The influence of collimators on SPECT center of rotation measurements: artifact generation and acceptance testing.

M D Cerqueira1, D Matsuoka, J L Ritchie, G D Harp.   

Abstract

Misalignment between the electronic and mechanical axes of rotation will result in artifact generation and image degradation during single photon emission computed tomography (SPECT) reconstruction. Acceptance and quality control testing procedures have not emphasized the variability in center of rotation (COR) measurements caused by collimators and the need to verify uniformity across the full collimator field of view (FOV). Variation from the mean COR across the FOV was tested in four different collimators using multiple point source acquisitions. The mean COR was different for each collimator and two of the four had a greater than 0.5 pixel difference from the mean COR on some area of the FOV. This variation makes these collimators unacceptable for SPECT acquisition. Thus, initial acceptance testing of SPECT collimators should verify a uniform COR across the full FOV and collimators with a variability from the mean COR greater than 0.5 pixels should be rejected.

Mesh:

Substances:

Year:  1988        PMID: 3261332

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  4 in total

1.  Single photon-emission computed tomography.

Authors:  Thomas A Holly; Brian G Abbott; Mouaz Al-Mallah; Dennis A Calnon; Mylan C Cohen; Frank P DiFilippo; Edward P Ficaro; Michael R Freeman; Robert C Hendel; Diwakar Jain; Scott M Leonard; Kenneth J Nichols; Donna M Polk; Prem Soman
Journal:  J Nucl Cardiol       Date:  2010-10       Impact factor: 5.952

2.  Instrumentation quality assurance and performance.

Authors:  Kenneth J Nichols; Stephen L Bacharach; Steven R Bergmann; S James Cullom; Edward P Ficaro; James R Galt; Gary V Heller; Jonathan Links; Josef Machac
Journal:  J Nucl Cardiol       Date:  2006-11       Impact factor: 5.952

Review 3.  Single Photon Emission Computed Tomography (SPECT) Myocardial Perfusion Imaging Guidelines: Instrumentation, Acquisition, Processing, and Interpretation.

Authors:  Sharmila Dorbala; Karthik Ananthasubramaniam; Ian S Armstrong; Panithaya Chareonthaitawee; E Gordon DePuey; Andrew J Einstein; Robert J Gropler; Thomas A Holly; John J Mahmarian; Mi-Ae Park; Donna M Polk; Raymond Russell; Piotr J Slomka; Randall C Thompson; R Glenn Wells
Journal:  J Nucl Cardiol       Date:  2018-10       Impact factor: 5.952

Review 4.  Multidetector single-photon emission tomography: are two (or three or four) heads really better than one?

Authors:  J M Links
Journal:  Eur J Nucl Med       Date:  1993-05
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.