Shaken not Stirred: A Pilot Study Testing a Gyroscopic Spoon Stabilization Device in Parkinson's Disease and Tremor.

Lauren E Ryden1, Elie Matar1, Jennifer Y Y Szeto1, Deborah A Hammond1, Paul Clouston1, Simon J G Lewis1.   

Abstract

Entities:  

Year:  2019        PMID: 32606559      PMCID: PMC7313572          DOI: 10.4103/aian.AIAN_251_19

Source DB:  PubMed          Journal:  Ann Indian Acad Neurol        ISSN: 0972-2327            Impact factor:   1.383


× No keyword cloud information.
Dear Sir, As practising clinicians, patients frequently ask for advice regarding non-mainstream treatment modalities for their neurological conditions. With the vast array of information available to patients of varying quality and accuracy, we often forewarn caution for treatments without a robust evidence base. This dilemma has led us to conduct a pilot study assessing a gyroscopic spoon stabilization device in 10 patients with Parkinson's disease (PD) with tremor. There are currently multiple gyroscopic spoon assist devices on the market that advertise tremor cancellation technology to assist patients with tremor in eating tasks. These include the Liftware Steady Spoon and Gyenno Spoon. These devices attempt to stabilize spoon transfers using an embedded gyroscopic mechanism. Gyenno Science kindly donated their device for our testing [Figure 1].
Figure 1

Gyenno Spoon, Gyroscopic Spoon Assist Device

Gyenno Spoon, Gyroscopic Spoon Assist Device Other assistive options for patients with tremor include large grip cutlery,[12] weighted cutlery,[12] weighted lead wrist cuffs,[12] and swivel spoons.[1] These devices have variable individualized efficacy and a paucity of published literature demonstrating objective functional improvement. Pathak et al. have shown the gyroscopic spoon assist device to be effective in a cohort of patients with essential tremor.[3] The gyroscopic device has also been shown to be the patient-preferred device in a head-to-head trial with large grip, weighted cutlery, and swivel spoon for a cohort of PD and essential tremor patients who self-reported difficulty with eating tasks.[1] Our study recruited patients sequentially from the Parkinson's Disease Research Clinic at the Brain and Mind Centre, Sydney. All 10 patients had a confirmed diagnosis of PD[4] and significant levels of tremor as measured by the Movement Disorder Society, Unified Parkinson's Disease Rating Scale (MDS-UPDRS).[5] Patient selection was specifically designed to reflect the market of the device, that being all tremor. No other specific inclusion criteria were required. The authors obtained ethics approval for this study before commencement through the University of Sydney Human Research Ethics Committee. Clinicians consented all participants before commencement of testing on the day. The patients were tested in their on state in the morning or afternoon (depending on their availability) following what they perceived as an adequate response to their usual medication at that time. Unfortunately, we had to exclude the data collected on two patients. One patient had shoulder pain during the testing and testing was ceased. The other patient had tremor predominantly in their non-dominant hand. Table 1 presents patient demographics. Table 2 presents the cohort's tremor phenomenology with average ratings, calculated standard deviations, and median interquartile ranges.
Table 1

Patient Demographics (n=10)

DemographicResult (Standard Deviation)Range
Age, years71.3 (6.8)63-74
Gender (Male), %9 (90%)-
Daily Levodopa Equivalent Dose, mg511.25 (228.4)137.5mg - 987.5mg
Handedness (Right), %10 (100%)-
Dominant Disease Side (Right), %8 (80%)-
Duration of Parkinson’s Disease, years6 (5.3)1-20
Treatment with a dopamine agonist (%)3 (30%)-
Dyskinesia (Number of patients) (%)2 (20%)-
Functional Impact of Dyskinesia0.3 (0.67)0-3
Time Spent with dyskinesia0.2 (0.4)0-1
Motor Fluctuations (%)1 (10%)0-1
Functional Impact of Fluctuations0.1 (0.3)0-1
MDS-UPDRS I score, raw score11.8 (5)1-14
MDS-UPDRS II score, raw score10.18 (6.6)0-23
MDS-UPDRS III score, raw score28.8 (3.05)26-36
MDS-UPDRS IV score, raw score0.8 (1.9)0-6
Hoehn and Yahr stageAll patients rated 2-

MDS-UPDRS I=Non-Motor Experiences of Daily Living, MDS-UPDRS II=Motor Experiences of Daily Living, MDS-UPDRS III=Motor Examination, MDS-UPDRS IV=Motor Complications

Table 2

Tremor Phenotyping of Cohort (n=10)

Tremor ratingAverage Tremor Rating (Standard Deviation)RangeTotal Tremors Rated Across Cohort
3.15a, Postural RUE Tremor0.8 (0.8)0-26
3.15b, Postural LUE Tremor0.5 (0.5)0-15
3.16a Kinetic RUE Tremor0.5 (0.7)0-24
3.16b Kinetic LUE Tremor0.6 (0.5)0-16
3.17a Rest Tremor RUE Tremor1.5 (1.2)0-37
2.17b Rest Tremor LUE Tremor0.4 (0.7)0-23
3.18 Constancy of Rest Tremor1.9 (0.6)1-3-

RUE=Right Upper Extremity, LUE=Left Upper Extremity

Patient Demographics (n=10) MDS-UPDRS I=Non-Motor Experiences of Daily Living, MDS-UPDRS II=Motor Experiences of Daily Living, MDS-UPDRS III=Motor Examination, MDS-UPDRS IV=Motor Complications Tremor Phenotyping of Cohort (n=10) RUE=Right Upper Extremity, LUE=Left Upper Extremity Two timed functional assessment tasks tested the device switched on and off. We blinded the state of the device to the patients during testing. The sixty-second-timed-transfer-task involved transferring rice between two bowls 50 cm apart with the spoon over 60 s. The three-attempt-transfer-task involved a timed three spoon transfer of rice between the two bowls 50 cm apart. The patients repeated the tasks twice; the results presented are averages. Differences between grams transferred and time taken for the tasks with the device switched on and off were compared using a paired t-test [Tables 3 and 4].
Table 3

