Precision and accuracy of the Mindray BC-5000Vet hematology analyzer for canine and feline blood.

Sorawat Thongsahuan1, Lalita Fonghoi2, Sunita Kaewfai2, Kanitta Srinoun2.   

Abstract

BACKGROUND: The Mindray BC-5000Vet hematology analyzer is a flow cytometry-based automated hematology analyzer that generates a complete blood count with a five-part white blood cell (WBC) differential count.
OBJECTIVES: We aimed to validate reliability results of the Mindray BC-5000Vet for use in dog and cat blood.
METHODS: Imprecision was performed using the manufacturer's quality control material at low, normal, and high levels. Blood sample results of healthy and ill dogs and cats were studied for comparability between manual methods and the Mindray BC-5000Vet analyzer. Forty dogs and 40 cats were included in the study.
RESULTS: Precision for red blood cell (RBC) parameters was excellent, with a coefficient of variation within-run (%CVw ) and between-run (%CVb ) of <2%. WBC count and differential count showed %CVw and %CVb <8%; however, %CVw and %CVb of low-level control material gave >9% eosinophils. The correlation between the BC-5000 Vet and manual methods in normal and abnormal canine and feline blood samples showed excellent correlations for the RBC counts, hemoglobin concentrations, hematocrits, and WBC counts (r > .93). The differential WBC analysis of canine blood showed good correlation (r = .80-.92). Feline blood samples showed excellent correlations for neutrophils and lymphocytes, with a good correlation for monocytes and eosinophils.
CONCLUSIONS: The Mindray BC-5000Vet hematology analyzer proved a suitable instrument for routine analysis in dogs and cats with various hematologic abnormalities.
© 2020 American Society for Veterinary Clinical Pathology.

Entities:  

Keywords:  accuracy; blood count; cat; dog; precision

Year:  2020        PMID: 32524640     DOI: 10.1111/vcp.12861

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  2 in total

1.  Hematological profile of blood parasitic infected dogs in Southern Thailand.

Authors:  Sorawat Thongsahuan; Usa Chethanond; Siriwat Wasiksiri; Vannarat Saechan; Wichaya Thongtako; Tipayaratn Musikacharoen
Journal:  Vet World       Date:  2020-11-10

2.  Evaluation of the scil vCell 5, a novel laser- and impedance-based point-of-care hematology analyzer, for use in dogs and cats.

Authors:  Kim-Lina Charlotte Zelmer; Natali Bauer; Andreas Moritz
Journal:  J Vet Diagn Invest       Date:  2022-03-24       Impact factor: 1.279