Device-Related Complications Associated with Magec Rod Usage for Distraction-Based Correction of Scoliosis.

Aakash Agarwal1, Amey Kelkar1, Ashish Garg Agarwal2, Daksh Jayaswal1, Arvind Jayaswal2, Vithal Shendge1.   

Abstract

INTRODUCTION: Recent literature identifies similar failure rates such as anchor pull-out and rod breakage, but a higher unplanned revision surgery with MAGEC rods than with traditional growth rods. Besides known failure modes such as rod fracture, infection, etc., failure to noninvasively distract the rods was cited as the main cause of such unplanned surgeries. The source of these data ranges from multicenter cohort studies to singular case series. These studies included explanted implants that had undergone failure in distraction mechanism, rod fracture, or infection, or had reached their maximum length. Nevertheless, in addition to identifying the overall mode of failure, it is equally important to identify the large-scale incidence of exclusive failures in comparison with standard instrumentation failure modes in spine surgery.
METHODS: The US Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) databases were searched for reports on MAGEC rods, and on standard instrumentation used for spinal fusion. The adverse events were recorded, tabulated, and analyzed.
RESULTS: A search of the US FDA MAUDE database yielded reports of 163 device-related adverse events. These included distraction mechanism failure (n=129), rod fracture (n=24), and minor voluntary reports of infection and tissue discoloration (n=10). For standard instrumentation usage in spine surgery, pedicle screw breakage post surgery (n=336), set screw damage during surgery (n=257), rod breakage post surgery (n=175), interbody cage breakage during surgery (n=118), and pedicle screw breakage during surgery (n=75) were identified as the top 5 failure modes.
CONCLUSIONS: The study identified the distraction mechanism failure as the most common and growing complication associated with MAGEC rod usage in children with scoliosis, leading to unplanned invasive revision surgeries.
Copyright © 2020 by The Japanese Society for Spine Surgery and Related Research.

Entities:  

Keywords:  MAGEC rods; complications; distraction; early onset scoliosis; rod fracture; traditional growth rods

Year:  2019        PMID: 32405561      PMCID: PMC7217671          DOI: 10.22603/ssrr.2019-0041

Source DB:  PubMed          Journal:  Spine Surg Relat Res        ISSN: 2432-261X


Introduction

The noninvasive distraction mechanism of MAGEC rods used to lengthen the spine in children with scoliosis have made traditional growth rods (an invasive distraction mechanism) a second choice, that is, used only for underprivileged patients or healthcare systems[1],[2]). Recent literature identifies similar failure rates such as anchor pull-out (11.8%) and rod breakage (10.6%), but a higher unplanned revision surgery with MAGEC rods than with traditional growth rods[3]). Besides known failure modes such as rod fracture, infection, etc., failure to noninvasively distract the rods (mostly due to breakage of actuator pin or reduced force generation) was cited as the main cause (11.7%-50%) of such unplanned surgeries[3],[4]). A common observation among these implants was the generation of titanium wear debris (leading to metallosis) due to off-axis loading[4],[5]). This is further escalated by failure of radial bearings and corrosion of underlying components. The source of these data ranges from multicenter cohort studies to singular case series[4-7]). These studies included explanted implants that had undergone failure in distraction mechanism, rod fracture, or infection, or had reached their maximum length. Nevertheless, in addition to identifying the overall mode of failure, it is equally important to identify the large-scale incidence reporting of exclusive failures in comparison with other standard failure modes in spine surgery. Therefore, the objective of the current study is to search the Manufacturer and User Facility Device Experience (MAUDE) database, which is a voluntary reporting system of adverse events involving medical devices since 1993, to identify the trend and frequency of adverse events related to the usage of MAGEC rods and standard instrumentation in spine surgery.

Materials and Methods

The United States Food and Drug Administration (FDA) of the Department of Human Services runs the MAUDE database. Adverse events are defined as potential and actual product use errors and product quality problems. The database contains information on medical devices that may have malfunctioned or caused a death or a serious injury. These adverse events are reported by the manufacturer or a healthcare professional (an operating room nurse or a surgeon). MAUDE may not include reports made according to exemptions, variances, or alternative reporting requirements granted under 21 Code of Federal Regulations 803.19. The FDA website was last accessed in June 2019, in preparation of data for this manuscript. The product code used was PGN (identifier for MAGEC rods), and all the categorial data based on event type were retrieved and analyzed to the date specified above. It is important to note that MAUDE data are not intended to be used either to evaluate absolute rates of adverse events (due to underreporting) or to compare adverse event occurrence rates between different devices (due to discrepancy in reporting between two devices). In addition to MAGEC rods, we also retrieved MAUDE data on the top 5 failure modes associated with standard instrumentation usage in spinal fusion; this does not include MAGEC rod failures, and may pertain to any possible diagnoses such as degenerative disc disease, spondylolisthesis, trauma, scoliosis, etc.

