm; Self inflating bag; Surfactant; T- piece resuscitator

Mesh:

Substances:

Year:  2020        PMID: 32666238      PMCID: PMC7358992          DOI: 10.1007/s12098-020-03393-x

Source DB:  PubMed          Journal:  Indian J Pediatr        ISSN: 0019-5456            Impact factor:   1.967


Introduction

Self-inflating bag (SIB) and T-piece resuscitator (TPR) are the two most common manual ventilation devices (MVDs) used in neonates in the delivery room and intensive care unit. Various studies on mechanical models have shown that TPR, as compared to SIB, generates inflation pressures closer to the targeted values and minimizes breath-to-breath variability in the inspiratory pressures [1-4]. Sustained lung inflations could be more easily achieved with TPR than SIB [5, 6]. TPR also delivers lower and less variable tidal volumes than SIB [7]. Currently, there is limited data on the pressures and volumes achieved in neonatal patients. Dawson et al. observed a lower tidal volume provided by TPR in preterm neonates less than 29 wk of gestation [8]. Whether this effect translates into reduced risk of hypocarbia in the early neonatal period has not yet been evaluated. Studies have also raised concerns regarding generation of inadvertent positive end expiratory pressures (PEEP) with TPR [9-11]. The effect of these two MVDs on PaCO2 levels still remains unclear. Hypocarbia is defined by a PaCO2 level below 35 mmHg [12]. Two or more PaCO2 values less than 25 mmHg in first 48 h of life constitutes severe early hypocarbia [13]. Major fluctuations in PaCO2 can significantly affect systemic blood pressure and cerebral blood flow. A retrospective analysis of 314 preterm neonates <29 wk of gestation revealed that a single PaCO2 value <30 mmHg in first 48 h of life was associated with increased incidence of severe intraventricular hemorrhage and periventricular leukomalacia. Prolonged hypocarbia during this period also increased the odds of bronchopulmonary dysplasia [14]. Given the long term implications of hypocarbia, it is essential to assess the impact of these two MVDs on carbon dioxide elimination in preterm neonates. There is also a paucity of comparative studies on MVDs outside the context of delivery room resuscitation. In centres where minimally invasive surfactant therapy is not practiced, MVDs are regularly being used during surfactant administration while on non-invasive respiratory support. Hence, the authors planned to conduct a randomized controlled trial (RCT) comparing the effect of SIB vs. TPR on PaCO2 levels, when used for brief manual ventilation during administration of first dose of surfactant. They hypothesized that manual ventilation with TPR, as compared to SIB, results in smaller reduction in PaCO2 levels.