Three-Attempt-Transfer-Task, Grams Transferred and Time Taken, difference calculated between assist device in on and off state

Participant (P1-P10)

P1P2P3P4P5P6P7P8P9P10
Grams Transferred, Difference between on state and off state01-3-13-90-1-1-3
Time Taken, Difference between on state and off state-2.6-2-2.64.70.5-1.11.65.32.1-2.2

Three-attempt-transfer-task, grams transferred (p 0.200, 95% CI [-3.69, 0.89]). Three-attempt-transfer-task, time taken (p 0.703, 95% CI [-1.757, 2.497])

Table 4

Sixty-Second-Timed-Transfer-Task, Grams Transferred, assist device on versus off state and difference calculated

Participant (P1-P10)

P1P2P3P4P5P6P7P8P9P10
On state11812711296112113180125155131
Off state129141127138105128192150160130
*Difference (On-Off)-11-14-15-427-15-12-25-51

*Sixty-second-timed-transfer-task, (p 0.014, 95% CI [-22.81, - 3.39]). Statistically significant result, less rice has been transferred with the device in the on state compared to the off state

Three-Attempt-Transfer-Task, Grams Transferred and Time Taken, difference calculated between assist device in on and off state Three-attempt-transfer-task, grams transferred (p 0.200, 95% CI [-3.69, 0.89]). Three-attempt-transfer-task, time taken (p 0.703, 95% CI [-1.757, 2.497]) Sixty-Second-Timed-Transfer-Task, Grams Transferred, assist device on versus off state and difference calculated *Sixty-second-timed-transfer-task, (p 0.014, 95% CI [-22.81, - 3.39]). Statistically significant result, less rice has been transferred with the device in the on state compared to the off state The results for the three-attempt-transfer-task did not reach significance (P < 0.05) for the amount of rice transferred in grams or the time taken to achieve the task. Paradoxically, less rice was transferred with the device switched on compared with switched off for the sixty-second-timed-transfer-task (P = 0.0138, 95% confidence interval: −22.81 to −3.39). This result suggests that turning on the gyroscope in the device decreased the efficiency overall. We postulate that the negative and non-statistically significant results obtained are due to the type and severity of tremor affecting the recruited PD cohort. Predominantly rest and low-amplitude postural tremor was noted [Table 2]. Indeed, it seems intuitive that the device would not significantly assist patients with rest tremor. We would, therefore, argue that future research for these devices should focus on testing patients with tremors that are on average more severe and significantly affecting motor activity such as position-independent postural, severe kinetic, and task-specific tremor. This subset of patients may find improved functional outcomes using the device. In addition, patients were not provided any training time with the device before testing. We would recommend further studies into these devices allow for a period of motor learning before formal testing. This pilot study suggests that clinicians should advise caution when patients are considering the purchase of tremor assist devices. Based on the results of this pilot study, we can make no recommendation for the widespread use of this device for patients with PD and tremor. Subgroups may find functional improvement if given enough practice time with the device; however, identifying this group will require further study.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  5 in total

Review 1.  MDS clinical diagnostic criteria for Parkinson's disease.

Authors:  Ronald B Postuma; Daniela Berg; Matthew Stern; Werner Poewe; C Warren Olanow; Wolfgang Oertel; José Obeso; Kenneth Marek; Irene Litvan; Anthony E Lang; Glenda Halliday; Christopher G Goetz; Thomas Gasser; Bruno Dubois; Piu Chan; Bastiaan R Bloem; Charles H Adler; Günther Deuschl
Journal:  Mov Disord       Date:  2015-10       Impact factor: 10.338

2.  Aids for activities of daily living in people with Parkinson's disease.

Authors:  Amelia Jackson
Journal:  Br J Community Nurs       Date:  2019-05-02

3.  Adapted Feeding Utensils for People With Parkinson's-Related or Essential Tremor.

Authors:  Joyce Sabari; Dimitre G Stefanov; Judy Chan; Lorraine Goed; Joyce Starr
Journal:  Am J Occup Ther       Date:  2019 Mar/Apr

4.  Movement Disorder Society-sponsored revision of the Unified Parkinson's Disease Rating Scale (MDS-UPDRS): scale presentation and clinimetric testing results.

Authors:  Christopher G Goetz; Barbara C Tilley; Stephanie R Shaftman; Glenn T Stebbins; Stanley Fahn; Pablo Martinez-Martin; Werner Poewe; Cristina Sampaio; Matthew B Stern; Richard Dodel; Bruno Dubois; Robert Holloway; Joseph Jankovic; Jaime Kulisevsky; Anthony E Lang; Andrew Lees; Sue Leurgans; Peter A LeWitt; David Nyenhuis; C Warren Olanow; Olivier Rascol; Anette Schrag; Jeanne A Teresi; Jacobus J van Hilten; Nancy LaPelle
Journal:  Mov Disord       Date:  2008-11-15       Impact factor: 10.338

5.  A noninvasive handheld assistive device to accommodate essential tremor: a pilot study.

Authors:  Anupam Pathak; John A Redmond; Michael Allen; Kelvin L Chou
Journal:  Mov Disord       Date:  2013-12-27       Impact factor: 10.338