Results

A total of 163 incidences were recorded for MAGEC rod-related adverse events, until June 2019. Of these, 129 events were failure of noninvasive distraction mechanism, 24 were for rod fracture, and 10 for other medical complications such as infection, tissue necrosis, wear particles, etc., (Fig. 1). The top 5 failure modes associated with standard instrumentation usage in spinal fusion included pedicle screw breakage post surgery (n=336), set screw damage during surgery (n=257), rod breakage post surgery (n=175), interbody cage breakage during surgery (n=118), and pedicle screw breakage during surgery (n=75) (Fig. 2).
Figure 1.

Adverse event reporting from the MAUDE database for MAGEC rod usage until June 2019.

Figure 2.

Adverse event reporting from the MAUDE database for the top 5 failure modes associated with standard instrumentation usage in spinal fusion until June 2019.

Adverse event reporting from the MAUDE database for MAGEC rod usage until June 2019. Adverse event reporting from the MAUDE database for the top 5 failure modes associated with standard instrumentation usage in spinal fusion until June 2019.

Discussion

The current study presents all the MAUDE reporting for adverse events related to the usage of MAGEC rods in growing children with scoliosis[8]). Rod fracture is a known problem with all growth sparring systems; however, the failure to noninvasively distract the spine is exclusive to MAGEC rods only[9],[10]). MAGEC rod's main benefit is that it allows for the noninvasive distraction of the growth rods. This benefit is realized by a drastic reduction in the number of consecutive surgeries, and in some cases, the possibility of reducing growth rod fracture and autofusion (caused due to gradual trauma to the spine via excessive distraction in traditional growth rod systems)[11-16]). However, the failure of this attribute, noninvasive distraction mechanism, reduces the overall efficacy of the device with newer studies even questioning if there is a real quality-of-life difference with use of MAGEC rods[17]). The results of the current study show that the most frequent clinical problem associated with the MAGEC rod is the failure of noninvasive distraction mechanism, which leads to invasive revision surgeries required to replace the device. These results also highlight the exponential increase in such failure rates (Fig. 1). Furthermore, the MAUDE report on the top 5 failure modes associated with standard instrumentation usage in spinal fusion proves that such failures are substantially underreported. Although the MAGEC rod has not completely replaced the traditional growth rod system, due to it unaffordability in underprivileged patients, it is still much more advantageous than its predecessor. The purpose of this study and many other clinical trials, or case series, is only to improve upon the existing noninvasive technology by thoroughly identifying the failure modes. In the case of MAGEC rods, it is the failure of its noninvasive distraction mechanism that needs attention. Further immediate research and development work should be undertaken to reduce these failure incidences in a clinical scenario. The limitation of the current study is that the patient demographics, surgical parameters, and the total number of surgeries performed (using MAGEC rods) in the United States within this period is unknown. The voluntary nature of such reporting also presents a challenge against excluding infrequently reported complications such as necrosis, which although is present in all cases (off-axis loading and wear), has only been reported 10 times in total. Despite these limitations, the type of adverse events and relatively high frequency of these events when compared to the top 5 failure modes associated with standard instrumentation usage in spinal fusion are a matter of great concern. In conclusion, the result of this study suggests a growing number of distraction mechanism failure associated with MAGEC rods. Better technical and clinical controls need to be set in place to avoid such adverse events, which leads to unplanned open surgeries. A higher distraction magnitude results in the generation of higher distraction forces, and this in combination with off-axis loading (exemplified by “growth marks”) result in wear and breakage of MAGEC rod's components. Therefore, one method to reduce the propensity of such failures would be to apply minimum distraction at higher frequency, for example, 1.5-2.0 mm every month, instead of 4.5-6.0 mm every 3 months. This would also reduce tissue trauma and its effects, such as autofusion[11-16]). Other technical methods to reduce attrition could include ceramic coating at wear generating surfaces, also applicable for growth guidance techniques[18],[20]).

Conflicts of Interest: AA reports royalties from Paradigm Spine, Joimax, consultancy from Spinal Balance, and editorial board membership from Clinical Spine Surgery, Spine, outside the submitted work; the rest of the authors have nothing to disclose. Author Contributions: Each co-author satisfied the 4 criteria as defined by ICMJE.
  19 in total

1.  Outcomes of Optimal Distraction Forces and Frequencies in Growth Rod Surgery for Different Types of Scoliotic Curves: An In Silico and In vitro Study.