Material and Methods

This single centre, parallel group RCT was conducted in the Neonatal Intensive Care Unit of Surya Children’s Hospital, Mumbai, Western India from August 2015 through February 2018. Preterm neonates having respiratory distress and/or FiO2 requirements of >35%, despite being on positive end expiratory pressure (PEEP) of 6 cm H2O within first 24 h of life, qualified for the first dose of surfactant and were considered eligible for enrolment into the study. Neonates with major congenital anomalies, complex congenital heart disease, pneumothoraces and those who had received first dose of surfactant prior to transfer to the hospital were excluded. The trial was approved by the institutional ethics committee. The trial was prospectively registered with Clinical Trials Registry of India (CTRI/2015/08/006067) following which the first subject was enrolled. Preterm neonates requiring respiratory support were transported to authors’ unit on Intermittent Mandatory Ventilation or nasal Continuous Positive Airway Pressure through a transport ventilator (T1500 Globe Trotter, Drager, Lubeck, Germany). Neonates with severe respiratory distress and/or FiO2 requirement of >50% were transported on Intermittent Mandatory Ventilation. Those infants requiring ≤40% FiO2 on admission were directly extubated to Continuous Positive Airway Pressure following administration of the first dose of surfactant. Neonates retrieved on Continuous Positive Airway Pressure and requiring >35% FiO2 were considered for surfactant treatment. They were continued on non-invasive ventilation unless they met previously specified ventilation criteria. Second dose of surfactant was given if the FiO2 requirement was ≥35% at 6 h after the first dose. Random number sequence was generated in variable block sizes of two or four each, using Stata software. The random sequence was generated by a statistician who was not a part of the study. The random number codes were placed in sequentially numbered sealed opaque envelopes. Neonates were enrolled by a senior resident after obtaining written informed consent from the parents/guardians. The resident then opened the sealed envelope to determine the infant’s MVD assignment. Masking the nurse and senior resident to allocated MVD was not possible due to the nature of the intervention. However, the study investigators and the laboratory technicians processing the blood gas were unaware of the allocation. Treating clinicians had to be informed the blood gas report since it was crucial for further management. The intervention was brief manual ventilation with either SIB or TPR during administration of first dose of surfactant. Baseline preductal arterial blood gas was obtained prior to the intervention. The staff nurse involved in the care of the neonate, endotracheally administered the first dose of surfactant while the resident provided positive pressure ventilation using the appropriate MVD. All senior residents had formal training on the use of MVDs. Bovine surfactant (25 mg phospholipids/ml) at the dose of 4 ml/kg was instilled in four aliquots through a 5 or 6 French infant feeding tube. After each aliquot, neonates were ventilated with SIB/TPR for either 30 s or the time to achieve target saturations of 92–95%, whichever was later. Infant’s vital signs (heart rate, respiratory rate) and saturations were continuously monitored during the procedure. Preductal arterial blood gas was obtained at 1 h and 6 h post intervention. Second dose of surfactant, if indicated, was administered after the 6 h ABG. Infant silicone manual resuscitator (Zeal Medical, Mumbai, India) without a positive end expiratory valve was used in SIB group and Neopuff (Fisher & Paykel Healthcare, Auckland, New Zealand) was used in the TPR group. The SIB had a capacity of 250 ml with the pop off valve set at 35 cm H2O. In the TPR group, initial peak inspiratory pressure and PEEP was set at 20 cm H2O and 5 cm H2O respectively. Subsequently, peak pressure was adjusted to achieve optimal chest rise. Maximal pressure release valve was set at 40 cm H2O. Manual ventilation was provided at the rate of 40 breaths per min in both the groups. Initial FiO2 of 30% was titrated to a target saturation of 92–95%. Neonates requiring ventilation were commenced on volume guarantee mode with a tidal volume of 4–6 ml/kg, inspiratory time of 0.3 s, PEEP of 5 cm H2O and FiO2 to target SpO2 of 92–95%. Extubation was considered if peak inspiratory pressure required to generate the set tidal volume was ≤18 cm H2O, FiO2 requirement ≤30% and blood gas was suggestive of a pH ≥ 7.25 with PaCO2 ≤ 55 mmHg. The primary outcome was the difference between baseline and 1 h PaCO2. This was quantified as mean change 1. Secondary outcomes were: (a) The difference between baseline PaCO2 and 6 h PaCO2 - mean change 2; (b) Blood gas variables (pH, PaCO2 and PaO2) at 1 and 6 h post intervention; (c) Clinical outcomes such as mortality prior to discharge, requirement of mechanical ventilation, need for second dose of surfactant, duration of invasive and non-invasive ventilation, duration of oxygen therapy and length of hospital stay; (d) Neonatal morbidities - severe intraventricular hemorrhage, patent ductus arteriosus requiring medical treatment, necrotizing enterocolitis, late onset sepsis, retinopathy of prematurity requiring treatment, and bronchopulmonary dysplasia. Observations from authors’ neonatal unit indicated that the average reduction in PaCO2 following a brief period of manual ventilation with SIB during surfactant administration was 12 mmHg (standard deviation of 8.5 mmHg). In order to detect a 6 mmHg difference in the change in PaCO2 between the two groups with an assumed standard deviation of 8 mmHg, a sample size of 39 neonates in each group was estimated for a study power of 90% and two tailed alpha of 0.05. Statistical analysis was conducted using the Stata 13.1 software (Statacorp LP, 4905, Lakeway Drive, College Station, Texas, USA). Descriptive statistics were used to define the baseline and outcome variables in both groups. Categorical variables were compared using the Fisher’s exact test. Continuous measures were compared using two sample t-test or Mann-Whitney U-test as appropriate. Analysis was performed by applying intention to treat principle. A p value of <0.05 was considered statistically significant. Factorial analysis of variance was performed to check the dependence of the mean change in PaCO2 on the two factors- type of MVD and duration of respiratory support (hours of life) and interaction between them, if any. Line charts and box plots were constructed to provide a visual impression of the difference in the pre and post intervention PaCO2.

