| Literature DB >> 32466323 |
Nuno P Silva1,2, Anna Bottiglieri1, Raquel C Conceição3, Martin O'Halloran1, Laura Farina1,4.
Abstract
Electromagnetic-based hyperthermic therapies induce a controlled increase of temperature in a specific tissue target in order to increase the tissue perfusion or metabolism, or even to induce cell necrosis. These therapies require accurate knowledge of dielectric and thermal properties to optimise treatment plans. While dielectric properties have been well investigated, only a few studies have been conducted with the aim of understanding the changes of thermal properties as a function of temperature; i.e., thermal conductivity, volumetric heat capacity and thermal diffusivity. In this study, we experimentally investigate the thermal properties of ex vivo ovine liver in the hyperthermic temperature range, from 25 °C to 97 °C. A significant increase in thermal properties is observed only above 90 °C. An analytical model is developed to model the thermal properties as a function of temperature. Thermal properties are also investigated during the natural cooling of the heated tissue. A reversible phenomenon of the thermal properties is observed; during the cooling, thermal properties followed the same behaviour observed in the heating process. Additionally, tissue density and water content are evaluated at different temperatures. Density does not change with temperature; mass and volume losses change proportionally due to water vaporisation. A 30% water loss was observed above 90 °C.Entities:
Keywords: density; electromagnetic-based therapies; hyperthermia; thermal ablation; thermal properties; water loss
Mesh:
Year: 2020 PMID: 32466323 PMCID: PMC7285484 DOI: 10.3390/s20103004
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Experimental setup: TEMPOS thermal properties analyser, metallic container, thermal bath and fibre optic temperature sensor [33].
Figure 2Sketch of the container used to hold the samples, with the related dimensions. Left: The dual-needle SH-3 sensor from the TEMPOS thermal properties analyser is sketched. The sensor was placed at least 15 mm from the border of the container, according to the manufacturer’s specification. A fibre optic temperature sensor used to monitor the temperature during the experiment is also shown. Right: The container’s lid is sketched together with the holes dedicated to the insertion of the SH-3 sensor and the fibre optic temperature sensor [33].
Thermal properties (i.e., thermal conductivity, k, volumetric heat capacity, , and thermal diffusivity, D) for each ex vivo ovine liver sample (n) at different increasing temperatures (average values, avg). The associated uncertainty of each thermal property is given by .
|
| ||||||||
|---|---|---|---|---|---|---|---|---|
| avg |
| avg |
| avg |
| avg |
| |
|
| 25.35 | 0.10 | 0.50 | 0.06 | 3.39 | 0.40 | 0.15 | 0.02 |
|
| 26.93 | 0.76 | 0.48 | 0.06 | 3.32 | 0.39 | 0.15 | 0.02 |
|
| 35.68 | 0.03 | 0.46 | 0.05 | 3.23 | 0.37 | 0.14 | 0.02 |
|
| 36.27 | 0.12 | 0.49 | 0.06 | 3.25 | 0.41 | 0.15 | 0.02 |
|
| 39.55 | 0.11 | 0.53 | 0.06 | 3.52 | 0.41 | 0.15 | 0.02 |
|
| 40.24 | 0.07 | 0.53 | 0.06 | 3.52 | 0.41 | 0.15 | 0.02 |
|
| 44.11 | 0.33 | 0.49 | 0.06 | 3.42 | 0.41 | 0.14 | 0.02 |
|
| 44.58 | 0.09 | 0.52 | 0.06 | 3.50 | 0.41 | 0.15 | 0.02 |
|
| 48.87 | 0.22 | 0.53 | 0.06 | 3.52 | 0.41 | 0.15 | 0.02 |
|
| 49.49 | 0.18 | 0.53 | 0.06 | 3.55 | 0.41 | 0.15 | 0.02 |
|
| 55.05 | 0.06 | 0.46 | 0.05 | 3.33 | 0.39 | 0.14 | 0.02 |
|
| 55.37 | 0.14 | 0.48 | 0.06 | 3.38 | 0.39 | 0.14 | 0.02 |
|
| 59.59 | 0.21 | 0.53 | 0.06 | 3.42 | 0.43 | 0.16 | 0.02 |
|
| 60.34 | 0.17 | 0.54 | 0.06 | 3.53 | 0.42 | 0.15 | 0.02 |
|
| 70.37 | 0.26 | 0.55 | 0.06 | 3.50 | 0.41 | 0.16 | 0.02 |
|
| 70.52 | 0.46 | 0.54 | 0.06 | 3.53 | 0.42 | 0.15 | 0.02 |
|
| 79.49 | 0.44 | 0.54 | 0.06 | 3.42 | 0.40 | 0.16 | 0.02 |
|
| 80.46 | 0.25 | 0.56 | 0.07 | 3.37 | 0.39 | 0.17 | 0.02 |
|
| 91.97 | 0.52 | 0.61 | 0.07 | 3.41 | 0.41 | 0.18 | 0.02 |
|
| 92.03 | 0.16 | 0.58 | 0.07 | 3.55 | 0.41 | 0.16 | 0.02 |
|
| 96.79 | 0.15 | 1.08 | 0.16 | 5.05 | 0.89 | 0.23 | 0.05 |
Figure 3Thermal properties of ex vivo ovine liver as a function of the increasing temperature: (a) thermal conductivity, (b) thermal diffusivity and (c) volumetric heat capacity. Average values are reported with the associated uncertainty (error bars). Experimental data (red) as well as literature data (blue) are shown. The best fit model is also presented by a dashed line. The estimated error of how well the least square model fits the data is reported in (d).
Regression coefficients and R-squared of fit function modelling each thermal property with temperature.
| Thermal Property |
|
|
|
|
|---|---|---|---|---|
| Thermal conductivity, | 0.502 | 1.447 | 0.256 | 0.93 |
| Volumetric heat capacity, | 3.415 | 1.278 | 0.289 | 0.95 |
| Thermal diffusivity, | 0.150 | 1.379 | 0.164 | 0.81 |
Figure 4Thermal properties of ex vivo ovine liver as a function of the decreasing temperature: (a) thermal conductivity, (b) thermal diffusivity and (c) volumetric heat capacity. Raw data measured during the cooling of each sample are reported; their nominal associated uncertainty is 10%, which corresponds to the device accuracy. The best fit model described in Table 2 is also reported by a dashedline. The estimated error of how well the least square model fits the data is reported in (d).