Purpose: To pursue high precision dose in lesions and steeper dose fall-off in healthy tissues of brain metastases stereotactic radiotherapy (SRT), this study investigated an opitimized planning by comparison different prescription dose line in the treatment of brain metastases using Cyberknife (CK) Robotic Radiosurgery System.Entities:
Keywords: CyberKnife; brain metastases; prescription dose line; stereotactic radiotherapy; treatment plan
Year: 2020 PMID: 32368299 PMCID: PMC7196273 DOI: 10.7150/jca.41102
Source DB: PubMed Journal: J Cancer ISSN: 1837-9664 Impact factor: 4.207
Clinical characteristics of 77 patients.
| 53(27-81) | |
| Age≥60 | 28 |
| Age<60 | 49 |
| Male | 32 |
| Female | 45 |
| Liver | 15 |
| Breast | 20 |
| Lung | 45 |
| 72(51-100) | |
| KPS≥70 | 54 |
| KPS <70 | 23 |
| Multiple metastases | 15 |
| Single metastasis | 62 |
| Cerebral hemisphere | 47 |
| Cerebellum | 35 |
| Brainstem | 10 |
| 4.6(0.05-9.89) | |
| 5.0-10.0 | 17 |
| 1.0-5.0 | 64 |
| <1.0 | 11 |
| 12-32(24) | |
| 1-4(3) |
Figure 1Different CK SRT plans for brain metastases patients. The representative patient had axial images taken, (A) and (D) were Patient 1 with metastases in the right lateral ventricle and Patient 2 with brainstem metastases. The red and purple lines area indicate the GTV and the PTV, respectively. Blue, orange and yellow lines represent Brainstem, Optic Nerves, and Eyes, respectively. (B) and (E) represent 50% prescription isodose covered more than 95% of the Patient 1 PTV and Patient 2 GTV (Plan_50%). (C) and (F) represent 80% prescription isodose covered more than 95% of the Patient 1 PTV or Patient 2 GTV (Plan_80%).Green line zones were covered 6-mm thick healthy brain tissue adjacent to the PTV or the GTV.
Figure 2PTV+2, PTV+6 and brainstem dose volume histogram (DVH) of patient 2 with brainstem metastases. (A) and (B) were dosimetry distribution of 2-mm-thick and 6-mm-thick healthy brain tissue adjacent to the PTV. (C) was the dosimetry distribution of Brainstem.
The dose distribution of healthy brain tissue around the PTV (using percent of the global maximum dose).
| PTV | PTV+2 | PTV+6 | ||||||
|---|---|---|---|---|---|---|---|---|
| Plan_50% | 63.62%±3.80% | 53.51%±1.46% | 57.28%±3.20% | 47.57%±2.56% | 53.40%±3.11% | 35.86%±2.03% | ||
| Plan_55% | 65.13%±2.71% | 54.61%±1.96% | 58.02%±4.01% | 48.50%±2.17% | 56.12%±3.49% | 36.24%±1.96% | ||
| Plan_60% | 67.34%±2.92% | 55.08%±1.73% | 60.25%±3.14% | 50.09%±2.87% | 58.84%±2.99% | 39.8%±3.23% | ||
| Plan_65% | 68.26%±3.16% | 56.12%±2.04% | 65.45%±3.28% | 51.27%±3.58% | 63.35%±3.50% | 39.96%±3.67% | ||
| Plan_70% | 70.82%±2.75% | 62.39%±1.80% | 69.63%±3.77% | 54.47%±4.01% | 67.37%±3.37% | 45.32%±2.15% | ||
| Plan_75% | 71.94%±3.03% | 63.61%±1.54% | 70.82%±3.65% | 57.12%±3.44% | 69.34%±2.86% | 53.42%±4.21% | ||
| Plan_80% | 76.56%±3.17% | 69.10%±1.74% | 72.49%±3.09% | 62.06%±3.97% | 70.88%±3.24% | 57.35%±2.65% | ||
*PTV+2 meant 2-mm-thick healthy brain tissue adjacent to the PTV, PTV+6 meant 4-mm-thick healthy brain tissue adjacent to the PTV+2
The dosimetry distribution of OAR (using percent of the global maximum dose).
| Brainstem | Optic Nerves | Optic Chiasm | ||||||
|---|---|---|---|---|---|---|---|---|
| Plan_50% | 63.62%±3.17% | 20.75%±1.55% | 14.84%±4.17% | 7.51%±3.76% | 18.84%±4.07% | 5.39%±1.17% | ||
| Plan_55% | 65.13%±2.44% | 21.20%±1.34% | 15.05%±2.98% | 7.59%±2.91% | 19.25%±4.30% | 5.44%±1.49% | ||
| Plan_60% | 67.34%±3.09% | 22.13%±1.46% | 16.12%±3.70% | 7.62%±2.95% | 20.43%±4.12% | 5.80%±1.96% | ||
| Plan_65% | 68.26%±2.91% | 22.40%±1.19% | 16.45%±3.24% | 8.03%±3.12% | 22.17%±3.37% | 6.11%±2.31% | ||
| Plan_70% | 70.82%±2.84% | 25.29%±2.07% | 19.05%±2.78% | 8.23%±3.51% | 23.75%±3.37% | 7.28%±2.15% | ||
| Plan_75% | 71.94%±3.15% | 29.02%±2.13% | 20.46%±3.14% | 9.25%±3.18% | 24.73%±2.16% | 7.42%±1.98% | ||
| Plan_80% | 76.56%±2.97% | 31.25%±1.83% | 22.49%±3.09% | 10.06%±3.62% | 25.88%±3.24% | 7.51%±2.14% | ||
*Brainstem dose statistics were from 10 patients(10 lesions) with brainstem metastasis. Optic nerves were superposition of both sides.
The statistical result of index for different plans
| PTV | |||||
|---|---|---|---|---|---|
| Plan_50% | 96.02%±0.81% | 1.20±0.10 | 1.41±0.37 | 108±11 | 7276±256 |
| Plan_55% | 95.83%±0.52% | 1.20±0.09 | 1.40±0.39 | 102±8 | 7154±217 |
| Plan_60% | 96.00%±0.74% | 1.15±0.07 | 1.36±0.21 | 105±14 | 6990±187 |
| Plan_65% | 95.98%±0.36% | 1.16±0.09 | 1.36±0.30 | 110±12 | 6951±135 |
| Plan_70% | 96.12%±0.29% | 1.14±0.08 | 1.34±0.28 | 106±9 | 6869±211 |
| Plan_75% | 95.69%±0.38% | 1.15±0.09 | 1.35±0.33 | 113±15 | 6834±158 |
| Plan_80% | 95.84%±0.37% | 1.16±0.07 | 1.33±0.32 | 109±12 | 6738±167 |
*The value of machine hop were in single fraction radiosurgery.
Figure 3The dosimetry distribution of the whole brain tissue in different plans for the same patients. (A) and (B) were the volume of healthy brain tissue received 12 Gy/14 Gy (V12Gy/V14Gy) in a single fraction.