C E Spritzer1, J B Vogler, S Martinez, W E Garrett, G A Johnson, M J McNamara, J Lohnes, R J Herfkens.
Abstract
The knees of 17 patients (18 extremities) with possible meniscal, cruciate ligament, and articular cartilage abnormalities were examined with a three-dimensional Fourier transform (3DFT), gradient-refocused acquisition in a steady state (GRASS) pulse sequence. Arthroscopic confirmation was available in all cases and was the standard for comparison. Thirteen of these extremities were also examined by using a two-dimensional Fourier transform spin-echo pulse sequence with a 2000-msec repetition time and 20- and 80-msec echo time. In these 13 cases, both pulse sequences correctly identified seven of eight meniscal abnormalities. However, interpretation of the 3DFT GRASS images resulted in fewer false-positive meniscal tears (three vs six). Cruciate ligament tears were detected more readily on the 3DFT GRASS images (six vs three with two possible tears on the spin-echo images). These preliminary findings suggest that the overall accuracy of MR imaging of the knee could be improved by including 3DFT gradient-refocused pulse sequences.Entities:
Mesh:
Year: 1988 PMID: 3257617 DOI: 10.2214/ajr.150.3.597