The estimation of plasma valproate by gas-liquid chromatography.

M H Wood, D C Sampson, W J Hensley.   

Abstract

A gas chromatographic method for the plasma assay of the anticonvulsant, sodium valproate, is described. Derivatization is not necessary. 200 microliter plasma are required for a single estimation. The method involves a chloroform extraction of valproate and the internal standard, cyclopentane carboxylic acid, from acidified plasma. Gas-liquid chromatography using the stationary phase 10% SP-216-PS gives complete separation of valproate and the internal standard in eight minutes. The limit of detection is 20 mumol valproate/1 plasma (equivalent to 40 pmol on column). This is well below the lower therapeutic plasma level. The between-run precision of the method indicates a variation for each sample within+/-3% of its mean value.

Entities:  

Mesh:

Substances:

Year:  1977        PMID: 326442     DOI: 10.1016/0009-8981(77)90239-x

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  1 in total

1.  Free sodium valproate monitoring.

Authors:  E J Roman; P Ponniah; J B Lambert; N Buchanan
Journal:  Br J Clin Pharmacol       Date:  1982-03       Impact factor: 4.335