Abutment screw torque changes with straight and angled screw-access channels.

Rajesh S Swamidass1, Joseph Y K Kan2, Mathew T Kattadiyil3, Charles J Goodacre4, Jaime Lozada5.   

Abstract

STATEMENT OF PROBLEM: Angle-correcting options allow the use of screw-retained implant prostheses in situations where an implant has been placed with a facial inclination. However, manufacturers have different recommended torque values, and it is unclear whether the performance of these designs is equivalent to that of the traditional screw-retained crowns (SRCs) when subjected to cyclic loading forces.
PURPOSE: The purpose of this in vitro study was to compare torque differences between conventional straight-line screw access and angulated access SRCs before and after simulated functional loading.
MATERIAL AND METHODS: Five groups consisting of 10 SRCs and implants were formed: Nobel Biocare zirconia crowns with 20-degree access channels (NB-20); Dynamic Abutment Solution zirconia crowns (DA-20) with 20-degree access channels; Core3dcentre angle correction zirconia crowns with 20-degree access channels (C3D-20); Nobel Biocare zirconia crowns with 0-degree access channels (NB-0); and gold alloy crowns cast to Nobel Biocare Gold-Adapt abutments (GA-0). Each specimen underwent thermocycling before cyclic loading. A preload torque based on the manufacturer's recommendation was applied to each crown placed on an implant. Reverse torque measurements were obtained for each specimen before cyclic loading. Each implant-abutment assembly was then cyclic loaded at 0 to 100 N at 10 Hz for 1 million cycles. Reverse torque measurements were obtained after cyclic loading and the percentage difference calculated.
RESULTS: No significant percentage torque loss differences were observed between the 0-degree and 20-degree SRCs after cyclic loading. No significant differences were seen among the angulated access channel crowns. DA-20 and C3D-20 specimens had significantly higher torque loss compared with the NB-0 group. The C3D-20 group reported the largest percentage torque loss (34.5%) among the angulated access screw channel groups. The GA-0 group reported the largest percentage torque loss of all the groups (35.9%). No crown mobility or other complications were observed in any of the groups after cyclic loading.
CONCLUSIONS: Angulated access channel crowns performed comparably with conventional straight-line screw access SRCs with regard to percentage torque values after cyclic loading. Angulated access channel crowns with lower manufacturer recommended torque values had higher percentage torque differences.
Copyright © 2020 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2020        PMID: 32439126     DOI: 10.1016/j.prosdent.2020.01.018

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  2 in total

1.  Management of Buccal Screw Access Hole Positioning for Implant Fixed Complete Dentures: A Report of Two Patients and a Proposed Decision Tree.

Authors:  Fawaz Alzoubi
Journal:  Case Rep Dent       Date:  2021-12-21

2.  Aesthetic Outcomes and Peri-Implant Health of Angled Screw Retained Implant Restorations Compared with Cement Retained Crowns: Medium Term Follow-Up.

Authors:  Livia Nastri; Ludovica Nucci; Vincenzo Grassia; Rino Miraldi
Journal:  J Funct Biomater       Date:  2021-05-16
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.