Literature DB >> 32436457

Woven EndoBridge Device immediate post-detachment tilt with later displacement: Case report, rescue techniques, and review of the literature.

Fortunato Di Caterino1, Panagiotis Primikiris1, Giovanni Vitale1, Alessandra Biondi1.   

Abstract

The Woven EndoBridge Device (WEB) is efficient and safe in the treatment of wide-neck bifurcation intracranial aneurysms. A crucial step in operative planning is establishing the appropriate dimension of the device, and there are limitations to the therapeutic solutions provided by WEBs. We describe a case of a right middle cerebral artery bifurcation wide-neck aneurysm treated with WEB. After the initial deployment of a WEB proven oversized, we substituted it with a smaller one that presented immediate post-detachment intra-aneurysmal tilt probably resulting from undersizing. The 24-h angiographic control identified a partial displacement of the device in the superior middle cerebral artery branch. We describe rescue techniques with review of the literature and our treatment strategy, including effort to reposition the device followed by stent deployment. The treatment was electively completed by the coil-through technique. The post-detachment WEB tilt should be considered an unstable position and treated either by removal of the device or with adjunctive implants. Each WEB size adapts to a range of aneurysmal height and width. This range is not always the same for each specific WEB dimension, based on the table provided by the manufacturer. This eventually predisposes to a different behavior of different WEB sizes regarding the modification of the device's height in relation to the modification of the diameter after deployment.

Entities:  

Keywords:  Woven EndoBridge Device; dislocation; intracranial aneurysm; rescue techniques

Mesh:

Year:  2020        PMID: 32436457      PMCID: PMC7446579          DOI: 10.1177/1591019920926334

Source DB:  PubMed          Journal:  Interv Neuroradiol        ISSN: 1591-0199            Impact factor:   1.610


  10 in total

1.  Should indications for WEB aneurysm treatment be enlarged? Report of a series of 20 patients with aneurysms in "atypical" locations for WEB treatment.

Authors:  L Pierot; A Biondi; A-P Narata; C Mihalea; A-C Januel; G Metaxas; R Bibi; J Caroff; S Soize; C Cognard; L Spelle; D Herbreteau
Journal:  J Neuroradiol       Date:  2017-03-03       Impact factor: 3.447

2.  Letter to the editor.

Authors:  Laurent Spelle; Thomas Liebig
Journal:  Neuroradiol J       Date:  2014-06-17

3.  Treatment of Wide-Neck Intracranial Aneurysms with the Woven EndoBridge Device Associated with Stenting: A Single-Center Experience.

Authors:  F Cagnazzo; R Ahmed; C Dargazanli; P-H Lefevre; G Gascou; I Derraz; S A Kalmanovich; C Riquelme; A Bonafe; V Costalat
Journal:  AJNR Am J Neuroradiol       Date:  2019-04-11       Impact factor: 3.825

4.  Rescue Maneuver for Dislocated Woven EndoBridge Device in Middle Cerebral Artery.

Authors:  Seby John; Ramon Navarro; Syed Irteza Hussain; Khalil Zahra; Mohammed Samy Elhammady
Journal:  World Neurosurg       Date:  2019-06-26       Impact factor: 2.104

5.  Safety and Efficacy of Aneurysm Treatment with the WEB: Results of the WEBCAST 2 Study.

Authors:  L Pierot; I Gubucz; J H Buhk; M Holtmannspötter; D Herbreteau; L Stockx; L Spelle; J Berkefeld; A-C Januel; A Molyneux; J V Byrne; J Fiehler; I Szikora; X Barreau
Journal:  AJNR Am J Neuroradiol       Date:  2017-04-27       Impact factor: 3.825

6.  Safety and efficacy of aneurysm treatment with WEB: results of the WEBCAST study.

Authors:  Laurent Pierot; Vincent Costalat; Jacques Moret; Istvan Szikora; Joachim Klisch; Denis Herbreteau; Markus Holtmannspötter; Werner Weber; Anne-Christine Januel; Thomas Liebig; Vojtech Sychra; Christoph Strasilla; Christophe Cognard; Alain Bonafé; Andrew Molyneux; James V Byrne; Laurent Spelle
Journal:  J Neurosurg       Date:  2015-09-18       Impact factor: 5.115

Review 7.  Endovascular techniques for the management of wide-neck intracranial bifurcation aneurysms: A critical review of the literature.

Authors:  Laurent Pierot; Alessandra Biondi
Journal:  J Neuroradiol       Date:  2016-03-11       Impact factor: 3.447

8.  Woven Endobridge (WEB) device for endovascular treatment of complex unruptured aneurysms-a single center experience.

Authors:  Stephanie Lescher; Richard du Mesnil de Rochemont; Joachim Berkefeld
Journal:  Neuroradiology       Date:  2016-01-15       Impact factor: 2.804

9.  Dislocation of a WEB Device into the Middle Cerebral Artery : Removal with the Alligator Retrieval Device.

Authors:  I König; A Weber; W Weber; S Fischer
Journal:  Clin Neuroradiol       Date:  2018-05-11       Impact factor: 3.649

10.  Clinical and Anatomical Follow-up in Patients With Aneurysms Treated With the WEB Device: 1-Year Follow-up Report in the Cumulated Population of 2 Prospective, Multicenter Series (WEBCAST and French Observatory).

Authors:  Laurent Pierot; Laurent Spelle; Andrew Molyneux; James Byrne
Journal:  Neurosurgery       Date:  2016-01       Impact factor: 4.654

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.