Biomechanical Response under Stress-Controlled Tension-Tension Fatigue of a Novel Carbon Fiber/Epoxy Intramedullary Nail for Femur Fractures.

Saeid Samiezadeh1, Emil H Schemitsch2, Radovan Zdero3, Habiba Bougherara4.   

Abstract

Metallic intramedullary nails are the "gold standard" implant for repairing femur shaft fractures. However, their rigidity may eliminate axial micromotion at the fracture (causing delayed healing) and they may carry too much load relative to the femur (causing "stress shielding"). Consequently, some researchers have proposed fiber-reinforced composite nails, but only one evaluated cyclic fatigue performance. Therefore, this study assessed the cyclic fatigue response of a carbon fiber/epoxy nail with a novel ply stacking sequence of [02/-45/452/-45/0/-45/452/-452/452/-45/902] previously developed by the present authors. Nails were cyclically loaded in tension-tension at 5 Hz with a stress ratio of R=0.1 from 30% - 85% of the material's ultimate tensile strength (UTS). Thermographic stress analysis, rather than conventional fatigue testing, was used to obtain high cycle fatigue strength (HCFS), below which the nail can be cyclically loaded indefinitely without damage. Also, the mechanical test machine's built-in load cell and an extensometer were used to create stress-strain curves, from which the change in static EO and dynamic E* moduli were obtained. Results showed that HCFS was 70.3% of UTS (or about 283 MPa), while EO and E* remained at 42 GPa without any dRegradation during testing. The current nail shows potential for clinical use.
Copyright © 2020 IPEM. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Biomechanics; Composite; Fatigue; Femur; Intramedullary nail

Year:  2020        PMID: 32430231     DOI: 10.1016/j.medengphy.2020.04.001

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  1 in total

1.  From Bench to Bedside: Radiolucent Implants-Better Visualization or Camouflaged Gimmick?

Authors:  Benjamin K Potter
Journal:  Clin Orthop Relat Res       Date:  2022-03-01       Impact factor: 4.755

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.