| Literature DB >> 32373753 |
Zackary Herbst1, Coleman T Turgeon2, Chad Biski3, Hamid Khaledi1,4, Nancy B Shoemaker5, Patrick D DeArmond5, Sara Smith6, Joseph Orsini3, Dietrich Matern2, Michael H Gelb1.
Abstract
Measurement of the absolute concentration of the biomarker psychosine in dried blood spots (DBS) is useful for diagnosis and prognosis of Krabbe disease and to support newborn screening of this leukodystrophy. As for assays for more common diseases, it is important to achieve congruence when multiple clinical laboratories provide testing. Four clinical laboratories, one research laboratory, and a commercial vendor collaborated with the goal to achieve congruence in quantitative psychosine measurement in DBS. A set of DBS calibrators was prepared by a single vendor and used in each reference laboratory to make a standard curve of measured psychosine in DBS versus the stated calibrator psychosine level. Congruence between the participating five laboratories was achieved using the psychosine DBS calibrators. This allowed application of disease-specific reference ranges obtained by the reference laboratory with the most extensive data by the other reference laboratories. Congruence between multiple reference laboratories in the measurement of the absolute concentration of biomarkers in DBS (and by extension other samples) is possible by the use of a common set of DBS calibrators.Entities:
Keywords: Krabbe disease; biomarkers; mass spectrometry; newborn screening
Year: 2020 PMID: 32373753 PMCID: PMC7199469 DOI: 10.3390/ijns6020029
Source DB: PubMed Journal: Int J Neonatal Screen ISSN: 2409-515X
Figure 1Galactocerebrosidase (GALC)-catalyzed reaction and direct conversion of galactosylceramide to psychosine when GALC is deficient as proposed [4].
Figure 2The y-axis gives the psychosine value (nM) measured in each laboratory by their LC-MS/MS method. The x-axis gives the vendor stated psychosine values (nM) for each dried blood spot (DBS) calibrator: Filled circles/solid line (Nationwide), open circles/solid line (Mayo), filled triangles/dotted line (PerkinElmer), open squares/solid line (NY Wadsworth Center), plus signs/solid line (GelbChem LLC).