Results

A total of 176 eligible neonates were admitted to authors’ unit from August 2015 through February 2018. Of these, 80 were enrolled in the study (40 in the TPR group and 40 in the SIB group) (Fig. 1). All the neonates were analyzed for the primary outcome. There were no adverse events associated with surfactant administration or manual ventilation.
Fig. 1

CONSORT 2010 flow diagram

CONSORT 2010 flow diagram Majority of the preterm neonates were outborn (55%) and cesarean delivered (85%). The gestational age of the study patients ranged from 26 to 36 wk and 88% of them were > 28 wk. Most of them were vigorous (83%) at birth and were managed with non-invasive ventilatory support (80%). Neonates who were primarily commenced on non-invasive respiratory support did not require mechanical ventilation subsequently. Sixteen (20%) were on mechanical ventilation at the time of primary outcome assessment. Eleven (14%) continued to require ventilation at 6 h after intervention. Baseline maternal and neonatal characteristics were similar in the study groups (Table 1).
Table 1

Patients’ demographic and baseline characteristics

CharacteristicsSIB(n = 40)TPR(n = 40)P
Mothers
    Cesarean section35 (87%)33 (82%)0.75
    Pregnancy induced hypertension10 (25%)9 (22%)>0.99
    Gestational diabetes mellitus6 (15%)00.03
    Antenatal steroid exposure*37 (93%)33 (83%)0.31
Neonates
    Gestation, weeks31.57 (2.68)31.13 (2.4)0.43
    Gestation <28 wk5 (12.5%)5 (12.5%)0.63
    Birth weight, grams1605 (589)1498 (542)0.40
    Male sex27 (67.5%)22 (55%)0.36
    Outborn21 (52%)23 (57%)0.82
    Resuscitated at birth7 (17%)7 (17%)>0.99
    APGAR at 5 min8 (7–8)8 (7–8)0.1
    Hypoglycemia8 (20%)4 (10%)0.35
    Duration of intervention (seconds)§170 (150, 200)160 (148, 185)0.57
    Time to surfactant (minutes)§50 (15, 120)60 (22, 160)0.48
    Ventilated till primary outcome10 (25%)6 (15%)0.40
    Minutes of ventilation (till primary outcome)§112 (85, 180)135 (92, 188)0.89
    Baseline FiO240 (38, 45)40 (35, 42)0.17
    Baseline PEEP6 (5, 6)6 (5, 7)0.54
    Baseline pH7.27 (0.07)7.28 (0.07)0.42
    Baseline PaCO2 (mmHg)46.5 (9.8)45.3 (9.4)0.59
    Baseline PaO2 (mmHg)§80.2 (61.7, 152.1)91.4 (70.4, 203.1)0.22

Data are presented as mean (SD) or n (%), unless indicated otherwise

PEEP Positive end expiratory pressure; RR Risk ratio; SIB Self-inflating bag; TPR T-piece resuscitator

*Received at least 1 dose of antenatal steroids 6 h prior to delivery

† Required more than 2 min of positive pressure ventilation in the delivery room

‡Blood sugar value less than 40 mg/dl

Variables with § are expressed as median (25th centile, 75th centile)

Patients’ demographic and baseline characteristics Data are presented as mean (SD) or n (%), unless indicated otherwise PEEP Positive end expiratory pressure; RR Risk ratio; SIB Self-inflating bag; TPR T-piece resuscitator *Received at least 1 dose of antenatal steroids 6 h prior to delivery † Required more than 2 min of positive pressure ventilation in the delivery room Blood sugar value less than 40 mg/dl Variables with § are expressed as median (25th centile, 75th centile) Blood gas variables of both groups are displayed in Table 2. A significant drop in PaCO2 at 1 h was found in the SIB group as compared to the TPR group [SIB: 8.96 ± 9.06 mmHg vs. TPR: 1.37 ± 9.06 mmHg, mean difference 7.58 mmHg, (95% CI: 3.78–11.4), p < 0.01]. However, the differential reduction did not persist at 6 h (Table 2).
Table 2

