Neeraj Kulkarni1, Deepti Pinto Rosario1, Shruthi Akkal2, Manisha Madhai Beck1, Liji Sarah David1, Reeta Vijayaselvi1, Bhaskar Mohan Murari2.
Abstract
KEY MESSAGE: The new NB scope aids in better visualization of the scalp and blood collection and analysis at bed side.
OBJECTIVE: Caesarean section rates and inherent complications are on the rise all over the world. One way to avoid a caesarean is to measure fetal scalp blood lactate levels. The methods available to visualize fetal scalp, obtain the blood sample and perform the blood test are separate, cumbersome and expensive, needing a certain level of expertise. We propose a device that incorporates all the steps of obtaining a fetal scalp blood lactate into one sleek, easy to use device.
METHODS: The initial design, 3-D print and was tried on mannequin. After ethics committee approval, the prototype was experimented on patients in labour with singleton live fetus in cephalic presentation with no evidence of distress.
RESULTS: There were (n = 9) patients recruited. There were (n = 5) primigravida and (n = 4) multigravida all of whom were in active labour. Parity did not seem to influence ease of instrumentation. Of the (n = 9) mothers (n = 2) had meconium-stained liquor and the rest (n = 7) had clear liquor, meconium-stained liquor did not affect visualization. The mean time taken to collect the sample was 184.11(± 33.04) seconds.
CONCLUSION: The Neeraj-Bhaskar (NB) scope is an easy to use, affordable device that can be used time and again to decide on cases where emergency caesarean section can be avoided due to fetal distress. © Federation of Obstetric & Gynecological Societies of India 2020.
KEY MESSAGE: The new NB scope aids in better visualization of the scalp and blood collection and analysis at bed side.
OBJECTIVE: Caesarean section rates and inherent complications are on the rise all over the world. One way to avoid a caesarean is to measure fetal scalp blood lactate levels. The methods available to visualize fetal scalp, obtain the blood sample and perform the blood test are separate, cumbersome and expensive, needing a certain level of expertise. We propose a device that incorporates all the steps of obtaining a fetal scalp blood lactate into one sleek, easy to use device.
METHODS: The initial design, 3-D print and was tried on mannequin. After ethics committee approval, the prototype was experimented on patients in labour with singleton live fetus in cephalic presentation with no evidence of distress.
RESULTS: There were (n = 9) patients recruited. There were (n = 5) primigravida and (n = 4) multigravida all of whom were in active labour. Parity did not seem to influence ease of instrumentation. Of the (n = 9) mothers (n = 2) had meconium-stained liquor and the rest (n = 7) had clear liquor, meconium-stained liquor did not affect visualization. The mean time taken to collect the sample was 184.11(± 33.04) seconds.
CONCLUSION: The Neeraj-Bhaskar (NB) scope is an easy to use, affordable device that can be used time and again to decide on cases where emergency caesarean section can be avoided due to fetal distress. © Federation of Obstetric & Gynecological Societies of India 2020.
Entities:
Keywords:
Fetal scalp blood sampling; Fetal scalp lactate; Fetal scalp visualization; NB scope
Year: 2020
PMID: 32476767 PMCID: PMC7239983 DOI: 10.1007/s13224-020-01313-9
Source DB: PubMed Journal: J Obstet Gynaecol India ISSN: 0975-6434