Effect of manual, preloaded, and automated preloaded injectors on corneal incision architecture after IOL implantation.

Michela Cennamo1, Eleonora Favuzza, Maria Cristina Salvatici, Gabriele Giuranno, Matilde Buzzi, Rita Mencucci.   

Abstract

PURPOSE: To analyze the effects on corneal morphology of manual, preloaded, and automated preloaded intraocular lens (IOL) injectors in eye bank human corneas by environmental scanning electron microscopy (ESEM) and in patients after phacoemulsification using anterior segment optical coherence tomography (AS-OCT). SETTINGS: Eye Clinic, Careggi University Hospital, Florence, Italy.
DESIGN: Retrospective and experimental study.
METHODS: Seventy-eight corneal incisions were examined after IOL implantation: 30 in human corneas mounted on an artificial chamber using ESEM (ex vivo); 48 in patients undergoing phacoemulsification (in vivo). Three different injectors were used for both analyses: manual (Monarch III, n = 26), manual preloaded (UltraSert, n = 26), and automated preloaded system (AutonoMe, n = 26). Thirty IOLs were implanted in the ex vivo study: 5 intermediate and 5 high dioptric powers for AcrySof IQ (Monarch and UltraSert) and for Clareon (AutonoMe) IOLs. In the in vivo analysis, 16 corneal wounds for each injector were evaluated using AS-OCT; in the ex vivo study, incision width was measured and Descemet membrane detachment, posterior wound retraction, and posterior gape were analyzed.
RESULTS: In the eye bank corneas, the incision width was significantly wider in the high dioptric power IOL manual subgroup (P < .05), with more Descemet tearing compared with AutonoMe. In the in vivo study, the incidence of Descemet membrane detachment, posterior gape, and wound retraction was lower in the automated preloaded group at 1 postoperative day 1.
CONCLUSIONS: The automated preloaded injector ensured less trauma to the wound and contributed to preserving the endothelial side of the incision even during the implantation of high-power IOLs and in the early postoperative period.

Entities:  

Mesh:

Year:  2020        PMID: 32649432     DOI: 10.1097/j.jcrs.0000000000000295

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

1.  Experimental Study on Delivery Performance of an Automated Preloaded Intraocular Lens Injector System for Corneal and Sclerocorneal Incisions.

Authors:  Tetsuro Oshika; Noriyuki Sasaki
Journal:  J Ophthalmol       Date:  2021-03-31       Impact factor: 1.909

2.  One-year multicenter evaluation of a new hydrophobic acrylic intraocular lens with hydroxyethyl methacrylate in an automated preloaded delivery system.

Authors:  Tetsuro Oshika; Noriyuki Sasaki
Journal:  J Cataract Refract Surg       Date:  2022-03-01       Impact factor: 3.528

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.