Accurate Fiducial Point Detection Using Haar Wavelet for Beat-by-Beat Blood Pressure Estimation.

Muskan Singla1, Syed Azeemuddin1, Prasad Sistla2.   

Abstract

Pulse Arrival Time (PAT) derived from Electrocardiogram (ECG) and Photoplethysmogram (PPG) for cuff-less Blood Pressure (BP) measurement has been a contemporary and widely accepted technique. However, the features extracted for it are conventionally from an isolated pulse of ECG and PPG signals. As a result, the estimated BP is intermittent.
OBJECTIVE: This paper presents feature extraction from each beat of ECG and PPG signals to make BP measurements uninterrupted. These features are extracted by employing Haar transformation to adaptively attenuate measurement noise and improve the fiducial point detection precision.
METHOD: the use of only PAT feature as an independent variable leads to an inaccurate estimation of either Systolic Blood Pressure (SBP) or Diastolic Blood Pressure (DBP) or both. We propose the extraction of supplementary features that are highly correlated to physiological parameters. Concurrent data was collected as per the Association for the Advancement of Medical Instrumentation (AAMI) guidelines from 171 human subjects belonging to diverse age groups. An Adaptive Window Wavelet Transformation (AWWT) technique based on Haar wavelet transformation has been introduced to segregate pulses. Further, an algorithm based on log-linear regression analysis is developed to process extracted features from each beat to calculate BP.
RESULTS: The mean error of 0.43 and 0.20 mmHg, mean absolute error of 4.6 and 2.3 mmHg, and Standard deviation of 6.13 and 3.06 mmHg is achieved for SBP and DBP respectively.
CONCLUSIONS: The features extracted are highly precise and evaluated BP values are as per the AAMI standards. Clinical Impact: This continuous real-time BP monitoring technique can be useful in the treatment of hypertensive and potential-hypertensive subjects.

Entities:  

Keywords:  Adaptive algorithm; blood pressure; detection algorithm; electrocardiography; feature extraction; photoplethysmography; regression analysis; signal processing; wavelet transforms

Year:  2020        PMID: 32596063      PMCID: PMC7316202          DOI: 10.1109/JTEHM.2020.3000327

Source DB:  PubMed          Journal:  IEEE J Transl Eng Health Med        ISSN: 2168-2372            Impact factor:   3.316


  17 in total

Review 1.  Arterial and cardiac aging: major shareholders in cardiovascular disease enterprises: Part I: aging arteries: a "set up" for vascular disease.

Authors:  Edward G Lakatta; Daniel Levy
Journal:  Circulation       Date:  2003-01-07       Impact factor: 29.690

2.  Continuous and noninvasive measurement of systolic and diastolic blood pressure by one mathematical model with the same model parameters and two separate pulse wave velocities.

Authors:  Yan Chen; Changyun Wen; Guocai Tao; Min Bi
Journal:  Ann Biomed Eng       Date:  2011-11-19       Impact factor: 3.934

3.  Photoplethysmogram intensity ratio: A potential indicator for improving the accuracy of PTT-based cuffless blood pressure estimation.

Authors:  Xiao-Rong Ding; Yuan-Ting Zhang
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2015

4.  Noninvasive Blood Pressure Estimation Using Ultrasound and Simple Finite Element Models.

Authors:  Aaron M Zakrzewski; Brian W Anthony
Journal:  IEEE Trans Biomed Eng       Date:  2017-06-12       Impact factor: 4.538

5.  Cuff-Less Blood Pressure Estimation Using Pulse Waveform Analysis and Pulse Arrival Time.

Authors:  Young-Zoon Yoon; Jae Min Kang; Yongjoo Kwon; Sangyun Park; Seungwoo Noh; Younho Kim; Jongae Park; Sung Woo Hwang
Journal:  IEEE J Biomed Health Inform       Date:  2017-06-12       Impact factor: 5.772

6.  A low-complexity ECG feature extraction algorithm for mobile healthcare applications.

Authors:  Evangelos B Mazomenos; Dwaipayan Biswas; Amit Acharyya; Taihai Chen; Koushik Maharatna; James Rosengarten; John Morgan; Nick Curzen
Journal:  IEEE J Biomed Health Inform       Date:  2013-01-25       Impact factor: 5.772

7.  Cuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring.

Authors:  Mohammad Kachuee; Mohammad Mahdi Kiani; Hoda Mohammadzade; Mahdi Shabany
Journal:  IEEE Trans Biomed Eng       Date:  2016-06-14       Impact factor: 4.538

8.  Design of a Continuous Blood Pressure Measurement System Based on Pulse Wave and ECG Signals.

Authors:  Jian-Qiang Li; Rui Li; Zhuang-Zhuang Chen; Gen-Qiang Deng; Huihui Wang; Constandinos X Mavromoustakis; Houbing Song; Zhong Ming
Journal:  IEEE J Transl Eng Health Med       Date:  2018-01-18       Impact factor: 3.316

9.  Worldwide trends in blood pressure from 1975 to 2015: a pooled analysis of 1479 population-based measurement studies with 19·1 million participants.

Authors: 
Journal:  Lancet       Date:  2016-11-16       Impact factor: 79.321

10.  A Novel Neural Network Model for Blood Pressure Estimation Using Photoplethesmography without Electrocardiogram.

Authors:  Ludi Wang; Wei Zhou; Ying Xing; Xiaoguang Zhou
Journal:  J Healthc Eng       Date:  2018-03-07       Impact factor: 2.682

View more
  2 in total

1.  Boosted-SpringDTW for Comprehensive Feature Extraction of PPG Signals.

Authors:  Jonathan Martinez; Kaan Sel; Bobak J Mortazavi; Roozbeh Jafari
Journal:  IEEE Open J Eng Med Biol       Date:  2022-05-12

2.  Design and Implementation of a Wireless Medical Robot for Communication Within Hazardous Environments.

Authors:  Nashat Maher; G A Elsheikh; A N Ouda; W R Anis; Tamer Emara
Journal:  Wirel Pers Commun       Date:  2021-08-26       Impact factor: 1.671

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.