Blood gas variables and other neonatal outcomes

SIB Group(n = 40)TPR Group(n = 40)RR/ Mn diff./Md diff.P
Mean change in PaCO2 (1 h)
    Overall (n = 80)8.96 (9.06)1.37 (9.06)7.58 (3.78, 11.4)<0.01
    NIV subgroup (n = 55)8.74 (9.22)0.71 (8.69)8.02 (3.18, 12.87)<0.01
Mean change in PaCO2 (6 h)
    Overall (n = 80)13.01 (11.59)10.72 (10.37)2.29 (−2.60, 7.18)0.35
    NIV subgroup (n = 55)14.41 (11.35)11.37 (9.86)3.05 (−2.69, 8.79)0.29
1 h after manual ventilation
    pH7.34 (0.08)7.31 (0.06)0.03 (−0.009, 0.06)0.06
    PaCO237.5 (9.74)44.0 (8.13)−6.4 (−10.4, −2.4)<0.01
    PaO2*105.9 (76, 158.3)99.9 (83,138.4)2.05 (−18.7, 23.8)0.85
    FiO2*35 (32, 36)34 (30, 36)0 (0, 2)0.39
    PEEP*5 (5, 6)5 (5, 6)0 (−1, 0)0.14
6 h after manual ventilation
    pH7.37 (0.09)7.38 (0.077)−0.01 (−0.05,0.028)0.6
    PaCO233.50 (8.48)34.63 (8.67)−1.12 (−4.94, 2.69)0.55
    PaO2*81.1 (67.1,118.4)92.3 (64.6,124)−4.2 (−22.1, 16.7)0.63
    FiO2*31 (30–32)30 (25–32)1.5 (0, 4)0.06
    PEEP *5 (5, 6)5 (5, 6)00.70
Need of 2nd dose of surfactant31 (77%)22 (55%)1.41 (1.02, 1.94)0.03
Need of ventilation beyond 6 h7 (17%)4 (10%)1.75 (0.56, 5.43)0.52
Days of invasive ventilation*0 (0, 0)0 (0, 10)00.15
Days of non-invasive ventilation*5 (3, 9)4 (3, 12)0 (−1.0, 2.0)0.77
Days of oxygen support*5.5 (4, 10)4 (3, 8)1.0 (−1.0, 3.0)0.36
Days of hospital stay*25 (14, 35)29 (19, 36)−2.0 (−10.0, 5.0)0.63
Death2 (5%)2 (5%)1.0>0.99
Severe IVH (grade 3 or 4)2 (5%)2 (5%)1.0>0.99
PDA requiring medical treatment12 (30%)11 (27%)1.09 (0.54, 2.18)>0.99
ROP requiring treatment5 (12.5%)4 (10%)1.25 (0.36, 4.34)>0.99
BPD9 (22%)6 (15%)1.5 (0.58, 3.8)0.57
NEC2 (5%)1 (2.5%)2 (0.19, 20.4)>0.99
Late onset sepsis6 (15%)3 (8.1%)0.5 (0.14, 1.8)0.29

Data are presented as mean (SD) or n (%), unless indicated otherwise, Variables with * are presented as median (25th centile, 75th centile)

BPD Bronchopulmonary dysplasia; IVH Intraventricular hemorrhage; Md diff. Median difference; Mn diff. Mean difference; NEC Necrotising enterocolitis; NIV Non-invasive ventilation; PDA Patent ductus arteriosus; PEEP Positive end expiratory pressure; ROP Retinopathy of prematurity; RR Risk ratio; SIB Self-inflating bag; TPR T-piece resuscitator

Blood gas variables and other neonatal outcomes Data are presented as mean (SD) or n (%), unless indicated otherwise, Variables with * are presented as median (25th centile, 75th centile) BPD Bronchopulmonary dysplasia; IVH Intraventricular hemorrhage; Md diff. Median difference; Mn diff. Mean difference; NEC Necrotising enterocolitis; NIV Non-invasive ventilation; PDA Patent ductus arteriosus; PEEP Positive end expiratory pressure; ROP Retinopathy of prematurity; RR Risk ratio; SIB Self-inflating bag; TPR T-piece resuscitator Neonates on NIV who received surfactant and then returned to NIV were analyzed as a separate subgroup (64 infants; 30 in SIB group and 34 in TPR group). The SIB group still demonstrated a significant drop in PaCO2 [SIB 8.74 mmHg, TPR 0.71 mmHg, Mean difference 8.02 mmHg, (95% CI: 3.18–12.87), p < 0.01]. The box plot shows that the magnitude of drop in the post intervention PaCO2 was significantly greater in the SIB group especially at 1 h (Fig. 2).
Fig. 2

Drop in PaCO2 levels in SIB vs. TPR group. SIB Self-inflating bag; TPR T-piece resuscitator

Drop in PaCO2 levels in SIB vs. TPR group. SIB Self-inflating bag; TPR T-piece resuscitator Analysis of covariance model of linear regression was used to assess the effect of the MVD on the primary outcome after adjusting for the baseline PaCO2. That the regression lines for the SIB group/TPR group are nearly parallel and well separated (Fig. 3), confirms the drop in PaCO2 to be independent of the baseline values.
Fig. 3

