Robotic Transfer of the Latissimus Dorsi.

Jose Carlos Garcia1, Eduardo Ferreira Cordeiro1, Maurício de Paiva Raffaelli1, Marcelo Boulos Dumans Mello1, Márcio Eduardo Kozonara1, Álvaro da Motta Cardoso1, Marcelo Cerdan Torres2.   

Abstract

Robotic surgery has been used for a long time; it is earning space and its use is expanding in daily medical practice in several surgical specialties, with advantages over traditional surgical methods. This Technical Note presents an endoscopic robotic posterior shoulder approach that allows the surgeon to perform latissimus dorsi transfer endoscopically. This Technical Note describes the use of the da Vinci robot (Intuitive Surgical, Sunnyvale, CA) for transfers related to rotator cuff tears.
© 2020 by the Arthroscopy Association of North America. Published by Elsevier.

Entities:  

Year:  2020        PMID: 32577350      PMCID: PMC7301275          DOI: 10.1016/j.eats.2020.02.008

Source DB:  PubMed          Journal:  Arthrosc Tech        ISSN: 2212-6287


Robotic surgery has been used for a long time,; it is earning space and its use is expanding in daily medical practice in several surgical specialties, with advantages over traditional surgical methods., Within orthopaedics, we highlight the use of robotics in brachial plexus treatment, and neurologic releases.7, 8, 9 The association of robotic technology with endoscopy further allowed a faster recovery for the patient with many applications, with a shorter time of hospitalization and minimally invasive approach. The advantages of this method include movement accuracy, high-resolution imaging with 3-dimensional vision, gas infusion rather than saline solution (better visualization), filtering of the surgeon's tremor when manipulating objects, movement scaling, and hands-free camera manipulation.11, 12, 13, 14 In addition, in the future, there is the possibility of remote surgery (telesurgery) in which a surgical team can treat a patient far away, or a surgical team may be composed of professionals located in different cities or countries, treating the same patient simultaneously. Some shoulder pathologies that require a posterior shoulder approach may need aggressive and traumatic exposure with extensive manipulation of soft tissues. The possibility of using a minimally invasive approach can potentially be important for both the timing of rehabilitation and avoidance of local soft-tissue adhesions. In addition, when performing a large posterior open approach, one needs to use tensioned retractors to properly maintain the surgical field. The use of these tensioned retractors can eventually damage the deeper muscle layer as well as other neurovascular structures., Minimally invasive procedures have shown a decrease in adhesions, avoiding reoperation and physical therapy over a long period. Indeed, this advantage makes these procedures cost-effective. Literature regarding robotic assisted orthopedic surgery is scarce, as opposed to other surgical specialties.17, 18, 19 The use of robotic-assisted surgery for better identification of the quadrangular space of the shoulder, identification of the axillary and radial nerves, and better identification of the latissimus dorsi muscle was described previously in a cadaveric trial on shoulder surgery. The visualization and partial manipulation of the latissimus dorsi muscle have already been reported to aid the transportation of the muscular pedicle, with a technique that we used as a reference for our study. Axillary nerve identification has also been described, making a contribution to our study and confirming the viability of the method. Regarding bleeding, studies in live patients have shown that air insufflation was effective at avoiding it. This Technical Note is based on a previous cadaveric trial and aims to present the use of the da Vinci robot for transfers of the latissimus dorsi tendon to the humeral head; we have performed this technique to treat massive rotator cuff tears in 4 live patients.

Indications, Preoperative Evaluation, and Imaging

This procedure has the same indications as traditional open transfer of the latissimus dorsi tendon: patients with massive rotator cuff tears with poor biological characteristics of the tendon. The ideal surgical indications are grade 3 or 4 fatty infiltration according to the Goutallier classification with retraction of more than 4 cm in a patient younger than 60 years presenting with a functional subscapularis tendon. The patient will present with pain, and shoulder elevation and abduction will be difficult or impossible to perform. The younger the patient is, the better the results tend to be.

