Does the Inion OTPSTM absorbable plating system induce higher foreign-body reaction than titanium implants? An experimental randomized comparative study in rabbits.

Byron Chalidis1, Dimitrios Kitridis1, Panagiotis Savvidis1, Apostolos Papalois2, Panagiotis Givissis1.   

Abstract

Hardware removal after complete fracture healing is a laborious and traumatic procedure that bears the risks of refracture, infection, and soft tissue damage. Absorbable materials for fracture fixation have the advantage of eliminating the need for removal; however, they may stimulate a considerable inflammatory response. For that reason, we compared in a prospective randomized experimental animal model study the incidence of foreign-body reactions between Inion OTPSTM absorbable plate implants, consisting of carbon trimethylene, L-polylactic acid, and D, L-polylactic acid, and conventional titanium plates. Seventy rabbits were equally divided into two groups. Titanium (Group A) or absorbable (Group B) plates and screws were implanted in the femoral shaft of the rabbits. Wound reaction and laboratory tests including white blood cell count (WBC), erythrocyte sedimentation rate (ESR), and C-reactive protein (CRP) values were measured and compared at different time points. Radiographs and histological analysis of tissue samples from all animals were also used to identify potential foreign-body reactions. WBC showed normal values in both groups. In Group A, ESR levels were higher at all time points but CRP values were normalized at 3 weeks postoperatively. In Group B, ESR and CRP levels reached a peak at 3 weeks and were above normal values throughout the whole experiment (p < 0.001). Radiologic examination revealed periosteal thickening and surrounding sclerosis in 23 out of 35 animals of Group B (66%). All Group A animals had normal radiographic findings. Histopathology showed a typical non-specific foreign-body reaction in all animals in Group B (100%) but none in Group A. However, only four of them (11.8%) developed non-infected wound complications and showed significantly elevated CRP and ESR values (p = 0.054). Although Inion OTPSTM Biodegradable Plating System stimulates a foreign-body reaction, only a small but not negligible percentage of almost 12% may lead to wound problems. However, this difference compared to titanium plates was of borderline not statistically significant. Future studies should focus on the development of absorbable implants with improved biocompatibility aiming to further decrease the potential of foreign-body reaction and eliminate the need for additional reoperative procedures.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 32629429     DOI: 10.1088/1748-605X/aba326

Source DB:  PubMed          Journal:  Biomed Mater        ISSN: 1748-6041            Impact factor:   3.715


  2 in total

1.  Breast implant causes allergic contact dermatitis or foreign body reaction?

Authors:  Hilde M Bosker; Jorrit B Terra; Martin M Stenekes
Journal:  Case Reports Plast Surg Hand Surg       Date:  2020-09-03

2.  Finite element analysis of different fixation methods of screws on absorbable plate for rib fractures.

Authors:  Hang Xue; Zhenhe Zhang; Mengfei Liu; Ze Lin; Yori Endo; Guodong Liu; Bobin Mi; Wu Zhou; Guohui Liu
Journal:  Front Bioeng Biotechnol       Date:  2022-07-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.