Jia Luo1, Jiaqi Liu1, Xiaoyan Guo1,2, Yuejiao Liu1, Haibo Jin1,2, Daidi Fan3.
Abstract
Barium sulfate (BaSO4) is a material with high reflectance for preparing bifunctional sheets used in dry reagent chemical tests. In this study, bifunctional BaSO4 sheets with scattering power and high reflectance were prepared with BaSO4 microspheres sized 1.3~1.8 μm and cellulose acetate (CA). Factors such as the BaSO4 morphology, CA dosage, mixing time, surfactant, solid content and wet sheet thickness were investigated. Scanning electron microscopy (SEM), a dynamic contact angle test, light reflection detector and light reflection densitometer were employed to characterize the structure and properties of bifunctional BaSO4 sheets. The optimal conditions for preparing bifunctional BaSO4 sheets under natural drying conditions were as follows: mass ratio of CA to BaSO4 microspheres was 0.03:1; mixing acetone solution was used to form liquid stock puree; 35 μL Tween-80 was used to improve the hydrophilicity of bifunctional sheets; solid content was 54.9%; mixing 2 h then 450 µm thick sheets were cast on a glass plate using a film applicator. The light reflectance value of bifunctional BaSO4 sheets in the range of 400~800 nm was higher than 97%. Serum diffused in bifunctional BaSO4 sheets reacted in reagent sheets and formed uniform colorful spots. Considering the repeatability of spot proportion and light reflectance value, the sheet offered a uniform serum scattering power and good repeatability. Therefore, the bifunctional BaSO4 sheets are regarded as a promising material for dry chemical diagnostic reagents.Entities:
Keywords: barium sulfate; bifunctional sheet; cellulose acetate; dry chemical diagnosis reagent
Year: 2020 PMID: 32605245 PMCID: PMC7372409 DOI: 10.3390/ma13132903
Source DB: PubMed Journal: Materials (Basel) ISSN: 1996-1944 Impact factor: 3.623
Figure 1Particle size distribution of BaSO4 micro-particle.
Figure 2Scanning electron microscope (SEM) micrographs of BaSO4 bifunctional sheets with different BaSO4 morphology: (A) BaSO4 microspheres; (B) amorphous BaSO4.
Figure 3SEM micrographs of bifunctional BaSO4 sheets with different mass ratio of cellulose acetate (CA) to BaSO4: (a) 0.02:1; (b) 0.03:1; (c) 0.04:1; (d) 0.05:1.
Figure 4BaSO4 bifunctional sheets prepared with different mixing times: (a) 0.5 h; (b) 1 h; (c) 1.5 h; (d) 2 h.
Figure 5The contrast diagram of effect of surfactant on serum scattering power: (A) without surfactant; (B) with 35 μL surfactant Tween-80.
Figure 6Contact angle of serum on the surface of scattering power sheets: (A) without surfactant; (B) with 35 μL surfactant Tween-80.
Effect of solid content on the surface and flatness of bifunctional sheets.
| Acetone Volume (mL) | Solid Content (%) | The Surface and Flatness of Bifunctional Sheet |
|---|---|---|
| 8 | 60.1 | uneven |
| 10 | 54.9 | uniform and flat |
| 12 | 50.5 | thin and porous |
Relationship between the thickness of wet sheet and scattering power.
| Sheets Code | Wet Sheet Thickness (µm) | Scattering Time (s) | Scattering Diameter (cm) |
|---|---|---|---|
| 1 | 200 | 27 | 2.3 |
| 2 | 250 | 20 | 2.0 |
| 3 | 300 | 17 | 1.9 |
| 4 | 350 | 14 | 1.6 |
| 5 | 400 | 12 | 1.3 |
| 6 | 450 | 11 | 1.2 |
| 7 | 500 | 11 | 1.2 |
Figure 7Light reflectivity of BaSO4 bifunctional sheet and TiO2 bifunctional sheet.
Figure 8Spot images of serum reaction.
Light reflection signal values of dry chemical diagnostic reagents in vitro.
| Serial | Single Number (V) | Mean Single Value | Deviation |
|---|---|---|---|
| 1 | 0.3295 | 0.3250 | 0.0045 |
| 2 | 0.3259 | 0.0009 | |
| 3 | 0.3238 | −0.0012 | |
| 4 | 0.3226 | −0.0024 | |
| 5 | 0.3230 | −0.0020 |
Figure 9Standard curves of glucose (GLU)-TiO2/CA and GLU-BaSO4/CA detection chips at different GLU concentrations.
Repeatability of TiO2/CA and BaSO4/CA detection chips.
| Chip Name | Concentration (mmol·L−1) | SD | CV (%) | |||||
|---|---|---|---|---|---|---|---|---|
| GLU-TiO2/CA | 8.44 | 8.81 | 8.68 | 8.19 | 8.27 | 8.47 | 0.24 | 2.8 |
| GLU-BaSO4/CA | 8.23 | 8.19 | 8.61 | 8.12 | 8.54 | 8.46 | 0.20 | 2.4 |
Accuracy of BaSO4/CA detection chip.
| Chip Name | Concentration (mmol·L−1) | Target Value (mmol·L−1) | Range of Target Values (mmol·L−1) | ||
|---|---|---|---|---|---|
| GLU-BaSO4/CA | 1.58 | 1.75 | 1.72 | 1.67 | 1.50~1.84 |