Reliable identification of cardiac liability in drug discovery using automated patch clamp: Benchmarking best practices and calibration standards for improved proarrhythmic assessment.

Nina Brinkwirth1, Kiyoshi Takasuna2, Masafumi Doi3, Nadine Becker1, Alison Obergrussberger1, Søren Friis1, Hatsue Furukawa2, Yuka Hasegawa2, Takayuki Oka4, Atsushi Ohtsuki4, Niels Fertig1, Sonja Stoelzle-Feix5.   

Abstract

INTRODUCTION: Screening compounds for activity on the hERG channel using patch clamp is a crucial part of safety testing. Automated patch clamp (APC) is becoming widely accepted as an alternative to manual patch clamp in order to increase throughput whilst maintaining data quality. In order to standardize APC experiments, we have investigated the effects on IC50 values under different conditions using several devices across multiple sites.
METHODS: APC instruments SyncroPatch 384i, SyncroPatch 384PE and Patchliner, were used to record hERG expressed in HEK or CHO cells. Up to 27 CiPA compounds were used to investigate effects of voltage protocol, incubation time, labware and time between compound preparation and experiment on IC50 values.
RESULTS: All IC50 values of 21 compounds recorded on the SyncroPatch 384PE correlated well with IC50 values from the literature (Kramer et al., 2013) regardless of voltage protocol or labware, when compounds were used immediately after preparation, but potency of astemizole decreased if prepared in Teflon or polypropylene (PP) compound plates 2-3 h prior to experiments. Slow acting compounds such as dofetilide, astemizole, and terfenadine required extended incubation times of at least 6 min to reach steady state and therefore, stable IC50 values. DISCUSSION: Assessing the influence of different experimental conditions on hERG assay reliability, we conclude that either the step-ramp protocol recommended by CiPA or a standard 2-s step-pulse protocol can be used to record hERG; a minimum incubation time of 5 min should be used and although glass, Teflon, PP or polystyrene (PS) compound plates can be used for experiments, caution should be taken if using Teflon, PS or PP vessels as some adsorption can occur if experiments are not performed immediately after preparation. Our recommendations are not limited to the APC devices described in this report, but could also be extended to other APC devices.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Automated patch clamp; Cardiovascular safety; Comprehensive in vitro Proarrhythmia assay (CiPA); Ion channels; Nonclinical safety, HTS; Proarrhythmic risk assessment; Safety assessment; Safety pharmacology; hERG

Mesh:

Substances:

Year:  2020        PMID: 32565325     DOI: 10.1016/j.vascn.2020.106884

Source DB:  PubMed          Journal:  J Pharmacol Toxicol Methods        ISSN: 1056-8719            Impact factor:   1.950


  3 in total

1.  Intracellular Binding of Terfenadine Competes with Its Access to Pancreatic ß-cell ATP-Sensitive K+ Channels and Human ether-à-go-go-Related Gene Channels.

Authors:  Bernd J Zünkler; Maria Wos-Maganga; Stefanie Bohnet; Anne Kleinau; Detlef Manns; Shivani Chatterjee
Journal:  J Membr Biol       Date:  2022-06-28       Impact factor: 1.843

2.  Biophysical characterization of light-gated ion channels using planar automated patch clamp.

Authors:  Elena G Govorunova; Oleg A Sineshchekov; Leonid S Brown; John L Spudich
Journal:  Front Mol Neurosci       Date:  2022-08-09       Impact factor: 6.261

3.  There is no F in APC: Using physiological fluoride-free solutions for high throughput automated patch clamp experiments.

Authors:  Markus Rapedius; Alison Obergrussberger; Edward S A Humphries; Stephanie Scholz; Ilka Rinke-Weiss; Tom A Goetze; Nina Brinkwirth; Maria Giustina Rotordam; Tim Strassmaier; Aaron Randolph; Søren Friis; Aiste Liutkute; Fitzwilliam Seibertz; Niels Voigt; Niels Fertig
Journal:  Front Mol Neurosci       Date:  2022-08-22       Impact factor: 6.261

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.