[Is high-density polyethylene suitable as an implant material in cement-free anchoring of hip endoprostheses? A histomorphologic study of an explanted polyethylene screw-in acetabula].

F Lintner1, G Böhm, P Bösch, G Brand, M Endler, K Zweymüller.   

Abstract

Firmly attached screw-in polyethylene acetabula which had been implanted for between 16 and 54 months were explanted after autopsies and subjected to macroscopic, radiologic and histologic examination. Metaplasias were seen around the threads and on the floor of the acetabulum. Their tendency to ossify represents an attempt at secondary stabilization. Due to the low stability of the polyethylene this causes increased wear on the floor of the acetabulum. The small defects in the polyethylene found in the threads, resembling damage done by mice, may be a sign of biodegradation. In view of the tissue reactions pointed out, the material stability of the polyethylene needs to be improved or implantation must be restricted to a very limited range of indications.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3245287     DOI: 10.1055/s-2008-1044509

Source DB:  PubMed          Journal:  Z Orthop Ihre Grenzgeb        ISSN: 0044-3220


  3 in total

1.  Long-term reaction of the osseous bed around silicone implants.

Authors:  A Wanivenhaus; F Lintner; C Wurnig; M Missaghi-Schinzl
Journal:  Arch Orthop Trauma Surg       Date:  1991       Impact factor: 3.067

2.  Implant materials for hip endoprostheses: old proofs and new trends.

Authors:  M Semlitsch; H G Willert
Journal:  Arch Orthop Trauma Surg       Date:  1995       Impact factor: 3.067

3.  Bone reaction to uncemented threaded polyethylene acetabular components.

Authors:  J Krugluger; R Eyb
Journal:  Int Orthop       Date:  1993       Impact factor: 3.075

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.