Acute Thrombus Burden on Coated Flow Diverters Assessed by High Frequency Optical Coherence Tomography.

Robert M King1,2, Erin T Langan1, Giovanni J Ughi1, Christopher M Raskett1, Ajit S Puri1, Hans Henkes3, Matthew J Gounis4.   

Abstract

PURPOSE: The implantation of flow diverters requires administration of dual anti-platelet therapy, posing the potential for complications. The p48MW HPC (phenox, Bochüm, Germany) hydrophilic-coated flow diverting stent is designed to be anti-thrombotic, thus opening the potential for single anti-platelet therapy. We deploy a novel intravascular high-resolution imaging technique, high-frequency optical coherence tomography (HF-OCT), to study in an animal model the acute thrombus formation on coated p48MW devices versus uncoated control devices.
METHODS: Three pigs were implanted with 4 flow diverters each, two test hydrophilic-coated devices, and two control uncoated devices (p48MW). Each pig was treated with a different anti-platelet regime: no anti-platelet therapy, aspirin only, aspirin and clopidogrel. Twenty minutes after the flow diverter was implanted, an HF-OCT data set was acquired. Acute clot formed on the flow diverter at each covered side branch was measured from the HF-OCT slices. Factors considered to be important were the device type (pHPC versus bare metal), aspirin, clopidogrel, and vessel location. A linear model was constructed from the significant factors.
RESULTS: Both coating (p < 0.001) and aspirin (p = 0.003) were significantly related to reduction in clot burden, leading to an approximate 100-fold and 50-fold reduction in clot, respectively.
CONCLUSIONS: This study shows the power of HF-OCT not only in the detection of clot but also the quantification of clot burden. In an animal model, the pHPC-coated p48MW significantly reduced acute thrombus formation over jailed side branches as compared to the bare metal p48MW that was nearly eliminated when combined with aspirin administration.

Entities:  

Keywords:  Flow diverter; Hydrophilic coating; Optical coherence tomography; Stents; Thrombus

Mesh:

Substances:

Year:  2020        PMID: 32385610     DOI: 10.1007/s00270-020-02482-w

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  1 in total

1.  Hydrophilic Surface Coating of Flow Diverters: A Possible Way to Omit Dual Platelet Aggregation Inhibition.

Authors:  Ansgar Berlis
Journal:  Cardiovasc Intervent Radiol       Date:  2020-08       Impact factor: 2.740

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.