Literature DB >> 32592881

A pilot study of the impact of facial skin protectants on qualitative fit testing of N95 masks.

Ai-Tram N Bui1, Zizi Yu1, Karen Lee2, Sara J Li2, William G Tsiaras3, Sherry H Yu2, Nicole R LeBoeuf4, Arash Mostaghimi5.   

Abstract

Mesh:

Year:  2020        PMID: 32592881      PMCID: PMC7313513          DOI: 10.1016/j.jaad.2020.06.069

Source DB:  PubMed          Journal:  J Am Acad Dermatol        ISSN: 0190-9622            Impact factor:   11.527


× No keyword cloud information.
To the Editor: The COVID-19 pandemic has necessitated prolonged use of N95 masks, leading to superficial wounds, purpura, and indentations on health care workers' faces. , The use of skin protectants may prevent skin irritation caused by N95 masks by providing a barrier and/or redistributing pressure; however, the impact on respirator fit has not been evaluated. This study assesses the impact of the use of skin protectants on N95 respirator qualitative fit test (QLFT) results and user comfort. We enrolled adult employees at Brigham Health and Dana-Farber Cancer Institute previously fit-tested for N95 masks (N = 25) via a standardized QLFT protocol (see Supplemental Materials; available via Mendeley at https://doi.org/10.17632/sj6tr3mp9r.1). Each participant underwent QLFT for 5 types of skin protectants on a 3M (St Paul, MN) 1860 N95 mask after self-application using a standardized protocol (Fig 1 and Supplemental Table I; available via Mendeley at https://doi.org/10.17632/sj6tr3mp9r.1). Participants underwent repeated QLFT of their respirator for each dressing and rated dressing comfort (Supplemental Fig 1; available via Mendeley at https://doi.org/10.17632/sj6tr3mp9r.1).
Fig 1

Standardized placement of facial protectants before donning N95 Masks (created with BioRender.com). DuoDERM (ConvaTec, Oklahoma City, OK) CGF, DuoDERM Extra Thin, and Mepitac Soft Silicone Tape (Mölnlycke, Gothenburg, Sweden) were precut into 3 uniform 2 × 5–cm dressings. Hydrocolloid bandages used were a combination of Band-Aid brand (Johnson & Johnson, New Brunswick, NJ) and readily available generic bandages, which came in a 2 × 5–cm rectangle size and a 3 × 5–cm oval size. For DuoDERM Control Gel Formula (CGF), DuoDERM Extra Thin, and Mepitac Soft Silicone Tape, participants were instructed to first place 1 dressing on the nasal bridge (starting centrally and moving laterally) and then 2 on the bilateral cheeks. For the hydrocolloid bandages, participants were instructed to apply 1 oval hydrocolloid bandage on the nasal bridge and two 2 × 5–cm rectangle bandages on the bilateral cheeks. For the Cavilon film, participants were instructed to apply the liquid protectant on the same 3 areas of the face. Cavilon film was tested last on each participant because it leaves a residue. Participants were instructed on removal, including lifting the bandages laterally to reduce discomfort and removing the Cavilon film with a facial cleanser or alcohol wipe.

Standardized placement of facial protectants before donning N95 Masks (created with BioRender.com). DuoDERM (ConvaTec, Oklahoma City, OK) CGF, DuoDERM Extra Thin, and Mepitac Soft Silicone Tape (Mölnlycke, Gothenburg, Sweden) were precut into 3 uniform 2 × 5–cm dressings. Hydrocolloid bandages used were a combination of Band-Aid brand (Johnson & Johnson, New Brunswick, NJ) and readily available generic bandages, which came in a 2 × 5–cm rectangle size and a 3 × 5–cm oval size. For DuoDERM Control Gel Formula (CGF), DuoDERM Extra Thin, and Mepitac Soft Silicone Tape, participants were instructed to first place 1 dressing on the nasal bridge (starting centrally and moving laterally) and then 2 on the bilateral cheeks. For the hydrocolloid bandages, participants were instructed to apply 1 oval hydrocolloid bandage on the nasal bridge and two 2 × 5–cm rectangle bandages on the bilateral cheeks. For the Cavilon film, participants were instructed to apply the liquid protectant on the same 3 areas of the face. Cavilon film was tested last on each participant because it leaves a residue. Participants were instructed on removal, including lifting the bandages laterally to reduce discomfort and removing the Cavilon film with a facial cleanser or alcohol wipe. Most participants were female (76%), with an average age of 28 years. QLFT passing rates ranged from 88.0% for Cavilon film (3M) to 56.0% for DuoDERM CGF (ConvaTec, Oklahoma City, OK), with the highest failure rates noted with movement maneuvers (Tables I and II ). Overall, 9 (36.0%) participants passed with all 5 materials. Mepitac tape (Mölnlycke, Gothenburg, Sweden) and DuoDERM CGF (88.0% positive rating) were reported to be more comfortable than Cavilon film (22.0%). Cavilon film and DuoDERM CGF had the most negative qualitative comments, with odor and impact on mask fit or seal quality as common concerns, respectively.
Table I

Study outcomes: Participant demographics/characteristics

CharacteristicsValue
Sex, n (%)
 Male6 (24)
 Female19 (76)
Race, n (%)
 White11 (44)
 Asian/Pacific Islander8 (32)
 Hispanic or Latino3 (12)
 Black2 (8)
 Other1 (4)
1860 mask size, n (%)
 Small16 (64)
 Regular9 (36)
Age, y
 Mean28
 Median27
BMI range, kg/m2, n (%)
 18.5-24.911 (44)
 25-29.96 (24)
 ≥308 (32)

BMI, Body mass index.

