Tough and injectable fiber reinforced calcium phosphate cement as an alternative to polymethylmethacrylate cement for vertebral augmentation: a biomechanical study.

Sónia de Lacerda Schickert1, João Castro Pinto1, John Jansen1, Sander C G Leeuwenburgh1, Jeroen J J P van den Beucken1.   

Abstract

Vertebral compression fractures (VCFs) are a very common problem among the elderly, which ultimately result in severe pain and a drastically reduced quality of life. An effective treatment for VCFs is the minimally invasive augmentation of the damaged vertebrae through vertebroplasty and/or kyphoplasty. These surgical procedures treat the affected vertebrae by injection of poly(methyl methacrylate) cement (PMMA) into the vertebral body. However, clinical use of PMMA cement is associated with major drawbacks. Bioceramic cements such as injectable calcium phosphate cements (CPC) exhibit a superior osteocompatibility over PMMA cements, but are too brittle for load-bearing applications. Here, we evaluated the handling and mechanical properties of a recently developed CPC formulation containing both poly(vinyl alcohol) (PVA) fibers and carboxymethyl cellulose (CMC) as an alternative to PMMA cement for vertebro- and kyphoplasty. Our results demonstrate that the addition of CMC rendered fiber-reinforced CPC injectable without negatively affecting its mechanical properties. Further, an ex vivo mechanical analysis clearly showed that extravasation of PVA fiber-reinforced CPC with CMC into trabecular bone was limited as compared to PMMA. Finally, we observed that the ex vivo biomechanical performance of vertebrae treated with CMC and PVA fibers was similar to PMMA-treated vertebrae. The obtained data suggests that PVA fiber-reinforced CPCs with CMC possesses adequate handling, mechanical and structural characteristics for vertebro- and kyphoplasty procedures. These data pave the way for future preclinical studies on the feasibility of treating vertebral compression fractures using PVA fiber-reinforced CPC with CMC.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 32579633     DOI: 10.1039/d0bm00413h

Source DB:  PubMed          Journal:  Biomater Sci        ISSN: 2047-4830            Impact factor:   6.843


  2 in total

1.  Injectable nanofiber-reinforced bone cement with controlled biodegradability for minimally-invasive bone regeneration.

Authors:  Peihao Cai; Shunyi Lu; Jieqin Yu; Lan Xiao; Jiayi Wang; Haifeng Liang; Lei Huang; Guanjie Han; Mengxuan Bian; Shihao Zhang; Jian Zhang; Changsheng Liu; Libo Jiang; Yulin Li
Journal:  Bioact Mater       Date:  2022-09-12

Review 2.  A 20-Year Review of Biomechanical Experimental Studies on Spine Implants Used for Percutaneous Surgical Repair of Vertebral Compression Fractures.

Authors:  Sairam Gajavelli; Aaron Gee; Z Shaghayegh Bagheri; Emil H Schemitsch; Christopher S Bailey; Parham Rasoulinejad; Radovan Zdero
Journal:  Biomed Res Int       Date:  2022-09-21       Impact factor: 3.246

  2 in total