Xiaoyi Zhu1, Zhiyu Huang1, Ziyuan Li1, Wenzhao Li1, Xi Liu1, Zhaolong Chen1, Jian Tian1, Changhui Li1.
Abstract
Photoacoustic microscopy (PAM) and optical coherence tomography (OCT) are sensitive to optical absorption and scattering characteristics, respectively. As such, the integration of these two modalities in order to combine important complementary information has garnered much attention. Due to the relatively low axial resolution ofEntities:
Keywords: Detection; Multiple imaging; Optical coherence tomography; Photoacoustic imaging
Year: 2020 PMID: 32577377 PMCID: PMC7300161 DOI: 10.1016/j.pacs.2020.100188
Source DB: PubMed Journal: Photoacoustics ISSN: 2213-5979
Comparison of axial resolutions from typical reported works.
| Methods | Axial resolution (μm) | |
|---|---|---|
| OCT | PA | |
| Fabry–Perot interferometry [ | 5–8 | 20–40 |
| Low-coherence interferometry [ | 8–13 | 20–60 |
Fig. 1Schematic of the PAM-OCT imaging system. SLD, super-luminescent diode; QWP, quarter-wave plate; DM, dichroic mirror; C1; 2, collimator; FC, fiber coupler; L1; 2; 3, lens; M1; 2; 3; 4, mirror; OL, objective lens; PBS, polarization beam splitter; P, prism; BD, balanced detector; STM32 F, a family of 32-bit microcontroller integrated circuits by STMicroelectronics.
Fig. 2Spatial resolution of the PAM-OCT system. (a) Axial resolution of PAM quantified using the shift-and-sum method; (b) axial resolution of SD-OCT quantified by the FWHM value of the A-line signal; (c) lateral resolution of the SD-OCT quantified using the edge spread function; (d) lateral resolution of PAM quantified using the edge spread function.
Fig. 3Dual modality imaging of four human hairs embedded in a scattering medium. (a) OCT MIP imaging result; (b) PAM MAP imaging result; (c) Photograph of the phantom; (d) B-scan images along the white dotted line in (a) and (b) by OCT and PAM, respectively. Scale bar: 500 μm.
Fig. 4Simultaneous in vivo imaging of a nude mouse ear. (a) OCT MIP image; (b) PAM MAP image; (c) B-scan image along the white dotted line in (a); (d) B-scan image along the white dotted line in (b); (e) fused B-scan image of the PAM-OCT system. SG, sebaceous gland; ED, epidermis; CT, cartilage; D, dermis. 3D visualizations of the fused image can be seen in Visualization 1 and Visualization 2. Scale bar ((a), (b)): 500 μm; scale bar ((c), (d), (e)): 250 μm.
Fig. 5Simultaneous in vivo imaging of a nude mouse hindlimb. (a) OCT MIP image; (b) PAM MAP image; (c) B-scan image along the white dotted line in (a); (d) B-scan image along the white dotted line in (b); (e) fused B-scan image of the dual modality PAM-OCT system; (f) Corresponding histology (H&E staining) of a nude mouse hindlimb. SG, sebaceous gland; ED, epidermis; D, dermis; HD, hypodermis; M, muscles. 3D visualizations of the fused image can be seen in Visualization 3 and Visualization 4. Scale bar ((a), (b)): 500 μm; scale bar ((c), (d), (e), (f)): 250 μm.