Image quality in dual-source multiphasic dynamic computed tomography of the abdomen: evaluating the effects of a low tube voltage (70 kVp) in combination with contrast dose reduction.

Keisuke Miyoshi1, Hideko Onoda1, Mashiro Tanabe1, Sei Nakao1, Mayumi Higashi1, Etsushi Iida1, Munemasa Okada1, Matakazu Furukawa1, Katsuyoshi Ito2.   

Abstract

PURPOSE: To compare the image quality of multiphasic (arterial, portal, and equilibrium phases) dynamic computed tomography (CT) of the abdomen obtained by a low tube voltage (70kVp) in combination with a half-dose iodine load using low-concentration contrast agent in high tube output dual-source CT with a standard tube voltage (120kVp) and full-dose iodine load using the same group of adult patients.
METHODS: Fifty-five patients who underwent both low-tube-voltage (70kVp) abdominal CT with a half-dose iodine load and standard-tube-voltage (120kVp) CT with a full-dose iodine load were analyzed. The mean CT values and signal-to-noise ratio (SNR) of the liver, aorta and portal veins were quantitatively assessed. In addition, the contrast enhancement of the abdominal organs and overall image quality were qualitatively evaluated.
RESULTS: The mean CT values and SNR of the liver parenchyma were significantly higher in 70-kVp protocol than in 120-kVp protocol in all 3 phases (p = 0.018 ~  < 0.001). Regarding the qualitative analysis, the overall image quality in the 70-kVp protocol was significantly better than in the 120-kVp protocol in all 3 phases (p < 0.001). In addition, the contrast enhancement scores of the liver parenchyma and hepatic vein in the equilibrium phase were also significantly higher in the 70-kVp protocol than in the 120-kVp protocol (p < 0.001).
CONCLUSION: A low tube voltage (70kVp) in combination with a half-dose iodine load using a low-concentration contrast agent and an iterative reconstruction algorithm in high tube output dual-source CT may improve the contrast enhancement and image quality in multiphasic dynamic CT of the abdomen in patients under 71 kg of body weight.

Entities:  

Keywords:  Diagnostic imaging; Iodine; Low tube voltage; X-ray computed tomography

Mesh:

Substances:

Year:  2020        PMID: 32382819     DOI: 10.1007/s00261-020-02565-9

Source DB:  PubMed          Journal:  Abdom Radiol (NY)


  2 in total

1.  Contrast media augmented apoptosis of cultured renal mesangial, tubular, epithelial, endothelial, and hepatic cells.

Authors:  Amir Peer; Zhan Averbukh; Sylvia Berman; David Modai; Michael Averbukh; Joshua Weissgarten
Journal:  Invest Radiol       Date:  2003-03       Impact factor: 6.016

Review 2.  Computed tomography in the evaluation for transcatheter aortic valve implantation (TAVI).

Authors:  Paul Schoenhagen; Jörg Hausleiter; Stephan Achenbach; Milind Y Desai; E Murat Tuzcu
Journal:  Cardiovasc Diagn Ther       Date:  2011-12
  2 in total
  2 in total

1.  Finding the optimal tube current and iterative reconstruction strength in liver imaging; two needles in one haystack.

Authors:  Bibi Martens; Joris G A Bosschee; Sander M J Van Kuijk; Cécile R L P N Jeukens; Maikel T H Brauer; Joachim E Wildberger; Casper Mihl
Journal:  PLoS One       Date:  2022-04-07       Impact factor: 3.240

2.  Low kV Computed Tomography of Parenchymal Abdominal Organs-A Systematic Animal Study of Different Contrast Media Injection Protocols.

Authors:  Daniel Overhoff; Gregor Jost; Michael McDermott; Olaf Weber; Hubertus Pietsch; Stefan O Schoenberg; Ulrike Attenberger
Journal:  Tomography       Date:  2021-11-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.