An Analysis of the Effects of Noisy Electrocardiogram Signal on Heartbeat Detection Performance.

Ziti Fariha Mohd Apandi1, Ryojun Ikeura2, Soichiro Hayakawa2, Shigeyoshi Tsutsumi2.   

Abstract

Heartbeat detection for ambulatory cardiac monitoring is more challenging as the level of noise and artefacts induced by daily-life activities are considerably higher than monitoring in a hospital setting. It is valuable to understand the relationship between the characteristics of electrocardiogram (ECG) noises and the beat detection performance in the cardiac monitoring system. For this purpose, three well-known algorithms for the beat detection process were re-implemented. The beat detection algorithms were validated using two types of ambulatory datasets, which were the ECG signal from the MIT-BIH Arrhythmia Database and the simulated noise-contaminated ECG signal with different intensities of baseline wander (BW), muscle artefact (MA) and electrode motion (EM) artefact from the MIT-BIH Noise Stress Test Database. The findings showed that signals contaminated with noise and artefacts decreased the potential of beat detection in ambulatory signal with the poorest performance noted for ECG signal affected by the EM artefacts. In conclusion, none of the algorithms was able to detect all QRS complexes without any false detection at the highest level of noise. The EM noise influenced the beat detection performance the most in comparison to the MA and BW noises that resulted in the highest number of misdetections and false detections.

Entities:  

Keywords:  ECG analysis; ambulatory ECG signal; cardiac monitoring; heartbeat detection; noisy signal

Year:  2020        PMID: 32517214     DOI: 10.3390/bioengineering7020053

Source DB:  PubMed          Journal:  Bioengineering (Basel)        ISSN: 2306-5354


  3 in total

1.  Influence of Artefact Correction and Recording Device Type on the Practical Application of a Non-Linear Heart Rate Variability Biomarker for Aerobic Threshold Determination.

Authors:  Bruce Rogers; David Giles; Nick Draper; Laurent Mourot; Thomas Gronwald
Journal:  Sensors (Basel)       Date:  2021-01-26       Impact factor: 3.576

2.  Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women.

Authors:  Marcelle Schaffarczyk; Bruce Rogers; Rüdiger Reer; Thomas Gronwald
Journal:  Sensors (Basel)       Date:  2022-08-30       Impact factor: 3.847

3.  Noise-Reducing Fabric Electrode for ECG Measurement.

Authors:  Takamasa Terada; Masahiro Toyoura; Takahide Sato; Xiaoyang Mao
Journal:  Sensors (Basel)       Date:  2021-06-23       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.