Accuracy and Reliability of Commercial Wrist-Worn Pulse Oximeter During Normobaric Hypoxia Exposure Under Resting Conditions.

Claire J Lauterbach1, Phebe A Romano1, Luke A Greisler1, Richard A Brindle1,2, Kevin R Ford1, Matthew R Kuennen1.   

Abstract

Purpose: The present study analyzed peripheral blood oxygen saturation (SpO2) and heart rate (HR) measurements taken on the Garmin fēnix® 5X Plus watch, comparing them to measurements taken on a standard medical-grade pulse oximeter during normobaric hypoxia exposure under resting conditions.
Methods: Thirteen women (mean ± SD: Age 20 ± 1 years, height 165 ± 5 cm, mass, 67 ± 9 kg) and ten men (mean ± SD: Age 21 ± 3 years, height 177 ± 6 cm, mass 78 ± 11 kg) sat inside a customized environmental chamber while the fraction of inspired oxygen (FIO2) was adjusted to simulate altitudes of 12,000; 10,000; 8,000; 6,000; and 900 ft. The novel commercial device (Garmin fēnix®) and a medical-grade pulse oximeter (Nonin® 7500) were used to measure SpO2 and HR in triplicate at each simulated altitude. Bland-Altman analyses were used to assess differences between methods.
Results: Bland-Altman analysis indicated 3.3% bias for SpO2 measurements taken on the Garmin fēnix® at 12,000 ft of simulated altitude (limits of agreement: -1.9-8.6%). Mean differences in SpO2 measurements were smaller at the remaining simulated altitudes, where bias measurements ranged from 0.7% to 0.8%. The Garmin fēnix® also underestimated heart rate, but those discrepancies were minimal (bias measurements at all simulated altitude exposures were < 1.0 bpm). Conclusions: With the exception of readings taken at 12,000 ft of simulated altitude, the Garmin fēnix® exhibits minimal overestimation of SpO2 and minimal underestimation of HR during simulated altitude exposure. These data suggest the Garmin fēnix® watch may be a viable method to monitor SpO2 and HR under most ambient environmental conditions.

Entities:  

Keywords:  Altitude; athletics; track and field; wearable technology

Year:  2020        PMID: 32633688     DOI: 10.1080/02701367.2020.1759768

Source DB:  PubMed          Journal:  Res Q Exerc Sport        ISSN: 0270-1367            Impact factor:   2.500


  3 in total

1.  Time Synchronization of Multimodal Physiological Signals through Alignment of Common Signal Types and Its Technical Considerations in Digital Health.

Authors:  Ran Xiao; Cheng Ding; Xiao Hu
Journal:  J Imaging       Date:  2022-04-21

2.  Modulation of thalamic network connectivity using transcranial direct current stimulation based on resting-state functional magnetic resonance imaging to improve hypoxia-induced cognitive impairments.

Authors:  Guo Dalong; Qin Yufei; Yang Lei; Li Pengfei; Ye Anqi; Guo Zichuan; Wang Cong; Zhou Yubin
Journal:  Front Neurosci       Date:  2022-08-25       Impact factor: 5.152

3.  Commercial smartwatch with pulse oximeter detects short-time hypoxemia as well as standard medical-grade device: Validation study.

Authors:  Jakub Rafl; Thomas E Bachman; Veronika Rafl-Huttova; Simon Walzel; Martin Rozanek
Journal:  Digit Health       Date:  2022-10-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.