Analysis of covariance (Baseline PaCO2 vs. 1 h PaCO2)

Analysis of covariance (Baseline PaCO2 vs. 1 h PaCO2) Repeated measures of PaCO2 (baseline, 1 h and 6 h) were subjected to a factorial analysis of variance to determine the device effect (Type of MVD - SIB or TPR) and time effect (1 h and 6 h after manual ventilation). The change in PaCO2 was found to be significantly associated with time [F (1, 77) = 34.47, p < 0.01- time effect] rather than the device F (1, 77) = 3.31, p = 0.07- device effect]. However, the interaction effect (device x time) was significant [F (1, 77) = 3.69, p = 0.03], indicating the effect of the MVD on the mean change in PaCO2 at 1 h. Outcomes related to neonatal morbidity were similar in both groups (Table 2). One neonate in the SIB group developed pneumothorax at 24 h of life requiring intercostal drainage.

Discussion

This clinical trial evaluates the effect of manual ventilation with a T-piece resuscitator or self-inflating bag during surfactant administration by measuring the mean change in arterial carbon dioxide levels after administration. The authors found that the fall in PaCO2 level was significantly greater with SIB as compared to TPR, 1 h post intervention. The strength of present study includes a robust randomized controlled trial design adequately powered for the primary outcome. To authors’ knowledge, there is no RCT until date comparing the effect of SIB/TPR on carbon dioxide elimination in neonates. Selection of an objective primary outcome and completeness of follow up eliminated detection and attrition bias. Several studies performed on manikins have evaluated MVDs for tidal volume and peak inspiratory pressure. Dawson et al. randomized 80 neonates <29 wk of gestation to the two MVDs. A trend towards lower tidal volumes was reported in the TPR group (6.6 ml/kg vs. 9.2 ml/kg, p = 0.34) [8]. A multicenter crossover trial by Szyld et al. reported that TPR generated lower and less variable inflation pressures than SIB (26 vs. 28 cm H2O) in neonates >26 wk of gestation. Fewer neonates in the TPR group received peak inspiratory pressure > 25 cm H2O (5% vs. 33%) [15]. In the present study, authors chose to assess the effect of MVDs on the PaCO2 levels, especially since surfactant could exaggerate the response to manual ventilation. The difference in 1 h PaCO2 values in present study could be attributed only to the type of MVD used during surfactant administration. Factors that could affect alveolar ventilation such as, extent of lung disease, initial ventilatory requirements, timing and dose of surfactant administered, were similar in the study groups. Furthermore, the main outcome was uninfluenced by the magnitude of individual PaCO2 reading. Significant drop in PaCO2 (in SIB group) even after restricting the analysis to the NIV subgroup strengthens the findings and captures the impact of MVD on the PaCO2 values. The lack of effect on the 6 h PaCO2 values could be explained by the ongoing non-invasive respiratory support in both groups. Very few studies have evaluated the effects of MVDs on clinical outcomes in neonates. Thakur et al. randomized 90 neonates >26 wk of gestation requiring delivery room resuscitation to TPR or SIB. They demonstrated that neonates in TPR group required positive pressure ventilation for a lesser duration and had lower intubation rates [16]. A Brazilian prospective cohort study of 1962 inborn preterm infants showed that TPR, as compared to SIB, increased the odds of survival to discharge without major morbidities [17]. In the present study, the requirement of second dose of surfactant was significantly lower in the TPR group. Consistent delivery of positive end expiratory pressure and better lung recruitment with the TPR could possibly explain this outcome. A trend towards lower FiO2 requirement at 6 h in the TPR group is in line with these findings. The arterial oxygen concentration in both groups, especially in the early phase, were high from a comparison with current practice. However, significant hyperoxia has also been reported in preterm neonates transported on non-synchronized mechanical ventilation [18]. The present study has certain limitations. Clinicians and nurses could not be blinded to the intervention. Real time monitoring and regulation of pressures generated by the SIB was not done in the study patients. In the absence of a manometer to monitor the peak inspiratory pressure generated, a bias towards administration of higher positive pressure in the SIB group could have existed. Minor group differences due to operator variability, especially with SIB, also could not be ruled out. A blood gas immediately after surfactant administration would have provided information specific to the effect of MVD. Only 12% of neonates in this study were below 28 wk of gestation and a large proportion of infants were outborn (55%). This could limit the generalizabilty of present study results. The present study highlights the need to regulate pressures generated by MVDs even during brief periods of manual ventilation. RCTs with large sample size addressing clinically important outcomes in high risk neonates (gestation <28 wk) and follow up beyond the neonatal period would be required to assess the implications of the findings from the present study.

