Filling of Fine and Core Biopsy Needles With the Contrast Agent Sulfur Hexafluoride: Ex Vivo and in vitro Evaluation of Ultrasound Needle Visibility.

Per Thunswärd1, Anders Nilsson1, Håkan Ahlström1.   

Abstract

OBJECTIVES: To investigate whether the ex vivo and in vitro ultrasound visibility of fine needles (FNs) and core biopsy needles (CNBs) can be improved by filling them with an ultrasound contrast agent.
METHODS: After needle filling with the contrast agent sulfur hexafluoride, punctures with FNs and CBNs were recorded in the B-mode and contrast-specific imaging mode (10 observations in each of the 4 groups). Recordings were made in both butchered bovine liver (experiment I) and a water bath (experiment II). Air and normal saline were used as controls (total n = 120 for each experiment). In experiment I, 4 ultrasound specialists subjectively assessed the relative needle visibility in the recordings by using an arbitrary scale (integers 0-10). In experiment II, the contrast-to-noise ratio was calculated for both the entire needle course and the needle tip area.
RESULTS: In experiment I, subjective visibility was increased compared with both controls only for CBNs in the contrast-specific imaging mode (P < .01). In experiment II, the contrast-to-noise ratio for both the entire needle course and the needle tip area increased compared with both controls for both FNs and CBNs in the contrast-specific imaging mode (P < .05).
CONCLUSIONS: Ultrasound contrast agent needle filling is a promising new method to increase the visibility of CBNs in the contrast-specific imaging mode. This finding needs to be confirmed in vivo before its clinical value can be assessed.
© 2020 The Authors. Journal of Ultrasound in Medicine published by Wiley Periodicals, Inc. on behalf of the American Institute of Ultrasound in Medicine.

Entities:  

Keywords:  biopsy; needle visibility; ultrasound contrast agent

Mesh:

Substances:

Year:  2020        PMID: 32395841     DOI: 10.1002/jum.15321

Source DB:  PubMed          Journal:  J Ultrasound Med        ISSN: 0278-4297            Impact factor:   2.153


  1 in total

1.  In vitro CT visualization of the biopsy location by filling the core biopsy needle with contrast media.

Authors:  Per Thunswärd; David Eksell; Håkan Ahlström; Anders Magnusson
Journal:  Acta Radiol       Date:  2021-10-22       Impact factor: 1.701

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.