Authors:  Aakash Agarwal; Anand K Agarwal; Arvind Jayaswal; Vijay K Goel
Journal:  Spine Deform       Date:  2017-01

2.  Growing rod fractures: risk factors and opportunities for prevention.

Authors:  Justin S Yang; Paul D Sponseller; George H Thompson; Behrooz A Akbarnia; John B Emans; Muharrem Yazici; David L Skaggs; Suken A Shah; Pooria Salari; Connie Poe-Kochert
Journal:  Spine (Phila Pa 1976)       Date:  2011-09-15       Impact factor: 3.468

3.  Quantitative Characteristics of Consecutive Lengthening Episodes in Early-onset Scoliosis (EOS) Patients With Dual Growth Rods.

Authors:  Aakash Agarwal; Ankur Goswami; Govindaraja Perumal Vijayaraghavan; Abhishek Srivastava; Pankaj Kandwal; Upendra Bidre Nagaraja; Vijay K Goel; Anand K Agarwal; Arvind Jayaswal
Journal:  Spine (Phila Pa 1976)       Date:  2019-03-15       Impact factor: 3.468

4.  Special article: Update on the magnetically controlled growing rod: tips and pitfalls.

Authors:  Jason Pui Yin Cheung; Patrick Cahill; Burt Yaszay; Behrooz A Akbarnia; Kenneth M C Cheung
Journal:  J Orthop Surg (Hong Kong)       Date:  2015-12       Impact factor: 1.118

5.  Do magnetic growing rods have lower complication rates compared with conventional growing rods?

Authors:  Kar H Teoh; Daniel M G Winson; Stuart H James; Alwyn Jones; John Howes; Paul R Davies; Sashin Ahuja
Journal:  Spine J       Date:  2016-02-02       Impact factor: 4.166

6.  Frequent small distractions with a magnetically controlled growing rod for early-onset scoliosis and avoidance of the law of diminishing returns.

Authors:  Jason Pui Yin Cheung; Cora Bow; Dino Samartzis; Kenny Kwan; Kenneth Man Chee Cheung
Journal:  J Orthop Surg (Hong Kong)       Date:  2016-12       Impact factor: 1.118

7.  Analysis of Explanted Magnetically Controlled Growing Rods From Seven UK Spinal Centers.

Authors:  Thomas J Joyce; Simon L Smith; Paul R P Rushton; Andrew J Bowey; Michael J Gibson
Journal:  Spine (Phila Pa 1976)       Date:  2018-01-01       Impact factor: 3.468

8.  Magnetic controlled growing rods for early-onset scoliosis: a 4-year follow-up.

Authors:  Kar Hao Teoh; Daniel M G Winson; Stuart H James; Alwyn Jones; John Howes; Paul R Davies; Sashin Ahuja
Journal:  Spine J       Date:  2016-02-01       Impact factor: 4.166

9.  Active Apex Correction With Guided Growth Technique for Controlling Spinal Deformity in Growing Children: A Modified SHILLA Technique.

Authors:  Aakash Agarwal; Loai Aker; Alaaeldin Azmi Ahmad
Journal:  Global Spine J       Date:  2019-06-23

Review 10.  The MAGEC system for spinal lengthening in children with scoliosis: A NICE Medical Technology Guidance.

Authors:  Michelle Jenks; Joyce Craig; Joanne Higgins; Iain Willits; Teresa Barata; Hannah Wood; Christine Kimpton; Andrew Sims
Journal:  Appl Health Econ Health Policy       Date:  2014-12       Impact factor: 2.561

View more
  3 in total

1.  Active Apex Correction: An overview of the modified SHILLA technique and its clinical efficacy.

Authors:  Alaaeldin Azmi Ahmad; Akash Agarwal
Journal:  J Clin Orthop Trauma       Date:  2020-07-23

2.  Towards a validated patient-specific computational modeling framework to identify failure regions in traditional growing rods in patients with early onset scoliosis.

Authors:  Aakash Agarwal; Manoj Kodigudla; Amey Kelkar; Daksh Jayaswal; Vijay Goel; Vivek Palepu
Journal:  N Am Spine Soc J       Date:  2020-12-13

3.  Premature spinal fusion after insertion of magnetically controlled growing rods for treatment of early-onset scoliosis: illustrative case.

Authors:  Anna H Green; Andrzej Brzezinski; Terrence Ishmael; Stephen Adolfsen; J Andrew Bowe
Journal:  J Neurosurg Case Lessons       Date:  2021-10-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.