Conclusions

T-piece resuscitator results in smaller reduction in PaCO2 levels compared to self inflating bag, when used for brief manual ventilation during surfactant administration.
  18 in total

Review 1.  Comparison of the T-piece resuscitator with other neonatal manual ventilation devices: a qualitative review.

Authors:  Colin Patrick Hawkes; C Anthony Ryan; Eugene Michael Dempsey
Journal:  Resuscitation       Date:  2012-01-04       Impact factor: 5.262

2.  Variation in inspiratory time and tidal volume with T-piece neonatal resuscitator: association with operator experience and distraction.

Authors:  S McHale; M Thomas; E Hayden; K Bergin; N McCallion; E J Molloy
Journal:  Resuscitation       Date:  2008-08-08       Impact factor: 5.262

3.  Sustained inflations: comparing three neonatal resuscitation devices.

Authors:  Claus Klingenberg; Jennifer A Dawson; Angela Gerber; C Omar F Kamlin; Peter G Davis; Colin J Morley
Journal:  Neonatology       Date:  2011-01-26       Impact factor: 4.035

4.  T-piece versus self-inflating bag ventilation in preterm neonates at birth.

Authors:  Ruth Guinsburg; Maria Fernanda Branco de Almeida; Junia Sampel de Castro; Walusa Assad Gonçalves-Ferri; Patricia Franco Marques; Jamil Pedro Siqueira Caldas; Vera Lucia Jornada Krebs; Ligia Maria Suppo de Souza Rugolo; João Henrique Carvalho Leme de Almeida; Jorge Hecker Luz; Renato S Procianoy; José Luiz Muniz Bandeira Duarte; Marcia Gomes Penido; Daniela Marques de Lima Mota Ferreira; Navantino Alves Filho; Edna Maria de Albuquerque Diniz; Juliana Paula Santos; Ana Lucia Acquesta; Cristina Nunes Dos Santos; Maria Rafaela Conde Gonzalez; Regina Pg Vieira Cavalcanti da Silva; Jucile Meneses; José Maria de Andrade Lopes; Franciscó Eulógio Martinez
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2017-06-29       Impact factor: 5.747

5.  Comparison of devices for newborn ventilation in the delivery room.

Authors:  Edgardo Szyld; Adriana Aguilar; Gabriel A Musante; Nestor Vain; Luis Prudent; Jorge Fabres; Waldemar A Carlo
Journal:  J Pediatr       Date:  2014-03-29       Impact factor: 4.406

6.  Oxygenation with T-piece versus self-inflating bag for ventilation of extremely preterm infants at birth: a randomized controlled trial.

Authors:  Jennifer A Dawson; Georg M Schmölzer; C Omar F Kamlin; Arjan B Te Pas; Colm P F O'Donnell; Susan M Donath; Peter G Davis; Colin J Morley
Journal:  J Pediatr       Date:  2011-01-15       Impact factor: 4.406

7.  Early hypocarbia of preterm infants: its relationship to periventricular leukomalacia and cerebral palsy, and its perinatal risk factors.

Authors:  M Murase; A Ishida
Journal:  Acta Paediatr       Date:  2005-01       Impact factor: 2.299

8.  Comparison of three manual ventilation devices using an intubated mannequin.

Authors:  S G Hussey; C A Ryan; B P Murphy
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-11       Impact factor: 5.747

9.  Manual ventilation devices in neonatal resuscitation: tidal volume and positive pressure-provision.

Authors:  Charles C Roehr; Marcus Kelm; Hendrik S Fischer; Christoph Bührer; Gerd Schmalisch; Hans Proquitté
Journal:  Resuscitation       Date:  2009-11-17       Impact factor: 5.262

10.  Manual Ventilation and Sustained Lung Inflation in an Experimental Model: Influence of Equipment Type and Operator's Training.

Authors:  Cristiane do Prado; Ruth Guinsburg; Maria Fernanda Branco de Almeida; Renata Suman Mascaretti; Luciana Assis Vale; Luciana Branco Haddad; Celso Moura Rebello
Journal:  PLoS One       Date:  2016-02-09       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.