Surgical Technique

The patient is placed in the ventral decubitus position; the arm is maintained in a position similar to 90° of elevation. The inferior border of the latissimus dorsi can be localized by palpation. A drawing of the muscle is made on the skin based on its inferior border and its known anatomy. The central line of the latissimus dorsi is also drawn. A 1-cm incision is made in the skin, 10 to 15 cm from the axilla, to establish the central portal. Two other portals are made 5 to 7 cm perpendicular medial and lateral to the central line. These portals are located 7 to 10 cm from the axilla. The central portal is used to insert the optics, and through the other 2 portals, the robotic hands are introduced to access the muscular fascia, where a cavity is formed through blunt dissection. This space is made for triangulation as an initial working cavity once there are no natural cavities in this region. A trocar and cannula are introduced into each incision in a common direction in the cavity the surgeon created (Fig 1). The camera of the da Vinci SI or Xi robot (Intuitive Surgical, Sunnyvale, CA), with an optic of 30° (Fig 2), is introduced in the first portal on the trapezius.
Fig 1

Patient in ventral decubitus position. Right shoulder and thorax. (A, B) Robotic hand trocars. (C) Optics trocar. (D) Shoulder. (E) Robotic exterior hand. (F) Robotic exterior hand for optics.

Fig 2

A robotic optic of 30° with 2 cameras allows a stereoscopic view.

Patient in ventral decubitus position. Right shoulder and thorax. (A, B) Robotic hand trocars. (C) Optics trocar. (D) Shoulder. (E) Robotic exterior hand. (F) Robotic exterior hand for optics. A robotic optic of 30° with 2 cameras allows a stereoscopic view. Carbon dioxide is inflated at a constant pressure of 8 to 14 mm Hg through the chamber portal into the working cavity, stretching the soft tissues and opening the cavity. The robotic arms use Cadiere Forceps (8 mm; Intuitive Surgical) and Hot Shears Monopolar Curved Scissors (8 mm; Intuitive Surgical). The first objective is to clean the area around the camera so that the best dissection and identification of the initial working cavity are achieved. After this first stage, we search for the superior border of the latissimus dorsi muscle and its division with the teres major. Dissection using this muscular plane is performed until its entrance deep into the medial border of the long head of the triceps (Fig 3).
Fig 3

Optics are placed in the central portal. (LD, latissimus dorsi; TM, teres major.)

Optics are placed in the central portal. (LD, latissimus dorsi; TM, teres major.) The latissimus is released and separated from the teres major (Fig 4); the radial nerve is just below the latissimus, and it is possible to visualize it but not required (Fig 5). Care must be taken to avoid damaging the neurovascular pedicle. A No. 0 Vicryl (J318; Johnson & Johnson, São Paulo, Brazil) is inserted by the cephalic robotic hand’s portal. The latissimus dorsi tendon is sutured by using Cadiere and DeBakey Forceps (Intuitive Surgical) (Fig 6, Video 1). The sutured tendon is pulled out of the body through the central portal of the optics using a gastric forceps.
Fig 4

With optics within the central portal, the latissimus dorsi (LD) is released. (TM, teres major.)

Fig 5

Optics within central portal showing radial nerve (A), teres major insertion (B), and retracted latissimus dorsi (C).

Fig 6

Optics within central portal showing latissimus dorsi (LD), long head of triceps (TLH), and teres major (TM).

With optics within the central portal, the latissimus dorsi (LD) is released. (TM, teres major.) Optics within central portal showing radial nerve (A), teres major insertion (B), and retracted latissimus dorsi (C). Optics within central portal showing latissimus dorsi (LD), long head of triceps (TLH), and teres major (TM). A small 3- to 5-cm incision is made on the lateral deltoid, and by use of a finger, dissection of the subdeltoid space is performed until the cavity created by the robot is reached. A long gastric grasper is inserted through the cephalic robotic hand’s portal until it reaches the subdeltoid space. A guide polyester No. 5 wire is passed by using this grasper, leaving the optics portal (Fig 7).
Fig 7

The patient is in the ventral decubitus position; the shoulder and scapular area are exposed. (A) Polyester guidewire through deltoid approach. (B) Gastric forceps. (C) Wires where latissimus dorsi was sutured. (D) Polyester guidewire exit.

