Allen Wing-Ho Chu1, Wan-Mui Chan1, Jonathan Daniel Ip1, Cyril Chik-Yan Yip2, Jasper Fuk-Woo Chan3, Kwok-Yung Yuen3, Kelvin Kai-Wang To4.
Abstract
BACKGROUND: The severe shortage of nucleic acid extraction kits during the current COVID-19 pandemic represents a key limiting factor in testing capacity.Entities:
Keywords: COVID-19; Detection; Nucleic acid extraction; Proteinase K; RT-PCR; SARS-CoV-2
Mesh:
Substances:
Year: 2020 PMID: 32629187 PMCID: PMC7309780 DOI: 10.1016/j.jcv.2020.104519
Source DB: PubMed Journal: J Clin Virol ISSN: 1386-6532 Impact factor: 3.168
Comparison of different respiratory specimen processing steps.
| Extraction methods | P values | |||||
|---|---|---|---|---|---|---|
| PKH | Heat only | Direct | PKH vs heat only | PKH vs direct | Heat only vs direct | |
| All Samples | ||||||
| (n = 50) | ||||||
| Positive | 40 (80) | 29 (58) | 28 (56) | 0.001 | 0.002 | 1.000 |
| Negative | 10 (20) | 21 (42) | 22 (44) | |||
| NPS | ||||||
| (n = 25) | ||||||
| Positive | 21 (84) | 15 (60) | 17 (68) | 0.031 | 0.0125 | 0.500 |
| Negative | 4 (16) | 10 (40) | 8 (32) | |||
| Saliva | ||||||
| (n = 25) | ||||||
| Positive | 19 (76) | 14 (56) | 11 (44) | 0.063 | 0.021 | 0.453 |
| Negative | 6 (24) | 11 (44) | 14 (56) | |||
Data are expressed as no. (%).
Fig. 1Comparison of RT-PCR results from different nucleic acid extraction methods. The Ct values were compared between the PKH, heat only and direct groups. PKH: specimens pre-incubated with proteinase K for 15 min at 56 °C, followed by 5 min at 98 °C. Heat only: specimens were heated for 5 min at 98 °C. Direct: no pre-processing step before RT-PCR. Original: Total nucleic acid was extracted using NucliSENS easyMAG extraction system. *, <0.05; ****, <0.0001.
Evaluation of the limit of detection (LOD) for the three processing methods using serially diluted NPS and saliva specimens from a COVID-19 patient.
| NPS Sample | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Sample Dilution | PKH | Heat Only | Direct | ||||||
| Test 1 | Test 2 | Test 3 | Test 1 | Test 2 | Test 3 | Test 1 | Test 2 | Test 3 | |
| 1:10 | 35.65 | 35.37 | 35.16 | 36.74 | 36.65 | 36.57 | 38.18 | 36.42 | 40.00 |
| 1:100 | 38.98 | 37.23 | 38.03 | – | – | – | 40.00 | 40.00 | 38.30 |
| 1:1000 | 37.77 | 40.00 | 40.00 | – | – | – | – | – | – |
| 1:10000 | – | – | – | – | – | – | – | – | – |
| 1:100000 | – | – | – | – | – | – | – | – | – |
Data represent Ct values. The Ct values for the total nucleic acid of the undiluted NPS and saliva specimen extracted using NucliSENS easyMAG extraction system is 21.59 and 21.14, respectively.