Anders Brostrøm1,2, Kirsten Inga Kling1,3, Karin Sørig Hougaard2, Kristian Mølhave1.
Abstract
Scanning electron microscopy, coupled with energy-dispersive X-ray spectroscopy (EDS), is a powerful tool used in many scientific fields. It can provide nanoscale images, allowing size and morphology measurements, as well as provide information on the spatial distribution of elements in a sample. This study compares the capabilities of a traditional EDS detector with a recently developed annular EDS detector when analyzing electron transparent and beam-sensitive NaCl particles on a TEM grid. The optimal settings for single particle analysis are identified in order to minimize beam damage and optimize sample throughput via the choice of acceleration voltage, EDS acquisition time, and quantification model. Here, a linear combination of two models is used to bridge results for particle sizes, which are neither bulk nor sufficiently thin to assume electron transparent. Additionally, we show that the increased count rate obtainable with the annular detector enables mapping as a viable analysis strategy compared with feature detection methods, which only scan segmented regions. Finally, we discuss advantages and disadvantages of the two analysis strategies.Entities:
Keywords: aerosol; electron microscopy; energy-dispersive X-ray spectroscopy; particle
Year: 2020 PMID: 32475372 DOI: 10.1017/S1431927620001464