The patient is in the ventral decubitus position; the shoulder and scapular area are exposed. (A) Polyester guidewire through deltoid approach. (B) Gastric forceps. (C) Wires where latissimus dorsi was sutured. (D) Polyester guidewire exit. The humeral head is drilled, and 2 anchors are inserted. The anchor wires are passed, through the deltoid approach, to the optics portal using the No. 5 polyester wire as a guide. The tendon is sutured to the anchor wires using Krackow stitches (Fig 8) and is passed to the subacromial space pulled by the anchor wires, and standard tendon-to-bone suturing is performed. More anchors and sutures can be placed after the tendon lies on the humerus’ greater tuberosity. The portals and deltoid lateral approach are sutured.
Fig 8

The patient is in the ventral decubitus position; the shoulder and scapular area are exposed. (A) The latissimus dorsi is sutured with suture anchor wires. (B) The other parts of the suture anchor wires are shown; these will pull the tendon to the humeral head.

The patient is in the ventral decubitus position; the shoulder and scapular area are exposed. (A) The latissimus dorsi is sutured with suture anchor wires. (B) The other parts of the suture anchor wires are shown; these will pull the tendon to the humeral head.

Rehabilitation

A sling is used for 5 weeks. Pendular movements and passive elevation until 90° are allowed 2 weeks after surgery. After this period, active exercises with isometric external rotation and elevation begin. Two weeks thereafter, isokinetic and proprioception movements begin. Scapular retraction and shoulder extension need to be stimulated in the initial movements, once the latissimus dorsi can also be activated during these movements. Better evolution is present in patients with better active movements before surgery. Pearls and pitfalls of the described technique are summarized in Table 1. Advantages and disadvantages are summarized in Table 2.
Table 1

Pearls and Pitfalls

Pearls and Pitfalls
Robot dockingThe da Vinci robot is designed for abdominal exploration; the robot will need to be cephalic with the optics pointed to the central line drawn on the patient’s skin.
Movements and subcutaneous dissectionThe movements in robotic surgery are different from those in regular endoscopy. Subcutaneous can be better dissected by using horizontal movements of both robotic hands, as usual in robotic surgery, just above the superior portion of the latissimus dorsi and teres major.
Triceps tendonThe triceps tendon tends to be close to the working space; the assistant surgeon within the surgical field can insert the aspirator through the triceps to aspirate blood and help the surgeon open the working space. The best location is chosen by an endoscopic view with a needle.
Radial nerveThe radial nerve is just a few centimeters under the latissimus dorsi; anatomic knowledge of this area and training on a cadaveric model are strongly recommended to avoid neurologic complications.
Table 2

Advantages and Disadvantages Comparing Robotic, Arthroscopic, and Open Surgical Procedures for Latissimus Dorsi Transfer

RoboticArthroscopicOpen
ScarringLowIntermediateHigh
Tendon visualizationGoodIntermediateIntermediate
Movement precisionHighLowIntermediate
Surgical timeLongerIntermediateFaster
CostIntermediateMore ExpensiveLess Expensive
Minimal invasivenessLess InvasiveIntermediateOpen
Pearls and Pitfalls Advantages and Disadvantages Comparing Robotic, Arthroscopic, and Open Surgical Procedures for Latissimus Dorsi Transfer

Discussion

Traditional approaches for the latissimus dorsi are wide, requiring large posterior incisions with cosmetic and scar-formation implications. Previous cadaveric and live-patient studies were used to establish the principles for the robotic latissimus dorsi transfer presented in this Technical Note. We aim to present a surgical technique that can be improved on and can even be used for other orthopaedic applications with the future introduction of robots, robotic arms, and smaller optics. Few studies have assessed the latissimus dorsi using the aid of robotics; all of them accessed the muscle and the origin of the latissimus for free flaps.17, 18, 19, 20 We describe the first in vivo robotic-assisted shoulder surgical procedure performed for transfer of the latissimus dorsi insertion to improve function after a rotator cuff tear. We previously robotically identified the tendon, neurovascular structures, and quadrangular space in cadaveric trials in other trials; thus, the surgical viability and safety of the procedure were previously examined., In addition, other robotic orthopaedic applications in live patients have shown the effectiveness of air insufflation in achieving better bleeding control. We have shown the viability of the introduction of robotics in shoulder surgery in the hope of encouraging further studies in the area. The limitations of this technique are the cost of the robot, robotic hands, and scissors. The necessity for specific training on robotic surgery, which is currently costly and not available in many hospitals, can also limit its current use. The surgical time is currently longer than that of the open procedure; however, this situation tends to improve in time.
  19 in total

