The Loopamp™ MTBC kit (TB-LAMP) is recommended by WHO for Mycobacterium tuberculosis complex detection in low-income countries with a still low drug-resistant tuberculosis (TB) rate. This study is aimed at testing its feasibility in Cambodia on sputa collected from presumptive tuberculosis patients. 499 samples were tested at a smear microscopy center and 200 at a central-level mycobacteriology laboratory. Using mycobacterial cultures as reference, TB-LAMP results were compared with those of LED fluorescent microscopy (LED-FM) and Xpert® MTB/RIF. At the microscopy center, TB-LAMP sensitivity was higher than that of LED-FM (81.5% [95% CI, 74.5-87.6] versus 69.4% [95% CI, 62.2-76.6]), but lower than that of the Xpert assay (95.5% [95% CI 92.3-98.8]). At the central-level laboratory, TB-LAMP sensitivity (92.8% [95% CI, 87.6-97.9]) was comparable to that of Xpert (90.7% [95% CI, 85.0-96.5]) using stored sample. No significant difference in terms of specificity between TB-LAMP and Xpert assays was observed in both study sites. In conclusion, our data demonstrate that TB-LAMP could be implemented at microscopy centers in Cambodia to detect TB patients. In addition, TB-LAMP can be a better choice to replace smear microscopy for rapid TB diagnosis of new presumptive TB patients, in settings with relative low prevalence of drug-resistant TB and difficulties to implement Xpert assay.Entities:
Year: 2020 PMID: 32596349 PMCID: PMC7288195 DOI: 10.1155/2020/6828043
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Study flowchart in smear microscopy center.
Demographic and clinical features of the study population and culture-confirmed tuberculosis cases in smear microscopy center study.
| Characteristics | Study population | Culture-confirmed TB cases | |
|---|---|---|---|
| Age (year) | Total | 499 | 157 |
| 18-24 | 36 (7.2) | 16 (10.2) | |
| 25-34 | 83 (16.6) | 31 (19.7) | |
| 35-44 | 67 (13.4) | 26 (16.6) | |
| 45-54 | 94 (18.8) | 33 (21.0) | |
| 55-64 | 98 (19.6) | 26 (16.6) | |
| ≥ 65 | 121 (24.2) | 25 (15.9) | |
|
| |||
| Sex | Total | 499 | 157 |
| Men | 280 (56.1) | 103 (65.6) | |
| Women | 219 (43.9) | 54 (34.4) | |
|
| |||
| Symptoms | Total | 499 | 157 |
| Fever | 497 (99.6) | 156 (98.1) | |
| Cough | 487 (97.6) | 154 (96.9) | |
| Chest discomfort | 463 (92.8) | 143 (89.9) | |
| Weight loss | 453 (90.8) | 150 (94.3) | |
| Breathlessness | 98 (19.6) | 35 (22.0) | |
| Hemoptysis | 63 (12.6) | 23 (14.5) | |
|
| |||
| Abnormal chest X-ray | 421/426 (98.8) | 141/141 (100) | |
|
| |||
| Previous treatment for TB disease | 129/482 (26.8) | 18/154 (11.7) | |
|
| |||
| TB contact | 74/496 (14.8) | 13/156 (8.3) | |
Sensitivity and specificity of LED-fluorescent microscopy, TB-LAMP, and Xpert MTB/RIF.
| Test | Culture result | Sensitivity among culture-positive MTB | Specificity | |||
|---|---|---|---|---|---|---|
| MTB-positive | MTB-negative | Smear-positive | Smear-negative | Overall | ||
| Smear microscopy center | ||||||
| Microscopy | 109/48 | 6/336 | — | — | 69.4 | 98.2 |
| TB-LAMP | 128/29 | 9/333 | 94.5 | 52.1 | 81.5 | 97.4 |
| Xpert | 150/7 | 11/331 | 99.1 | 87.5 | 95.5 | 96.8 |
| Central laboratory | ||||||
| TB-LAMP | 90/7 | 4/95 | 100 | 56.3 | 92.8 | 96.0 |
| Xpert | 88/9 | 3/96 | 100 | 43.8 | 90.7 | 97.0 |
Figure 2Visual detection of LAMP products under UV light. The combination of turbidity and bright green fluorescence observed in tube 2 and positive control (PC) indicated positive reactions, while a blue color without turbidity indicated a negative reaction in tubes 1, 3, and negative control (NC).
Matched comparison between TB-LAMP performed at the smear microscopy center and the Xpert and LED-fluorescent microscopy.
| Comparison | Difference | Lower 95% confidence limit | Upper 95% confidence limit |
|---|---|---|---|
| Sensitivity | |||
| TB-LAMP versus Xpert | 14.0 | 8.3 | 19.7 |
| TB-LAMP versus microscopy | 12.1 | 5.4 | 18.8 |
| Specificity | |||
| TB-LAMP versus Xpert | -0.6 | -2.0 | 0.8 |
| TB-LAMP versus microscopy | 0.9 | -1.2 | 2.9 |