Literature DB >> 32556821

Comparison of Two Specialized Histology Needles for Endoscopic Ultrasound (EUS)-Guided Liver Biopsy: A Pilot Study.

Rintaro Hashimoto1, David P Lee1, Jason B Samarasena1, Vishal S Chandan2, Wenchang Guo2, John G Lee1, Kenneth J Chang3.   

Abstract

BACKGROUND: EUS-guided liver biopsy (EUS-LB) has been shown to be a safe and effective alternative to percutaneous liver biopsy. The optimal needle device and technique for EUS-LB is still evolving. The aim of this study was to compare the efficacy of two second-generation 19G fine-needle biopsy (FNB) (Franseen- and Fork-tip) devices for EUS-LB.
METHODS: This is a repeated-measure crossover study with a prospectively maintained cohort of patients. We performed EUS-LB with a one-pass and single-actuation method using two 19G FNB needles in 22 consecutive patients between 10/2018 and 9/2019. Patients were randomized to left vs right liver lobes to be biopsied as well as the needle sequence. The specimens obtained were evaluated for adequacy for histologic diagnosis. The primary outcome was number of complete portal tracts (CPTs), post-fix aggregate, and longest specimen length. Secondary outcomes were prefix aggregate specimen length and the specimen adequacy judged by two expert pathologists.
RESULTS: A total of 44 liver biopsies were performed in 22 patients. The CPTs were higher in the Franseen-tip needle group compared to the Fork-tip needle group (14.4 vs 9.5, p = 0.043). Post-fix aggregate specimen length (44.9 mm vs 34.6 mm, p = 0.097), the post-fix longest specimen length (19.9 mm vs 13.7 mm, p = 0.175), and prefix aggregate specimen length (51.7 mm vs 45 mm, p = 0.265) were not significantly different. Both needles showed similarly high histologic adequacy (100% vs 95.5%, p = 0.312). Interestingly, the right of the liver showed higher yield of CPTs with both needles (Franseen, 16.2 vs. 12.8, p = 0.003, the Fork-tip, 12.8 vs. 7.0, p < 0.0001).
CONCLUSION: EUS-guided liver biopsy using the 19G Franseen-tip needle may provide more CPTs than 19G Fork-tip needle on a single-pass, single-actuation comparison.

Entities:  

Keywords:  Endoscopic ultrasound-guided fine-needle aspiration; Endosonography; Liver biopsy; Observational study

Year:  2020        PMID: 32556821     DOI: 10.1007/s10620-020-06391-3

Source DB:  PubMed          Journal:  Dig Dis Sci        ISSN: 0163-2116            Impact factor:   3.199


  1 in total

1.  Endoscopic Ultrasound-Guided Fine Needle Biopsy Using 22-Gauge Franseen Needle for the Histological Diagnosis of Solid Lesions: A Multicenter Prospective Pilot Study.

Authors:  Naoki Mita; Takuji Iwashita; Shinya Uemura; Yuhei Iwasa; Katsuhisa Toda; Tsuyoshi Mukai; Tatsuhiko Miyazaki; Ichiro Yasuda; Masahito Shimizu
Journal:  Dig Dis Sci       Date:  2019-09-18       Impact factor: 3.199

  1 in total
  2 in total

Review 1.  Update on endoscopic ultrasound-guided liver biopsy.

Authors:  Shiva Rangwani; Devarshi R Ardeshna; Khalid Mumtaz; Sean G Kelly; Samuel Y Han; Somashekar G Krishna
Journal:  World J Gastroenterol       Date:  2022-07-28       Impact factor: 5.374

2.  EUS-guided liver biopsy using a modified wet heparin suction technique.

Authors:  Ali Zakaria; Abdulrahman Diab; Michael Dowd; Pushpak Taunk; Ali Abbas
Journal:  VideoGIE       Date:  2022-08-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.