Evaluation of three VMAT-TMI planning methods to find an appropriate balance between plan complexity and the resulting dose distribution.

Joanna Litoborska1, Tomasz Piotrowski2, Julian Malicki3.   

Abstract

PURPOSE: Evaluation of different planning methods of treatment plan preparation for volumetric modulated arc therapy during total marrow irradiation (VMAT-TMI).
METHOD: Three different planning methods were evaluated to establish the most appropriate VMAT-TMI technique, based on organ at risk (OAR) dose reduction, conformity and plan simplicity. The methods were: (M1) the sub-plan method, (M2) use of eight arcs optimised simultaneously and (M3) M2 with monitor unit reduction. Friedman ANOVA comparison, with Nemenyi's procedures, was used in the statistical analysis of the results.
RESULTS: The dosimetric results obtained for the planning target volume and for most OARs do not differ statistically between methods. The M3 method was characterized by the lowest numbers of monitor units (3259 MU vs. 4450 MU for M1 and 4216 MU for M2) and, in general, the lowest complexity. The variability of the monitor units from control points was almost half for M3 than M1 and M2 (i.e. 0.33 MU vs. 0.61 MU for M1 and 0.58 for M2). Analysing the relationship between the dose distributions obtained for the plans and their complexity, the best result was observed for the M3 method.
CONCLUSION: The use of eight simultaneously optimised arcs with MU reduction allows to obtain VMAT-TMI plans that are characterized by the lowest complexity, with dose distributions comparable to the plans generated by other methods.
Copyright © 2020 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Keywords:  Plan optimization; Radiotherapy; Total marrow irradiation; Treatment plan complexity; Volumetric modulated arc therapy

Year:  2020        PMID: 32480353     DOI: 10.1016/j.ejmp.2020.05.018

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  The impact of different optimization strategies on the agreement between planned and delivered doses during volumetric modulated arc therapy for total marrow irradiation.

Authors:  Joanna Litoborska; Tomasz Piotrowski; Marta Kruszyna-Mochalska; Julian Malicki
Journal:  Contemp Oncol (Pozn)       Date:  2021-07-01

2.  Preclinical validation and treatment of volumetric modulated arc therapy based total bone marrow irradiation in Halcyon™ ring gantry linear accelerator.

Authors:  Tanweer Shahid; Sourav Mandal; Subhra Snigdha Biswal; Arundhati De; Mukti Mukherjee; Sandipan Roy Chowdhury; Anupam Chakrapani; Kirubha George; Jibak Bhattacharya; Prosenjit Soren; Tanmoy Ghosh; Biplab Sarkar; Luca Cozzi
Journal:  Radiat Oncol       Date:  2022-08-19       Impact factor: 4.309

3.  Gonadal-sparing total body irradiation with the use of helical tomotherapy for nonmalignant indications.

Authors:  Khaled Dibs; Austin J Sim; José A Peñagaricano; Kujtim Latifi; Genevieve A Garcia; Julia A Peters; Michael L Nieder; Sungjune Kim; Timothy J Robinson
Journal:  Rep Pract Oncol Radiother       Date:  2021-02-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.