Barrier enclosure using a Mayo stand and plastic sheet during cardiopulmonary resuscitation in patients and its effect on reducing visible aerosol dispersion on healthcare workers.

Hiren J Mehta1, Mary Patterson2, Nikolaus Gravenstein3.   

Abstract

Entities:  

Year:  2020        PMID: 32653587      PMCID: PMC7241360          DOI: 10.1016/j.jcrc.2020.05.007

Source DB:  PubMed          Journal:  J Crit Care        ISSN: 0883-9441            Impact factor:   3.425


× No keyword cloud information.
Hospitals around the world are ramping up their services to accommodate patients infected with severe acute respiratory distress syndrome coronavirus 2 (SARS-CoV-2). Although healthcare workers (HCW) are at significant risk for infection with SARS-CoV-2, the deficit of personal protective equipment (PPE) to protect HCWs from infection has been a major recurring concern. Recommendations for the use of personal protective equipment to protect against SARS-CoV-2 exposure by HCWs were recently published by the World Health Organization and the US Centers for Disease Control and Prevention. For aerosol-generating procedure, N95 respirators, eye protection, isolation gowns, and gloves were recommended. Coveralls, boots with a cover, and hair coverings were not part of the recommended protective clothing [1]. The World Health Organization (WHO) has classified cardiopulmonary resuscitation (CPR) as an aerosol-generating procedure [2]. The center for disease control (CDC) along with American Heart Association (AHA) has issued recommendations for CPR, endotracheal intubation, and noninvasive ventilation to minimize aerosol generation and protect the HCW [3]. Some of these recommendations especially for CPR include holding chest compression during intubation and trying to establish airway early in the resuscitation to minimize aerosol generation. Barrier enclosures have been previously described to minimize contamination of HCW during aerosol generating procedures [4]. However, no such modality has been described to protect HCWs while performing CPR. Here we describe a model to lower contamination by aerosol of HCWs during simulated CPR.

Method

The barrier used is Mayo stand with transparent plastic sheet. We used clear plastic shower liner (70  inchin. x 72 in.) covering the simulated patient. We pulled the mayo stand as shown in figure over the head of the patient, adjusted the height of the stand, and covered it with the plastic sheet to fully enclose the mouth and face of the patient before we started the CPR. In our simulation, the person performing chest compression was stationed to the right side of the manikin ‘victim’ who is not intubated. The model simulates the beginning of cardiac arrest resuscitation before a definitive airway is established. The HCW performing chest compression is donned in appropriate PPE as per CDC guidelines. Glow germ™ was used to simulate aerosols during the CPR. We performed this simulation with and without barrier and took photographs of HCW illuminated by ultraviolet light to assess the visible level of contamination with the fluorescent aerosolized particles.

Results

UV light illuminated images below demonstrate exposure to HCW using complete PPE as per CDC and WHO guidelines with and without the barrier described previously. Fig. 1 : Draped Mayo stand barrier during CPR. Fig. 2 shows barrier stand and HCW performing CPR with barrier sheet on to decrease contamination from any generated aerosol. Fig. 3, Fig. 4, Fig. 5, Fig. 6 show dispersion of aerosols on the PPE of the HCW performing CPR without barrier enclosure. Fig. 7, Fig. 8 show minimal contamination of HCW with aerosols when barrier stand and cover were used during CPR.
Fig. 1

Mayo stand with plastic drape decreasing aersol exposure to HCW performing CPR.

Fig. 2

Mayo stand with plastic drape decreasing aersol exposure to HCW performing CPR.

Fig. 3

Dispersion of aerosols on the PPE of the HCW performing CPR without barrier enclosure.

Fig. 4

Dispersion of aerosols on the PPE of the HCW performing CPR without barrier enclosure.

Fig. 5

Dispersion of aerosols on the PPE of the HCW performing CPR without barrier enclosure.

Fig. 6

Dispersion of aerosols on the PPE of the HCW performing CPR without barrier enclosure.

Fig. 7

Minimal contamination of HCW with aerosols when barrier stand and cover were used during CPR.

Fig. 8

Minimal contamination of HCW with aerosols when barrier stand and cover were used during CPR.

Mayo stand with plastic drape decreasing aersol exposure to HCW performing CPR. Mayo stand with plastic drape decreasing aersol exposure to HCW performing CPR. Dispersion of aerosols on the PPE of the HCW performing CPR without barrier enclosure. Dispersion of aerosols on the PPE of the HCW performing CPR without barrier enclosure. Dispersion of aerosols on the PPE of the HCW performing CPR without barrier enclosure. Dispersion of aerosols on the PPE of the HCW performing CPR without barrier enclosure. Minimal contamination of HCW with aerosols when barrier stand and cover were used during CPR. Minimal contamination of HCW with aerosols when barrier stand and cover were used during CPR.

Discussion

Despite personal protective equipment, fluorescent markers were found on the uncovered areas of neck, skin and hair of the HCW providing resuscitation to a simulated SARS-CoV-2 patient without barrier enclosure. The findings suggest that the current recommendations for personal protective equipment may not fully prevent exposures while performing CPR. Clothing that covers all skin and hair may further diminish exposure risk. However using a simple barrier stand with barrier sheet can significantly decrease aerosol dispersion in the air and onto the healthcare workers, potentially better protecting them from significant exposure. Inhalation of aerosols and exposure risks associated with doffing were not evaluated. The simulated healthcare setting and the surrogate measures of exposure are the primary limitations. Because this was a mannikin simulation, it is uncertain how the results may translate to actual patient care.
  1 in total

1.  Barrier Enclosure during Endotracheal Intubation.

Authors:  Robert Canelli; Christopher W Connor; Mauricio Gonzalez; Ala Nozari; Rafael Ortega
Journal:  N Engl J Med       Date:  2020-04-03       Impact factor: 91.245

  1 in total
  2 in total

1.  Simulation and evaluation of the protective barrier enclosure for cardiopulmonary resuscitation.

Authors:  Farah H Jamaludin; Shahridan M Fathil; Tuck-Whye Wong; Maryam S Termizi; Steven H Hsu; Hsien Y Lai
Journal:  Resusc Plus       Date:  2021-11-13

2.  Thin foil body-shield resuscitation barrier device to protect from blood: an experimental study.

Authors:  Martin Hermann; Wolfgang Lederer; Markus Isser; Valentin Bilgeri; Andreas Klinger
Journal:  Sci Rep       Date:  2022-08-09       Impact factor: 4.996

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.