| Literature DB >> 32599732 |
Dominika Krenczkowska1, Krystyna Mojsiewicz-Pieńkowska1, Bartosz Wielgomas2, Dagmara Bazar1, Zbigniew Jankowski3.
Abstract
Cyclic methylsiloxanes D4, D5, D6 (also called cyclic silicones) are widely used in various dermatological products and cosmetics, both for children and adults. As a result of their unique physicochemical properties, the production of cyclic methylsiloxanes has greatly increased over the last few years, which has resulted in increased exposure to mankind. The validated quantitative for gas chromatography-flame ionization detector (GC-FID) analysis with using the transdermal diffusion system with vertical Franz cells demonstrated that ex vivo human skin is not a barrier to cyclic siloxanes. D4, D5, and D6 have a specific affinity to stratum corneum (SC) (especially D6), and can even diffuse into the deeper layers of the skin (epidermis (E) and dermis (D)), or into the receptor fluid as well. An important achievement of this work was the observation of the characteristic ratio partitioning D4, D5, and D6 in skin layers and receptor fluid (RF). The studies have shown that, in order to thoroughly understand the mechanism, it is important to determine not only the differences in the amounts of cumulated doses in total in all skin layers and receptor fluid, but also the mutual ratios of analyte concentrations existing between matrices. For example, in the case of the stratum corneum, the cumulative doses of D4, D5, and D6 were 27.5, 63.9, and 67.2 µg/cm2/24 h, respectively, and in the epidermis, they were 6.9, 29.9, and 10.7 µg/cm2/24 h, respectively, which confirmed the highest affinity of D6 to stratum corneum as the amount diffused into the epidermis was 2.8 times smaller compared to D5. The calculated epidermis-to-stratum corneum ratios of analyte concentrations also confirm this. The largest ratio was identified for D5 (E/SC = 47), followed by D4 (E/SC = 25), and finally by D6 (E/SC = 16). The analysis of the next stage of diffusion from epidermis to dermis revealed that in dermis the highest cumulative dose was observed for D5 (13.9 µg/cm2/24 h), while the doses of D4 and D6 were similar (5.1 and 5.3 µg/cm2/24 h). Considering the concentration gradient, it can be concluded that the diffusion of D5 and D6 occurs at a similar level, while D4 diffuses at a much higher level. These observations were also confirmed by the dermis-to-epidermis concentration ratios. The final stage of diffusion from dermis to the receptor fluid indicated that D4 was able to permeate easily, while D5 exhibited a difficult diffusion and the diffusion of D6 was limited. The receptor fluid-to-dermis concentration ratios (RF/D) were calculated for D4, D5, and D6: 80, 53, and 17, respectively. Our results also revealed the increased risk of D4 and D5 absorption into the blood and lymphatic systems, whereas D6 demonstrated the lowest risk. Therefore, we can argue that, among the three tested compounds, D6 is the safest one that can be used in dermatological, cosmetic, and personal care products. This study demonstrates that the stratum corneum, epidermis, and dermis can be also considered reservoirs of cyclic methylsiloxanes. Therefore, these compounds can demonstrate potential long-term bioaccumulation, and can be absorbed to the bloodstream in a long-term and uncontrolled process.Entities:
Keywords: cyclic siloxanes; gas chromatography-flame ionization detector (GC-FID); human skin diffusion; percutaneous absorption; silicones; transdermal diffusion Franz cell
Year: 2020 PMID: 32599732 PMCID: PMC7355424 DOI: 10.3390/pharmaceutics12060586
Source DB: PubMed Journal: Pharmaceutics ISSN: 1999-4923 Impact factor: 6.321
Figure 1The methodology of sample preparation for gas chromatography-flame ionization detector (GC-FID) analysis.
Selected recommendations and acceptance criteria for the bioanalytical and analytical methods in accordance with the United States Food and Drug Administration (FDA), European Medicines Agency (EMA), and International Conference on Harmonization (ICH) guidelines [53,54,55,56,57,58].
