Bairen Du1, Jingjing Zhang2, Yanjie Dong3, Junwei Wang3, Longwen Lei3, Rengfei Shi4.
Abstract
A novel method for determining the testosterone/epitestosterone concentration ratio in human urine was established by capillary electrophoresis with diode-array detector. The urine samples were firstly purified by the solid extraction. The optimal experimental conditions were: running buffer pH = 4.74, 15.0 mmol L-1 HAc-NaAc, separation voltage 25 kV, temperature 25 °C, sample injection pressure 3.43 × 103 Pa, and duration 10 s. The testosterone and epitestosterone linear range were determined as 8.0-960.0 ng mL-1, respectively. The testosterone and epitestosterone detection limits were determined as 4.6 and 4.5 ng mL-1, respectively. The relative standard deviation was less than 0.36%.Entities:
Keywords: Capillary electrophoresis; Determination; Epitestosterone; Testosterone; Urine
Mesh:
Substances:
Year: 2020 PMID: 32603755 DOI: 10.1016/j.steroids.2020.108691
Source DB: PubMed Journal: Steroids ISSN: 0039-128X Impact factor: 2.668