Literature DB >> 32399160

Parametric Design and Hybrid Fabrication of Above-Knee Prosthesis.

Lalitrao Amrutsagar1, Gaurav Parit1, Rupesh Ghyar1, Ravi Bhallamudi1.   

Abstract

BACKGROUND: India has a large number of above-knee amputation patients who require a prosthetic leg that is affordable, conformal, functional, and durable. Available low-cost solutions, such as Jaipur foot, employ gypsum plaster in the process of fabricating a fitting socket. This has four shortcomings: (1) requirement of trained technicians, (2) high possibility of manual errors leading to rework, (3) long production time of several hours, and (4) difficulty in scaling up for widespread application.
MATERIALS AND METHODS: An improved approach is presented here, which combines computer-aided parametric design and numerically controlled machining with manual thermoforming to overcome the above issues. The socket is semi-automatically designed based on 60 parameters, derived from 23 measurements obtained on the natural stump of a patient. The three-dimensional (3D) computer-aided design model of the socket can be used for additive manufacturing (3D printing), which was found to be accurate, but time-consuming and expensive. Hence, a hybrid process was evolved with the following three steps: computer numeric control machining of the stump and shank replica in polyurethane (PU) foam, followed by coating with suitable epoxy, and finally high-density polyethylene pipe thermoforming over the PU foam replicas.
RESULTS: Three prostheses were fabricated using both conventional and hybrid processes and provided to volunteer patients. The hybrid process resulted in 28% reduction in overall fabrication time and improved satisfaction of patients due to better fit and comfort.
CONCLUSION: The proposed approach can be adapted for mass customization, required to meet the large gap in demand and supply, especially in resource-constrained settings. © Indian Orthopaedics Association 2020.

Entities:  

Keywords:  Computer numerical control machining; Computer-aided parametric design; Jaipur foot prosthesis

Year:  2020        PMID: 32399160      PMCID: PMC7205932          DOI: 10.1007/s43465-020-00059-w

Source DB:  PubMed          Journal:  Indian J Orthop        ISSN: 0019-5413            Impact factor:   1.251


  7 in total

1.  Functional considerations in the fitting of above-knee prostheses.

Authors:  C W RADCLIFFE
Journal:  Artif Limbs       Date:  1955-01

Review 2.  Lower-limb prosthetic technologies in the developing world: A review of literature from 1994-2010.

Authors:  Jan Andrysek
Journal:  Prosthet Orthot Int       Date:  2010-12       Impact factor: 1.895

3.  Factors affecting quality of life in lower limb amputees.

Authors:  Richa Sinha; Wim J A van den Heuvel; Perianayagam Arokiasamy
Journal:  Prosthet Orthot Int       Date:  2011-03       Impact factor: 1.895

4.  Socket design for the above-knee amputee.

Authors:  J Foort
Journal:  Prosthet Orthot Int       Date:  1979-08       Impact factor: 1.895

5.  Northwestern University Flexible Subischial Vacuum Socket for persons with transfemoral amputation-Part 1: Description of technique.

Authors:  Stefania Fatone; Ryan Caldwell
Journal:  Prosthet Orthot Int       Date:  2017-01-17       Impact factor: 1.895

6.  Northwestern University Flexible Subischial Vacuum Socket for persons with transfemoral amputation: Part 2 Description and Preliminary evaluation.

Authors:  Stefania Fatone; Ryan Caldwell
Journal:  Prosthet Orthot Int       Date:  2017-01-30       Impact factor: 1.895

7.  Intervention for phantom limb pain: A randomized single crossover study of mirror therapy.

Authors:  Shashikumar Ramadugu; Satish C Nagabushnam; Nagendra Katuwal; Kaushik Chatterjee
Journal:  Indian J Psychiatry       Date:  2017 Oct-Dec       Impact factor: 1.759

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.