FEASIBILITY OF DISPLAYED EXPOSURE INDEX IN IEC STANDARD FRAMEWORK AS A DOSE OPTIMISATION TOOL FOR DIGITAL RADIOGRAPHY SYSTEMS.

Hyemin Park1, Yongsu Yoon2, Nobukazu Tanaka2, Jungsu Kim3, Jungmin Kim1, Junji Morishita2.   

Abstract

International Electrotechnical Commission (IEC) established the framework for the use of exposure index (EI) for evaluating the exposure conditions with various digital systems. In this study, we investigated the feasibility of EI, as per the IEC, by comparing the EIs obtained through manual calculated and that displayed on the console of two computed radiography (CR) and digital radiography (DR) systems with radiation beam qualities of RQA3,5,7 and 9. As a result, both two systems indicated an uncertainty of less than 20% for both calculated and displayed EI with all beam qualities except displayed EI obtained by RQA3. However, the displayed EI values were different even under the same exposure conditions because of the characteristics of the imaging receptor materials, such as BaFI or CsI, of two systems. Therefore, when an operator attempts to introduce displayed EI for managing radiation dose, it is essential to understand the characteristics of the digital system.
© The Author(s) 2020. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Mesh:

Year:  2020        PMID: 32364600     DOI: 10.1093/rpd/ncaa052

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Proper Management of the Clinical Exposure Index Based on Body Thickness Using Dose Optimization Tools in Digital Chest Radiography: A Phantom Study.

Authors:  Yongsu Yoon; Hyemin Park; Jungmin Kim; Jungsu Kim; Younghoon Roh; Nobukazu Tanaka; Junji Morishita
Journal:  Int J Environ Res Public Health       Date:  2021-05-13       Impact factor: 3.390

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.