| Studies | Validation Parameter | Acceptance Criterion |
|---|---|---|
|
| Specificity | Blank samples of the appropriate biological matrix should be obtained from at least six individual sources. The response of the interfering compound should be less than 20% of the lowest concentration of the standard curve lower limit of quantification (LLOQ) and 5% standard response internal standard (IS). |
| Precision/repeatability/intermediate precision for retention time | Lack of the acceptance criterion in guidelines. Adopted ≤ 2.5%. | |
|
| Range | Analogically to linearity |
| Linearity | The lowest concentration of analyte from the calibration curve LLOQ ≥ 5 times higher than the blank sample, with a precision of up to 20% of the coefficient of variation (CV) and accuracy within ± 20% (80–120%); upper limit, the highest concentration of the standard curve upper limit of quantification (ULOQ) with a precision below 15% CV and accuracy within ± 15% (85–115%); the correlation coefficient (r) should be greater than 0.99 and the coefficient of determination be (r2) 0.98. Nonzero calibrators should be ± 15% of the nominal (theoretical) concentrations (NC), except at LLOQ where the calibrator should be ± 20% of the nominal concentrations in each validation run. In addition, 75% and a minimum of six nonzero calibrator levels should meet the above criteria in each validation run. | |
| Sensitivity | The lowest nonzero standard on the calibration curve defines the sensitivity (LLOQ). | |
| LOQ | A 10:1 signal-to-noise ratio. The quantitation limit is generally determined by the analysis of samples with known concentrations of analyte and by establishing the minimum level at which the analyte can be quantified with an acceptable accuracy and precision. | |
| LOD | A 3:1 signal-to-noise ratio. | |
| Precision/Repeatability | Precision should be established with at least 3 independent runs, 4 QC levels per run (LLOQ, L, M, and H QC), and ≥ 5 replicates per QC level. CV in the biological matrix should not exceed 15% and close to 20% of the LLOQ value. | |
| Accuracy/Trueness | Accuracy should be established with at least 3 independent runs, 4 QC levels per run (LLOQ, L, M, and H QC), and ≥ 5 replicates per QC level. The relative error should not exceed ± 15% of the actual value, except for the lowest concentration of the LLOQ analyte close to ± 20% of the value. | |
| QCs | Four QCs, including LLOQ, low (L: defined as three times the LLOQ), mid (M: defined as mid-range), and high (H: defined as high-range) from at least 5 replicates in at least 3 runs. | |
| Recovery (extraction efficiency) | Extracted samples at low quality control (LOQ), medium quality control (MOQ), high quality control (HQC) concentrations versus extracts of blanks spiked with the analyte postextraction (maximum of 3 times the LLOQ and close to the HLOQ). | |
| Matrix effect | Extracted samples including matrix at L, M, and H QC concentrations versus spiked solvent at L, M, and H QC concentrations used for the extraction. |
Figure 2(A) separation of chromatographic peaks of the analyzed compounds D4, D5, and D6 and internal standard (DNB, 5 µg/mL); (B) example of matrix sample (stratum corneum).
Figure 3Representative chromatograms of the extracts obtained from the fractionated layers of the skin (stratum corneum, epidermis, and dermis) and from the receptor fluid containing octamethylcyclotetrasiloxane (D4), decamethylcyclopentasiloxane (D5), dodecamethylcyclohexasiloxane (D6), respectively; the internal standard (1,4-dinitrobenzene (DNB) was added at a concentration of 5 µg/mL.
Validation results (intra-day precision, inter-day precision, and accuracy) of the analytical method for the determination of D4, D5, and D6 (n = 6/18).
| Cyclic Siloxane | Nominal Concertation NC [µg/mL] | Concentration Found CF (Mean) [µg/mL] | Intra-Day Precision CV [%] n = 6 | Inter-Day Precision CV [%] n = 18 | Accuracy [%] n = 6 | |
|---|---|---|---|---|---|---|
| D4 | LQC | 1 | 1.0 | 2.4 | 3.6 | 102.8 |
| MQC | 5 | 5.0 | 1.5 | 2.2 | 100.4 | |
| HQC | 25 | 24.8 | 0.6 | 1.1 | 99.3 | |
| D5 | LQC | 1 | 1.0 | 6.6 | 4.7 | 106.1 |
| MQC | 5 | 5.1 | 1.7 | 1.9 | 100.44 | |
| HQC | 25 | 25.3 | 0.4 | 1.1 | 100.7 | |
| D6 | LQC | 1 | 1.0 | 2.0 | 6.2 | 105.7 |
| MQC | 5 | 5.1 | 3.5 | 2.9 | 105.6 | |
| HQC | 25 | 24.2 | 6.5 | 5.7 | 99.7 | |
Matrix effect coefficient for D4, D5, and D6.
