Techniques for and uncertainties of MRI-based reconstruction of titanium tandem and ring brachytherapy applicators.

Irina Malajovich1, Shibu Anamalayil2, Otilia Valentina Dolney2, Boon-Keng Kevin Teo2, William Tristram Arscott3, Neil K Taunk4.   

Abstract

PURPOSE: Eliminating patient computed tomography (CT) scans for tandem and ring (T&R) brachytherapy can reduce overall procedure time and eliminates imaging dose. However, reconstructing titanium applicators in magnetic resonance imaging (MRI) is challenging. We evaluated the uncertainty of different applicator reconstruction workflows in MR-guided brachytherapy, and assessed the clinical impact of reconstruction uncertainties. METHODS AND MATERIALS: Titanium MRI-compatible T&Rs with aqueous gel in the buildup cap were reconstructed on CTs and MRIs to assess the uncertainties of four different workflows. Reconstruction was performed using (1) proton density-weighted MRIs with solid applicator from a library, (2) applicator-only reference CT fused with MRIs, (3) T2-weighted (T2W) MRIs following GEC-ESTRO guidelines, and (4) patient CTs fused with patient MRIs with in situ applicators. We evaluated dwell positions and plan quality differences using high-risk clinical target volume coverage, and EQD2 D2cc of rectum, sigmoid, bladder, and small bowel.
RESULTS: The 2σ uncertainty for dwell positions for each workflow were (1) 2.7 mm for both ring and tandem, (2) 1.4 mm ring and 0.8 mm tandem, (3) 0.2 mm ring and 0.8 mm tandem, and (4) 1.9 mm ring and 0.4 mm tandem. Reconstruction uncertainties resulted in dose variations within acceptable levels (below 10%) except for (1) which resulted in larger dose to the rectum (20%). Dose uncertainties were similar between reference CT and patient CT.
CONCLUSIONS: Reconstruction with a reference CT results in similar uncertainty to a patient CT. T2W MRI plans have acceptable uncertainty levels for the applicator reconstruction and resulting dose distributions.
Copyright © 2020 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Brachytherapy; Cervical cancer; Reconstruction; Titanium

Mesh:

Substances:

Year:  2020        PMID: 32624406     DOI: 10.1016/j.brachy.2020.05.005

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  1 in total

1.  Practical needle selection for Vienna-style applicators: improving therapeutic ratio in hybrid intracavitary-interstitial brachytherapy.

Authors:  David A Martin; Neil K Taunk; Shibu Anamalayil; Vatsal Mangal; Jaclyn Marcel; Emily Hubley
Journal:  J Contemp Brachytherapy       Date:  2021-10-29
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.