Eric Laugeman1, Ana Heermann1, Jessica Hilliard1, Michael Watts1, Marshia Roberson1, Robert Morris1, Sreekrishna Goddu1, Abhishek Sethi1, Imran Zoberi1, Hyun Kim1, Sasa Mutic1, Geoffrey Hugo1, Bin Cai1.
Abstract
PURPOSE: To perform a comprehensive validation of plans generated on a preconfigured Halcyon 2.0 with preloaded beam model, including evaluations of new features and implementing the patient specific quality assurance (PSQA) process with multiple detectors.Entities:
Keywords: double-stack MLC; patient-specific QA; ring gantry LINAC
Mesh:
Year: 2020 PMID: 32368862 PMCID: PMC7386180 DOI: 10.1002/acm2.12881
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
The number of plans per site and per treatment technique.
| Site | # IMRT | # VMAT | 3D with DFB |
|---|---|---|---|
| Whole Pelvis | 1 | 3 | 1 |
| Prostate | 2 | 2 | – |
| Brain | 2 | 2 | – |
| Head and Neck | 2 | 2 | – |
| Lung | 2 | 4 | – |
| Abdomen | 1 | – | – |
| Breast (single isocentre) | – | 5 | – |
| Breast (extended treatment field) | – | 6 | – |
| Breast (3D) | – | – | 10 |
| Spine | – | – | 1 |
| TG 119 | 5 | 5 | – |
DFB, dynamically flattened beam.
TG‐119 results for point dose (% difference calculated and measured) and radiographic film measurements (gamma analysis with 3%/2 mm gamma criteria).
| Test | location | Point dose (%) | Film (3%/2 mm) | ||
|---|---|---|---|---|---|
| IMRT | VMAT | IMRT | VMAT | ||
| MultiTarget | Isocentre | −0.6 | −0.2 | 99.6 | 99.4 |
| 4cm superior | −1.4 | −0.8 | |||
| 4cm inferior | −0.5 | −2.0 | |||
| Prostate | Target | −0.4 | 1.6 | 99.6 | 99.7 |
| OAR | 1.1 | 3.4 | 98.8 | 98.9 | |
| Head/Neck | Target | −2.5 | 2.3 | 96.9 | 99.0 |
| OAR | −0.8 | −1.6 | 98.5 | 99.6 | |
| CShape (easy) | Target | −1.8 | −0.6 | 99.8 | 99.9 |
| OAR | 1.0 | 1.7 | 99.9 | 98.6 | |
| CShape (hard) | Target | 5.6 | 2.9 | 94.7 | 100.0 |
| OAR | 1.6 | −0.1 | 98.5 | 99.4 | |
| Confidence Limit | Target | 6.4 | 4.1 | 6.3 | 1.2 |
| OAR | 2.5 | 4.2 | 2.4 | 2.0 | |
| Combined | 4.4 | 4.2 | 4.9 | 1.6 | |
Ion chamber (IC) % difference and 3%/2 mm gamma passing rates (mean ± standard deviation) for all measurement devices for IMRT and VMAT plans.
| n | IC (%) | Film | n | ArcCHECK | Delta4 | PD | |
|---|---|---|---|---|---|---|---|
| IMRT | 10 | −0.6 ± 1.0 | 98.1 ± 1.9 | 15 | 99.1 ± 0.7 | 99.0 ± 1.1 | 99.8 ± 0.7 |
| VMAT | 24 | −0.9 ± 1.6 | 96.3 ± 3.5 | 29 | 99.0 ± 0.9 | 99.3 ± 1.2 | 99.9 ± 0.4 |
| ALL | 34 | −0.8 ± 1.4 | 96.8 ± 3.3 | 44 | 99.1 ± 0.9 | 99.2 ± 1.2 | 99.9 ± 0.5 |
| Tolerance Limit | 90.3 | 96.2 | 96.8 | 99.5 | |||
| Action Limit | 86.3 | 96.1 | 95.7 | 98.4 |
IC and film results do not include TG‐119 plans, while TG‐119 plans were measured with ArcCHECK, Delta4, and PD.
Fig. 1Example film result for an extended treatment field breast VMAT plan exhibiting over‐response to low‐energy scatter. Five of six extended field VMAT breast plans were below 95% passing rate on film with majority of failing points in the 20–40% dose range.
Fig. 2Control chart for EDR2 Film PSQA. The final six plans were with extended treatment field and where film over‐response is assumed. After removing these six from the action limit calculation, the limit increases to above TG‐218 recommendations.
Fig. 3Extended treatment field delivered on ArcCHECK using the elongated treatment field measurement technique to capture the full dose distribution with automated couch shifts included in the delivery.
Mean ± SD point dose and true composite measured values for plans with DFB fields.
| DFB plans | N | IC (%) | ArcCHECK | Delta4 | PD |
|---|---|---|---|---|---|
| Breast | 10 | −1.5 ± 0.5 | 97.3 ± 1.7 | 99.4 ± 0.5 | 100 ± 0.0 |
| AP/PA | 1 | −1.2 | 100 | 96.5 | 100 |
| 4 field box | 1 | −0.8 | 98.6 | 99.0 | 100 |
Fig. 4Tangent breast plan with dynamically flattened beam on ArcCHECK.
Gamma passing rates for two plans (one IMRT, one VMAT) measured with ArcCHECK (AC) and Portal Dosimetry (PD) with at least one month in between measurements.
| IMRT Plan AC | IMRT Plan PD | VMAT Plan AC | VMAT Plan PD | |
|---|---|---|---|---|
| Measurement 1 | 99.7 | 100 | 98.9 | 100 |
| Measurement 2 | 99.5 | 100 | 99.7 | 100 |
| Measurement 3 | 98.9 | 99.8 |