| Literature DB >> 32445773 |
Ralph A C van Wezel1, Anne C T Vrancken1, Marieke Ernest1, Johan Laurensse1, Josephus P C M van Doornmalen Gomez Hoyos2.
Abstract
AIM: To develop a protocol to ensure the quality of respiratory protective devices for healthcare workers nursing and treating patients with possible or confirmed COVID-19 in the Catharina hospital.Entities:
Keywords: COVID-19; Corona; Respirators; particle counting; patient safety; staff safety
Year: 2020 PMID: 32445773 PMCID: PMC7239022 DOI: 10.1016/j.jhin.2020.05.023
Source DB: PubMed Journal: J Hosp Infect ISSN: 0195-6701 Impact factor: 3.926
Figure 1Flow chart of data acquisition for inward leakage of the respirators.
Figure 2Schematic overview of testing respirator efficacy for filtering particles. The set-up includes a metal channel with a total length of 2.30 m and a diameter of 0.16 m (Figure 3). The open end (left) of the tube has free access to the environment. At the closed end (right) a vacuum pump is installed. The green vertical bar represents a mould in which the respirator to be tested is placed (Figure 3c). The particles coming with the flow into the channel are measured at two locations with two separate particle counters (Figure 3d), one before and one after the location of the respirator. Additional to the iso-kinetic measurements system of the particle counters, the sampling for particle counting is performed via a horizontally placed Pitot tube. This ensures that the sampling is in the axial opposite direction of the air flow. The black dots between the mould and channel represent a seal to exclude air leaks into the channel. To exclude the possibility of accumulating static electric charge in the channel wall, the metal channel is grounded to earth. All equipment used in the test-rig was available in the technical department of the hospital.
Figure 3Test-rig for respirator testing. (a) Overview of test-rig. At the left end of the channel is the open end of the environment; at the right end, the hose to the vacuum pump. (b) Front view of test-rig with a respirator loaded in channel for testing. (c) Example of a respirator in a dedicated mould. (d) Particle counters, dirty and clean side.
Figure 4Data from 13 tested respirators in the test-rig (Figure 2, Figure 3): averaged values of the test sessions per particle size (Figure 1). In every session the measured particle sizes were 0.3, 0.5, 1.0, 5.0, and 10.0 μm. ‘ref’ stands for reference: the reference values of an A-brand FFP2 respirator. The horizontal magenta lines indicate the maximum particle values for FFP1, FFP2, and FFP3 respirators, according to the standard: 22%, 8%, and 2%, respectively [1]. Because the reference stays within the limit of the FFP2 respirator, this limit is used as the acceptance criterion for the full filter test. Based on these results the respirators D, E, G, F, H, J, and L meet the criteria. The brands and types are not disclosed because it does not add value in the development of the test protocol for respirators.
Example of a processed data set from a measurement
| Item | 0.3 μm | 0.5 μm | 1 μm | 3 μm | 5 μm | 10 μm |
|---|---|---|---|---|---|---|
| Zero measurement before | 0.75 | 0.92 | 0.91 | 0.55 | 0.58 | 1 |
| 0.75 | 0.92 | 0.96 | 0.83 | 0.60 | 0.60 | |
| 0.77 | 0.95 | 0.90 | 0.55 | 1.18 | 0.29 | |
| 0.77 | 0.90 | 0.90 | 0.53 | 0.52 | 0.86 | |
| 0.79 | 0.95 | 0.89 | 0.56 | 0.58 | 0.40 | |
| Respirator P-1 | 10.59 | 4.57 | 2.17 | 0 | 0 | 0 |
| 10.46 | 4.12 | 1.90 | 0 | 0 | 0 | |
| 9.94 | 4.20 | 0.69 | 0.80 | 0 | 0 | |
| 10.04 | 3.88 | 1.52 | 0.91 | 0 | 0 | |
| 9.60 | 4.02 | 1.87 | 1.12 | 0 | 0 | |
| Respirator P-2 | 10.48 | 5.07 | 2.57 | 0 | 0 | 0 |
| 10.48 | 4.22 | 2.61 | 0 | 0 | 0 | |
| 10.76 | 4.50 | 1.71 | 1.06 | 0 | 0 | |
| 10.39 | 4.68 | 0.76 | 0 | 0 | 0 | |
| 10.90 | 5.04 | 1.58 | 0 | 0 | 0 | |
| Respirator P-3 | 10.01 | 4.11 | 1.85 | 0 | 0 | 0 |
| 9.81 | 3.96 | 1.76 | 0 | 0 | 0 | |
| 10.28 | 4.76 | 2.77 | 0 | 0 | 0 | |
| 10.09 | 4.08 | 2.31 | 1.29 | 0 | 0 | |
| 9.50 | 3.93 | 1.58 | 0 | 0 | 0 | |
| Respirator P-4 | 15.56 | 8.13 | 4.64 | 1.12 | 1.25 | 0 |
| 14.55 | 6.88 | 4.24 | 0 | 0 | 0 | |
| 13.89 | 6.84 | 2.98 | 1.72 | 0 | 0 | |
| 13.36 | 6.20 | 3.55 | 0 | 0 | 0 | |
| 13.36 | 6.47 | 2.34 | 0.65 | 0 | 51.06 | |
| Respirator P-5 | 15.45 | 10.11 | 4.88 | 1.26 | 3.18 | 0 |
| 14.47 | 8.88 | 4.17 | 4.45 | 3.82 | Div.0! | |
| 13.82 | 7.45 | 3.89 | 5.25 | 0 | 0 | |
| 14.16 | 7.74 | 2.91 | 0 | 0 | 0 | |
| 13.38 | 7.55 | 2.77 | 1.44 | 0 | 0 | |
| Zero measurement after | 0.80 | 0.91 | 0.82 | 0.64 | 1 | 0.38 |
| 0.80 | 0.90 | 0.78 | 0.77 | 0.41 | 0.33 | |
| 0.81 | 0.90 | 0.82 | 0.91 | 0.58 | 0.50 | |
| 0.82 | 0.90 | 0.89 | 0.65 | 0.29 | 0.75 | |
| 0.81 | 0.90 | 0.83 | 0.94 | 0.22 | 1.50 | |
| Correction factor before | 0.76 | 0.93 | 0.91 | 0.61 | 0.69 | 0.63 |
| Correction factor after | 0.81 | 0.90 | 0.83 | 0.78 | 0.50 | 0.69 |
| Difference | –0.05 | 0.03 | 0.08 | –0.17 | 0.19 | –0.06 |
In this example the acceptable level of particles is 8%. The values for the zero measurements are ratios (Equation 1). The values for the respirator measurements are percentages (Equation 2).
Values >8% are assessed as fail. In the calculated correction factors the change in air composition can be verified.
Because of the ratio a value may result in Division by zero (Div.0!) when no particles are counted. This was typically observed with the larger particles (5.0 and 10 μm).