| Siloxane | Heading | Matrix Effect | |||
|---|---|---|---|---|---|
| Nominal Concentration [µg/mL] | Stratum Corneum | Epidermis | Dermis | Receptor Fluid | |
| D4 | 1 | 1.03 | 1.01 | 0.99 | 0.93 |
| 5 | 0.92 | 1.07 | 0.95 | 1.03 | |
| 25 | 0.98 | 1.03 | 1.10 | 1.04 | |
| D5 | 1 | 0.98 | 0.93 | 0.97 | 0.93 |
| 5 | 0.94 | 0.98 | 0.93 | 0.98 | |
| 25 | 0.94 | 0.97 | 0.99 | 0.99 | |
| D6 | 1 | 0.96 | 0.91 | 0.92 | 0.97 |
| 5 | 0.89 | 0.91 | 0.94 | 0.94 | |
| 25 | 0.94 | 0.96 | 0.97 | 1.07 | |
Results of quality control and recovery (extraction efficiency) of D4, D5, and D6 from the stratum corneum, epidermis, dermis, and receptor fluid, n = 6.
| Cyclic Siloxane | Heading | Sample | Stratum Corneum | Epidermis | Dermis | Receptor Fluid | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| NC [µg/mL] | CF [µg/mL] | CV | Accuracy | CF [µg/mL] | CV | Accuracy | CF [µg/mL] | CV | Accuracy | CF [µg/mL] | CV | Accuracy | ||
| (%) | (%) | (%) | (%) | (%) | (%) | (%) | (%) | |||||||
| D4 | LQC | 1 | 1.1 | 6.7 | 105.5 | 1.0 | 9.0 | 102.6 | 1.0 | 4.6 | 100.8 | 0.9 | 2.8 | 95.3 |
| MQC | 5 | 4.8 | 4.6 | 95.2 | 5.0 | 9.3 | 100.1 | 4.8 | 4.3 | 95.8 | 5.2 | 4.6 | 103.3 | |
| HQC | 25 | 24.3 | 5.3 | 97.1 | 25.4 | 9.9 | 101.6 | 25.5 | 5.2 | 102.4 | 25.6 | 3.5 | 102.2 | |
| D5 | LQC | 1 | 1.0 | 6.8 | 104.5 | 1.0 | 5.7 | 99.9 | 1.0 | 5.8 | 103.4 | 0.9 | 3.3 | 99.5 |
| MQC | 5 | 4.7 | 7.3 | 94.7 | 5,0 | 10.6 | 99.2 | 4.7 | 5.5 | 93.9 | 4.9 | 5.3 | 98.7 | |
| HQC | 25 | 23.7 | 5.9 | 94.9 | 24.7 | 11.3 | 98.6 | 25.1 | 6.3 | 100.5 | 25.3 | 9.3 | 101.0 | |
| D6 | LQC | 1 | 1.0 | 5.1 | 101.6 | 1,0 | 5.8 | 97.2 | 1.0 | 5.7 | 98.1 | 1.0 | 4.9 | 103.0 |
| MQC | 5 | 4.6 | 9.6 | 92.2 | 4.7 | 7.4 | 94.2 | 4.9 | 6.6 | 97.4 | 4.9 | 7.9 | 97.6 | |
| HQC | 25 | 24.1 | 7.4 | 96.2 | 25.6 | 9.2 | 97.4 | 24.7 | 5.9 | 98.8 | 25.7 | 6.5 | 102.7 | |
The cumulative doses (CD) of D4, D5, and D6 in various layers of the skin and in receptor fluid; total amount after diffusion as well as concentration in extracts (CE); calculated epidermis-to-stratum corneum, dermis-to-epidermis, dermis-to-receptor fluid ratios. mean obtained from n = 7.
| Sample | D4 | D5 | D6 | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| CD [µg/cm2/24 h] | CE [µg/mL] | CV [%] | Ratio [%] | CD [µg/cm2/24 h] | CE [µg/mL] | CV [%] | Ratio [%] | CD [µg/cm2/24 h] | CE [µg/mL] | CV [%] | Ratio [%] | |
| Stratum corneum | 27.5 | 3.5 | 4.9 | 25 | 63.9 | 8.1 | 5.5 | 47 | 67.2 | 8.6 | 20.3 | 16 |
| Epidermis | 6.9 | 4.4 | 21.4 | 29.9 | 19.1 | 9.4 | 10.7 | 6.8 | 6.2 | |||
| 74 | 46 | 50 | ||||||||||
| Dermis | 5.1 | 3.2 | 24.5 | 13.9 | 8.8 | 13.5 | 5.3 | 3.4 | 7.4 | |||
| 80 | 53 | 17 | ||||||||||
| Receptor fluid | 4.1 | 2.6 | 27.4 | 7.4 | 4.7 | 24.2 | 0.9 | 0.6 | 29.5 | |||
| TOTAL | 43.6 | 13.7 | 8.8 | 114.1 | 40.7 | 2.6 | 84.1 | 19.4 | 16.3 | |||
Figure 4The mechanism of the overcoming human skin’s barrier system by various cyclic siloxanes.