Literature DB >> 32498880

Interference-free, green microanalytical method for total mercury and methylmercury determination in biological and environmental samples using small-sized electrothermal vaporization capacitively coupled plasma microtorch optical emission spectrometry.

Simion Bogdan Angyus1, Eugen Darvasi2, Michaela Ponta2, Dorin Petreus3, Radu Etz3, Marin Senila4, Maria Frentiu4, Tiberiu Frentiu5.   

Abstract

An analytical method for the quantification of total Hg and CH3Hg+ in biological tissues (fish, mushroom) and water sediment was developed based on small-sized electrothermal vaporization capacitively coupled plasma microtorch optical emission spectrometry using a low-resolution microspectrometer as detector. Sample preparation was carried out according to the procedure recommended by JRC Technical Report of European Commission for the determination of CH3Hg+ in seafood and adapted by us for lower consumption of reagents. Amounts of 0.1 - 0.5 g sample were subjected to extraction in 5 ml of 47% HBr then CH3Hg+ was extracted in 2 × 1 ml toluene and back-extracted in 2 ml aqueous solution of 1% l-cysteine. Total Hg/CH3Hg+ were quantified in 10 μl of acidic extract/l-cysteine solution after electrothermal vaporization and measurement of 253.652 nm Hg signal in the episodic emission spectra. Under the optimal working conditions of system (70 °C sample drying, 1300 °C sample vaporization, 10 W plasma power and 150 ml min-1 Ar flow) the limits of detection were 7.0 μg kg-1 total Hg and 3.5 μg kg-1 CH3Hg+. Comparison of slopes in external calibration and standard addition procedure revealed the lack of non-spectral interferences of multimineral matrix, so that the calibration against Hg2+ standards was adopted. Pooled recovery of total mercury/methylmercury was 101 ± 7%/100 ± 7%, while precision assessed from measurements of real samples was in the range 1.6-9.6%/2.7-12.8%. The proposed method validated according to Eurachem Guide 2014 is selective and complies with demands in European legislation (Decisions 657/2002; 333/2007; 836/2011) and Association of Official Analytical Chemists Guide in terms of performances for food control. The method displays a high degree of greenness by circumventing cold vapor generation, use of small amounts of reagents and full-miniaturized instrumentation resulting in low analytical costs without reducing results quality. Besides, the method is simple and rapid, since it uses external calibration curves prepared from Hg2+standard solutions both for total Hg and CH3Hg+ determination.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Electrothermal vaporization; Green analytical method; Mercury; Methylmercury; Miniaturized instrumentation; Optical emission spectrometry

Year:  2020        PMID: 32498880     DOI: 10.1016/j.talanta.2020.121067

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  2 in total

1.  Improving the Voltammetric Determination of Hg(II): A Comparison Between Ligand-Modified Glassy Carbon and Electrochemically Reduced Graphene Oxide Electrodes.

Authors:  Matei D Raicopol; Andreea M Pandele; Constanţa Dascălu; Eugeniu Vasile; Anamaria Hanganu; Gabriela-Geanina Vasile; Ioana Georgiana Bugean; Cristian Pirvu; Gabriela Stanciu; George-Octavian Buica
Journal:  Sensors (Basel)       Date:  2020-11-28       Impact factor: 3.576

2.  Simultaneous Determination of As, Bi, Sb, Se, Te, Hg, Pb and Sn by Small-Sized Electrothermal Vaporization Capacitively Coupled Plasma Microtorch Optical Emission Spectrometry Using Direct Liquid Microsampling.

Authors:  Simion Bogdan Angyus; Erika Levei; Dorin Petreus; Radu Etz; Eniko Covaci; Oana Teodora Moldovan; Michaela Ponta; Eugen Darvasi; Tiberiu Frentiu
Journal:  Molecules       Date:  2021-04-30       Impact factor: 4.411