Evaluation of a new wireless technique for continuous electroencephalography monitoring in neurological intensive care patients.

Marco Meyer1, Sven Fuest2, Dominique Krain3, Martin Juenemann3, Tobias Braun3, Serge C Thal4, Patrick Schramm5.   

Abstract

A novel wireless eight-channel electroencephalography (EEG) headset specially developed for ICUs was tested in regard of comparability with standard 10/20 EEG systems. The continuous EEG (cEEG) derivations via CerebAir EEG headset (Nihon Kohden Europe, Rosbach, Germany) and internationally standardized 10/20 reference EEGs as the diagnostic standard were performed in a mixed collective on a neurointensive care unit (neuro-ICU). The derivations were verified for comparability in detection of EEG background activity, epileptiform discharges, and seizure patterns. Fifty-two patients with vigilance reduction following serious neurological or metabolic diseases were included, and both methods were applied and further analyzed in 47. EEG background activity matched in 24 of 45 patients (53%; p = 0.126), epileptiform discharges matched in 32 (68%) patients (p = 0.162), and seizure activity matched in 98%. Overall, in 89% of the patients, cEEG detected the same or additional ICU-relevant EEG patterns. The tested wireless cEEG headset is a useful monitoring tool in patients with consciousness disorders. The present study indicates that long-term measurements with the wireless eight-channel cEEG lead to a higher seizure and epileptiform discharge detection compared to intermittent 10/20 EEG derivations in the ICU setting.
© 2020. Springer Nature B.V.

Entities:  

Keywords:  Bedside monitoring; Electroencephalography; Intensive care; Wireless EEG headset

Year:  2020        PMID: 32488677     DOI: 10.1007/s10877-020-00533-8

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  13 in total

Review 1.  Admission to neurological intensive care: who, when, and why?

Authors:  Robin S Howard; Dimitri M Kullmann; Nicholas P Hirsch
Journal:  J Neurol Neurosurg Psychiatry       Date:  2003-09       Impact factor: 10.154

Review 2.  Recommendations on the use of EEG monitoring in critically ill patients: consensus statement from the neurointensive care section of the ESICM.

Authors:  Jan Claassen; Fabio S Taccone; Peter Horn; Martin Holtkamp; Nino Stocchetti; Mauro Oddo
Journal:  Intensive Care Med       Date:  2013-05-08       Impact factor: 17.440

Review 3.  American Clinical Neurophysiology Society's Standardized Critical Care EEG Terminology: 2012 version.

Authors:  L J Hirsch; S M LaRoche; N Gaspard; E Gerard; A Svoronos; S T Herman; R Mani; H Arif; N Jette; Y Minazad; J F Kerrigan; P Vespa; S Hantus; J Claassen; G B Young; E So; P W Kaplan; M R Nuwer; N B Fountain; F W Drislane
Journal:  J Clin Neurophysiol       Date:  2013-02       Impact factor: 2.177

4.  Continuous electroencephalogram monitoring in the ICU or how to bring the inflation under control.

Authors:  Nicolas Adam; Eric Magalhaes; Tarek Sharshar
Journal:  Crit Care Med       Date:  2013-04       Impact factor: 7.598

5.  Continuous EEG monitoring in comatose intensive care patients: epileptiform activity in etiologically distinct groups.

Authors:  G Bryan Young; Gordon S Doig
Journal:  Neurocrit Care       Date:  2005       Impact factor: 3.210

6.  Continuous electroencephalography monitoring for early prediction of neurological outcome in postanoxic patients after cardiac arrest: a prospective cohort study.

Authors:  Marleen C Cloostermans; Fokke B van Meulen; Carin J Eertman; Harold W Hom; Michel J A M van Putten
Journal:  Crit Care Med       Date:  2012-10       Impact factor: 7.598

7.  Prevalence of nonconvulsive status epilepticus in comatose patients.

Authors:  A R Towne; E J Waterhouse; J G Boggs; L K Garnett; A J Brown; J R Smith; R J DeLorenzo
Journal:  Neurology       Date:  2000-01-25       Impact factor: 9.910

Review 8.  Continuous electroencephalogram monitoring in the intensive care unit.

Authors:  Daniel Friedman; Jan Claassen; Lawrence J Hirsch
Journal:  Anesth Analg       Date:  2009-08       Impact factor: 5.108

Review 9.  EEG in postanoxic coma: Prognostic and diagnostic value.

Authors:  J Hofmeijer; M J A M van Putten
Journal:  Clin Neurophysiol       Date:  2016-02-11       Impact factor: 3.708

10.  Novel wireless electroencephalography system with a minimal preparation time for use in emergencies and prehospital care.

Authors:  Andrei Jakab; Antti Kulkas; Timo Salpavaara; Pasi Kauppinen; Jarmo Verho; Hannu Heikkilä; Ville Jäntti
Journal:  Biomed Eng Online       Date:  2014-05-08       Impact factor: 2.819

View more
  2 in total

1.  Continuous EEG monitoring by a new simplified wireless headset in intensive care unit.

Authors:  Anselmo Caricato; Giacomo Della Marca; Eleonora Ioannoni; Serena Silva; Tiziana Benzi Markushi; Eleonora Stival; Daniele Guerino Biasucci; Nicola Montano; Camilla Gelormini; Isabella Melchionda
Journal:  BMC Anesthesiol       Date:  2020-12-07       Impact factor: 2.217

Review 2.  Noninvasive mobile EEG as a tool for seizure monitoring and management: A systematic review.

Authors:  Andrea Biondi; Viviana Santoro; Pedro F Viana; Petroula Laiou; Deb K Pal; Elisa Bruno; Mark P Richardson
Journal:  Epilepsia       Date:  2022-03-27       Impact factor: 6.740

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.