Comparing the Mutual Interchangeability of ECOM, FloTrac/Vigileo, 3D-TEE, and ITD-PAC Cardiac Output Measuring Systems in Coronary Artery Bypass Grafting.

Adam J Milam1, Farhad Ghoddoussi2, Jon Lucaj3, Spurthy Narreddy3, Nakul Kumar4, Vennela Reddy3, Joffer Hakim3, Sandeep H Krishnan5.   

Abstract

OBJECTIVE: The aim of this study was to compare the mutual interchangeability of 4 cardiac output measuring devices by comparing their accuracy, precision, and trending ability.
DESIGN: A single-center prospective observational study.
DESIGN: Nonuniversity teaching hospital, single center. PARTICIPANTS: Forty-four consecutive patients scheduled for elective, nonemergent coronary artery bypass grafting (CABG).
INTERVENTIONS: The cardiac output was measured for each participant using 4 methods: intermittent thermodilution via pulmonary artery catheter (ITD-PAC), Endotracheal Cardiac Output Monitor (ECOM), FloTrac/Vigileo System (FLOTRAC), and 3-dimensional transesophageal echocardiography (3D-TEE).
MEASUREMENTS AND MAIN RESULTS: Measurements were performed simultaneously at 5 time points: presternotomy, poststernotomy, before cardiopulmonary bypass, after cardiopulmonary bypass, and after sternal closure. A series of statistical and comparison analyses including ANOVA, Pearson correlation, Bland-Altman plots, quadrant plots, and polar plots were performed, and inherent precision for each method and percent errors for mutual interchangeability were calculated. For the 6 two-by-two comparisons of the methods, the Pearson correlation coefficients (r), the percentage errors (% error), and concordance ratios (CR) were as follows: ECOM_versus_ITD-PAC (r = 0.611, % error = 53%, CR = 75%); FLOTRAC_versus_ITD-PAC (r = 0.676, % error = 49%, CR = 77%); 3D-TEE versus ITD-PAC (r = 0.538, % error = 64%, CR = 67%); FLOTRAC_versus_ECOM (r = 0.627, % error = 51%, CR = 75%); 3D-TEE_versus ECOM (r = 0.423, % error = 70%, CR = 60%), and 3D-TEE_versus_FLOTRAC (r = 0.602, % error = 59%, CR = 61%).
CONCLUSIONS: Based on the recommended statistical measures of interchangeability, ECOM, FLOTRAC, and 3D-TEE are not interchangeable with each other or to the reference standard invasive ITD-PAC method in patients undergoing nonemergent cardiac bypass surgery. Despite the negative result in this study and the majority of previous studies, these less-invasive methods of CO have continued to be used in the hemodynamic management of patients. Each device has its own distinct technical features and inherent limitations; it is clear that no single device can be used universally for all patients. Therefore, different methods or devices should be chosen based on individual patient conditions, including the degree of invasiveness, measurement performance, and the ability to provide real-time, continuous CO readings.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  3D transesophageal echocardiography; FloTrac/Vigileo; cardiac output (CO) monitoring; endotracheal cardiac output monitoring (ECOM); interchangeability; intermittent thermodilution method; pulmonary arterial catheter

Mesh:

Year:  2020        PMID: 32622708     DOI: 10.1053/j.jvca.2020.03.048

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  1 in total

1.  Comparison of accuracy of two uncalibrated pulse contour cardiac output monitors in off-pump coronary artery bypass surgery patients using pulmonary artery catheter-thermodilution as a reference.

Authors:  Ramakrishna Mukkamala; Benjamin A Kohl; Aman Mahajan
Journal:  BMC Anesthesiol       Date:  2021-07-10       Impact factor: 2.217

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.