Tomi Ryynänen1, Ropafadzo Mzezewa2, Ella Meriläinen1, Tanja Hyvärinen2, Jukka Lekkala1, Susanna Narkilahti2, Pasi Kallio1.
Abstract
Microelectrode array (MEA) is a tool used for recording bioelectric signals from electrically active cells in vitro. In this paper, ion beam assisted electron beam deposition (IBAD) has been used for depositing indium tin oxide (ITO) and titanium nitride (TiN) thin films which are applied as transparent track and electrode materials in MEAs. In the first version, both tracks and electrodes were made of ITO to guarantee full transparency and thus optimal imaging capability. In the second version, very thin (20 nm) ITO electrodes were coated with a thin (40 nm) TiN layer to decrease the impedance of Ø30 µm electrodes to one third (1200 kΩ 320 kΩ) while maintaining (partial) transparency. The third version was also composed of transparent ITO tracks, but the measurement properties were optimized by using thick (200 nm) opaque TiN electrodes. In addition to the impedance, the optical transmission and electric noise levels of all three versions were characterized and the functionality of the MEAs was successfully demonstrated using human pluripotent stem cell-derived neuronal cells. To understand more thoroughly the factors contributing to the impedance, MEAs with higher IBAD ITO thickness as well as commercial sputter-deposited and highly conductive ITO were fabricated for comparison. Even if the sheet-resistance of our IBAD ITO thin films is very high compared to the sputtered one, the impedances of the MEAs of each ITO grade were found to be practically equal (e.g., 300-370 kΩ for Ø30 µm electrodes with 40 nm TiN coating). This implies that the increased resistance of the tracks, either caused by lower thickness or lower conductivity, has hardly any contribution to the impedance of the MEA electrodes. The impedance is almost completely defined by the double-layer interface between the electrode top layer and the medium including cells.Entities:
Keywords: indium tin oxide (ITO); ion beam assisted electron beam deposition (IBAD); microelectrode array (MEA); neurons; titanium nitride (TiN); transparent
Year: 2020 PMID: 32423145 PMCID: PMC7281740 DOI: 10.3390/mi11050497
Source DB: PubMed Journal: Micromachines (Basel) ISSN: 2072-666X Impact factor: 2.891
Details of microelectrode array (MEA) batches.
| MEA Batch | 1 | 2 | 3 | 4 |
|---|---|---|---|---|
| ITO thickness (nm) | 20 | 20 | 150 | 180 |
| ITO deposition method | IBAD | IBAD | IBAD | Sputtering |
Details of MEA versions.
| MEA Version | 1 | 2 | 3 |
|---|---|---|---|
| Track material | ITO | ITO | ITO |
| Electrode material | ITO | TiN on ITO | TiN on ITO |
| TiN thickness (nm) | - | 40 | 200 |
Figure 1(a) Sideview of the ion beam assisted electron beam deposited (IBAD) indium tin oxide (ITO) MEA structure. The image is not to scale. The structure is the same in each batch and version except that the thickness of the ITO track (grey) and/or its deposition method are varied in different batches and the existence or thickness of the TiN coating (green) is varied in different versions. See Table 1 and Table 2 for more details. (b) Photograph of an IBAD ITO MEA with a polydimethylsiloxane (PDMS) microfluidic device. The dimensions of the glass substrate are 49 mm × 49 mm.
Figure 2Four-wire test pattern for sheet resistance measurements.
Figure 3Transmission of IBAD deposited 20 nm ITO and 40 nm TiN films at visible wavelengths.
Impedances at 1 kHz and estimates of RMS noise level and sheet resistance. In the normalized column the impedances of 35 × 50 µm2 electrodes are normalized to correspond to the area of Ø30 µm electrodes.
| Batch/ | ITO Deposition Method | ITO Thickness (nm) | TiN Thickness (nm) | Measured | Measured | Normalized Impedance (kΩ) | RMS Noise (µV) | ITO Sheet Resistance |
|---|---|---|---|---|---|---|---|---|
| 1/1 | IBAD | 20 | - | 1200 ± 260 | 820 ± 140 | 2030 | 12 ± 1 | 2.6 × 103 |
| 2/1 | IBAD | 20 | - | 1950 ± 810 | 1170 ± 120 | 2900 | Na | 9.4 × 106 |
| 3/1 | IBAD | 150 | - | 1940 ± 290 | 1180 ± 150 | 2920 | Na | 76.4 × 103 |
| 4/1 | Sputtering | 180 | - | 1420 ± 170 | 830 ± 30 | 2050 | Na | 8–10 * |
| 2/1 | IBAD | 20 | 40 | 320 ± 20 | 190 ± 10 | 470 | 6.1 ± 0.5 | 2.6 × 103 |
| 2/2 | IBAD | 20 | 40 | 370 ± 10 | 150 ± 10 | 370 | na | 9.4 × 106 |
| 3/2 | IBAD | 150 | 40 | 350 ± 10 | 160 ± 3 | 400 | na | 76.4 × 103 |
| 4/2 | Sputtering | 180 | 40 | 300 ± 10 | 130 ± 2 | 320 | na | 8–10 * |
| 1/3 | IBAD | 20 | 200 | 190 ± 10 | 150 ± 20 | 370 | 5.6 ± 0.5 | 2.6 × 103 |
| 2/3 | IBAD | 20 | 200 | 240 ± 10 | 120 ± 2 | 300 | na | 9.4 × 106 |
| 3/3 | IBAD | 150 | 200 | 210 ± 4 | 110 ± 2 | 270 | na | 76.4 × 103 |
| 4/3 | Sputtering | 180 | 200 | 190 ± 3 | 90 ± 1 | 220 | na | 8–10 * |
* Data from supplier (University Wafer Inc.).
Figure 4Inverted microscope images of neuronal cell networks growing on different versions of batch 1 MEAs (a–c) and a control culture (d) grown at the same time on standard plastic plate (29 days after plating). (a) The cells are fully visible through the tracks and circular electrodes of MEA version 1. (b) In MEA version 2 the visibility through the electrodes drops to partial. However, the cells above the electrodes can still be seen (see especially the larger cluster on the grounding electrode, pointed out by the arrow). (c) No visibility through the electrodes in MEA version 3. The control image was taken with a different microscope than the MEA images.
Figure 5Examples of neuronal cell field potential recordings with different MEA versions from batch 1. Offsets of the curves have been shifted for clarity.