Table II

Outcome measures by dressing type, n (%)∗

Outcome measuresCavilonHydrocolloid bandagesDuoDERM Extra ThinMepitac Soft Silicone Tape§DuoDERM Control Gel Formula
Qualitative mask fit test
 Passed all components22 (88)21 (84)18 (72)16 (64)14 (56)
 Stage failed
 Seal check0 (0)0 (0)0 (0)0 (0)0 (0)
 Regular breathing0 (0)0 (0)0 (0)0 (0)0 (0)
 Deep breaths1 (4)0 (0)1 (4)0 (0)2 (8)
 Head side to side2 (8)0 (0)1 (4)2 (8)0 (0)
 Head up and down0 (0)2 (8)2 (8)4 (16)5 (20)
 Bending forward0 (0)2 (8)1 (5.9)2 (8)4 (16)
 Reading passage0 (0)0 (0)1 (5.9)1 (4)0 (0)
Comfort of dressing
 Positive6 (24)21 (84)21 (84)22 (88)22 (88)
 Neutral13 (56)3 (12)4 (16)2 (8)1 (4)
 Negative5 (20)1 (4)0 (0)1 (4)2 (8)
Qualitative negative comments by category
 Sensation on skin4 (16)0 (0)0 (0)0 (0)0 (0)
 Feeling of mask fit/seal quality0 (0)3 (12)3 (12)3 (12)10 (40)
 Dressing adhesiveness4 (16)0 (0)4 (16)0 (0)1 (4)
 Dressing odor8 (32)0 (0)0 (0)0 (0)0 (0)

Primary outcome measures of qualitative fit test with failure rates for respective testing maneuvers and secondary outcomes of comfort of skin protectants and comments regarding comfort.

3M, St Paul, MN.

ConvaTec, Oklahoma City, OK.

Mölnlycke, Gothenburg, Sweden.

Study outcomes: Participant demographics/characteristics BMI, Body mass index. Outcome measures by dressing type, n (%)∗ Primary outcome measures of qualitative fit test with failure rates for respective testing maneuvers and secondary outcomes of comfort of skin protectants and comments regarding comfort. 3M, St Paul, MN. ConvaTec, Oklahoma City, OK. Mölnlycke, Gothenburg, Sweden. In this study, we found that the use of skin protectants to prevent skin irritation may interfere with N95 respirator fit. Mitigation of skin irritation from prolonged N95 use is a concern, but workers should not trade efficacy for comfort. Most fit test failures were observed with movement, suggesting that the impact of skin protectants on fit may not be obvious to the wearer. These results reinforce current guidelines, which recommend against placing any material(s) between the skin and mask or to retest fit if skin protectants are applied. , Proposed strategies for replacement of brand or style of N95 are reasonable, but these results show that current recommendations for liquid protectants should be reconsidered. Health care workers electing to use skin protectants should confirm appropriate fit before use in clinical settings. Although these data suggest heterogeneity in the impact of skin protectants on QLFT, the study was insufficiently powered to make definitive statements on the relative safety of these products. Nonetheless, it is important to consider how different materials affect respirator fit and comfort; no dressing had a complete QLFT passing rate. Passing rates were highest with Cavilon film (88%), but it had the most negative (24%) comfort ratings. This study is limited by sample size and a single N95 model tested. It does not evaluate the efficacy of skin protectants over longer periods of time. Larger studies examining fit and user experience across different mask and dressing types are warranted. Until these issues are further studied, health care workers and institutions electing to use facial skin protectants should ensure adequate fit by undergoing fit testing with facial skin protectants in place before use in high-risk clinical settings.
  4 in total

Review 1.  Adverse Effects of COVID-19 and Face Masks: A Systematic Review.

Authors:  Akshitha Thatiparthi; Jeffrey Liu; Amylee Martin; Jashin J Wu
Journal:  J Clin Aesthet Dermatol       Date:  2021-09-01

2.  De novo Powered Air-Purifying Respirator Design and Fabrication for Pandemic Response.

Authors:  Akshay Kothakonda; Lyla Atta; Deborah Plana; Ferrous Ward; Chris Davis; Avilash Cramer; Robert Moran; Jacob Freake; Enze Tian; Ofer Mazor; Pavel Gorelik; Christopher Van; Christopher Hansen; Helen Yang; Michael S Sinha; Ju Li; Sherry H Yu; Nicole R LeBoeuf; Peter K Sorger
Journal:  medRxiv       Date:  2021-03-29

Review 3.  Skin care and hygiene among healthcare professionals during and after the SARS-CoV-2 pandemic.

Authors:  Jason K Rivers; John P Arlette; Joel DeKoven; Lyn C Guenther; Channy Muhn; Vincent Richer; Nathan Rosen; Jean-François Tremblay; Marni C Wiseman; Catherine Zip; David Zloty
Journal:  SAGE Open Med       Date:  2021-12-08

Review 4.  Impact in Contact Dermatitis during and after SARS-CoV2 Pandemic.

Authors:  Graziella Babino; Giuseppe Argenziano; Anna Balato
Journal:  Curr Treat Options Allergy       Date:  2022-02-10
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.