| Literature DB >> 32514397 |
Joanna Reid1, Gert Van Zyl1, Michael Linström1, Stephen Korsman2, Gert Marais2, Wolfgang Preiser1.
Abstract
The consequences of falsely reactive HIV test results can be significant, for patients and healthcare providers. This case describes a diagnostic investigation of a patient with pronounced discordant HIV serological results, to determine HIV status. The fourth generation serological screening assay (Roche COBAS Elecsys HIV combiPT) had high positive results but confirmatory testing was negative (Abbott HIV Ag/Ab Combo). Five separate samples over 13 days were tested using multiple fourth generation HIV immunoassays and molecular tests for HIV-1 and HIV-2. Potential causes of falsely reactive serological results were investigated. Samples were sent to the manufacturer for analysis. The screening assay was positive on all samples with a very high signal to cut-off ratio (S/CO) of greater than 400. However, multiple serological and molecular assays did not detect HIV-1 or HIV-2 specific antibodies, antigen or nucleic acid. A recombinant immunochromatographic assay had faint reactivity to gp41 peptide and the manufacturer investigation reported cross-reactivity to one of the screening assay's synthetic peptides. Possible causes of the false positive result include cross reactivity to other antigens, including prior schistosomiasis infection, or the patient's previously excised ameloblastoma (a rare germ cell tumor of the jaw). This is a rare case of false high positive results on fourth-generation HIV serology testing due to high level non-specific reactivity to an isolated synthetic peptide component of the assay. It highlights the need for confirmatory testing even in settings with HIV high prevalence and awareness that false-positive serological results may have a high S/CO.Entities:
Keywords: False positive HIV serology; HIV Elisa; HIV diagnosis; HIV serology; Human immunodeficiency virus
Year: 2020 PMID: 32514397 PMCID: PMC7267741 DOI: 10.1016/j.idcr.2020.e00849
Source DB: PubMed Journal: IDCases ISSN: 2214-2509
Serological results.
| Date: April 2019 | 5/4 | 9/4 | 9/4 | 10/4 | 10/4 | 17/4 | |
|---|---|---|---|---|---|---|---|
| Sample ID | 1 | 2 | 3 | 4 | 5 | 6 | |
| Instrument /platform | Assay/ test | ||||||
| COBAS e 601 | Elecsys®HIV combi PT | R | R | R | R | R | |
| Abbott Architect i2000SR | HIV Ag/Ab Combo | NR | NR | NR | NR | NR | |
| Bio-Rad | Bio-Rad Genscreen | N | |||||
| ULTRA HIV Ag/Ab | |||||||
| Rapid test | Visitect HIV1/2 Rapid | NR | |||||
| Rapid test | Alere™ HIV Combo | N | |||||
| Rapid test | Abon HIV 1/2/O Tri-line | NR | |||||
| Rapid test | ADVANCED QUALITY™ | NR | |||||
| Rapid Anti-HIV(1&2) |
Molecular results.
| Date: April 2019 | 5/4 | 9/4 | 9/4 | 10/4 | 10/4 | 17/4 | |
|---|---|---|---|---|---|---|---|
| Sample ID | 1 | 2 | 3 | 4 | 5 | 6 | |
| Instrument /platform | Assay/ test | ||||||
| COBAS®AmpliPrep/ | HIV-1 Quantitative | LDL | LDL | ||||
| COBAS® TaqMan | Test V2.0 | ||||||
| COBAS®AmpliPrep/ | HIV-1 Qualitative | ND | ND | ||||
| COBAS® | |||||||
| TaqMan | Test V2.0 | ||||||
| Alere q | HIV-1/2 | ND | |||||
| Detect (quant PCR) |