1.  The da Vinci telerobotic surgical system: the virtual operative field and telepresence surgery.

Authors:  Garth H Ballantyne; Fred Moll
Journal:  Surg Clin North Am       Date:  2003-12       Impact factor: 2.741

2.  Three-dimensional imaging improves surgical performance for both novice and experienced operators using the da Vinci Robot System.

Authors:  John C Byrn; Stefanie Schluender; Celia M Divino; John Conrad; Brooke Gurland; Edward Shlasko; Amir Szold
Journal:  Am J Surg       Date:  2007-04       Impact factor: 2.565

3.  Telerobotic anterior translocation of the ulnar nerve.

Authors:  Jose Carlos Garcia; Gustavo Mantovani; Stephanie Gouzou; Philippe Liveneaux
Journal:  J Robot Surg       Date:  2011-02-18

4.  Endoscopic exploration and repair of brachial plexus with telerobotic manipulation: a cadaver trial.

Authors:  Gustavo Mantovani; Philippe Liverneaux; Jose Carlos Garcia; Stacey H Berner; Michael S Bednar; Catherine J Mohr
Journal:  J Neurosurg       Date:  2011-04-08       Impact factor: 5.115

5.  New Frontiers in Robotic Surgery: The latest high-tech surgical tools allow for superhuman sensing and more.

Authors:  Michele Solis
Journal:  IEEE Pulse       Date:  2016 Nov-Dec       Impact factor: 0.924

6.  Endoscopic robotic decompression of the ulnar nerve at the elbow.

Authors:  Jose Carlos Garcia; Edna Frasson de Souza Montero
Journal:  Arthrosc Tech       Date:  2014-06-09

7.  Supervised autonomous robotic soft tissue surgery.

Authors:  Azad Shademan; Ryan S Decker; Justin D Opfermann; Simon Leonard; Axel Krieger; Peter C W Kim
Journal:  Sci Transl Med       Date:  2016-05-04       Impact factor: 17.956

8.  Robotic Versus Laparoscopic Staging for Early Ovarian Cancer: A Case-Matched Control Study.

Authors:  Valerio Gallotta; Carla Cicero; Carmine Conte; Giuseppe Vizzielli; Marco Petrillo; Anna Fagotti; Vito Chiantera; Barbara Costantini; Giovanni Scambia; Gabriella Ferrandina
Journal:  J Minim Invasive Gynecol       Date:  2016-11-14       Impact factor: 4.137

9.  Colorectal surgery in elderly patients: our experience with DaVinci Xi® System.

Authors:  A Oldani; P Bellora; M Monni; B Amato; S Gentilli
Journal:  Aging Clin Exp Res       Date:  2016-11-26       Impact factor: 3.636

10.  Vulnerable neurovasculature with a posterior approach to the scapula.

Authors:  Coen A Wijdicks; Bryan M Armitage; Jack Anavian; Lisa K Schroder; Peter A Cole
Journal:  Clin Orthop Relat Res       Date:  2008-12-04       Impact factor: 4.176

View more
  2 in total

1.  Robotic Transfer of the Latissimus Dorsi for Irreparable Subscapularis Tear.

Authors:  Jose Carlos Garcia
Journal:  Arthrosc Tech       Date:  2022-05-17

2.  Technical Guide and Tips to Posterior Arthroscopic Latissimus Dorsi Transfer for Irreparable Posterosuperior Rotator Cuff Tears.

Authors:  Vanesa Lopez-Fernandez; Sandrine Mariaux; Laurent Lafosse; Thibault Lafosse
Journal:  Arthrosc Tech       Date:  2022-04-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.