| Literature DB >> 32364151 |
Yisong Zhang1,1, Song Zhang1,1, Lin Yang1, Yimin Yang1, Xuwen Li1, Dongmei Hao1, Mingzhou Xu2, Jing Shao3.
Abstract
BACKGROUND: Fetal electrocardiogram (FECG) can be obtained in a non-invasive manner to monitor fetal growth status.Entities:
Keywords: Fetal ECG; R-R interval; fetal QRS position correction; fetal heart rate
Mesh:
Year: 2020 PMID: 32364151 PMCID: PMC7369097 DOI: 10.3233/THC-209019
Source DB: PubMed Journal: Technol Health Care ISSN: 0928-7329 Impact factor: 1.285
Figure 1.The overall structure of the system. Left: MF-HOLTER. Right: FECG-MS.
Figure 2.The flowchart of the MF-HOLTER.
Figure 3.The flowchart of the FECG-MS.
Figure 4.The complete FECG separated (120 bpm).
Figure 5.The separated FECG with FQRS lost (120 bpm).
Figure 6.The separated FECG where the MQRS is misjudged as FQRS (120 bpm).
Figure 7.The step diagram.
The parameters of FLUKE’s PS320 FETAL SIMULATOR
| ECG type | Optional amplitude ( | Optional heart rate (bpm) |
|---|---|---|
| MECG | 500, 1000, 2000 | 60, 80, 100, 120, 140, 160 |
| FECG | 50, 100, 200, 500, 1000, 2000 | 30, 60, 90, 120, 150, 180, 210, 240 |
Figure 8.The three AECG signals. From top to bottom are case 1, case 2 and case 3, respectively.
The accuracy of the FHR calculation
| FHR value (bpm) | Duration (min) | Results |
|---|---|---|
| 120 | 15 | Total: 3616 Wrong: 8 Accuracy: 99.78% |
| 150 | 15 | Total: 3616 Wrong: 40 Accuracy: 98.89% |
| 180 | 15 | Total: 3616 Wrong: 40 Accuracy: 98.89% |
| 120/150/180 | 15 | Total: 3616 Wrong: 72 Accuracy: 98.01% |
The average FHR values from the proposed system and Monica AN24
| Data | Case 1 | Case 2 | Case 3 |
|---|---|---|---|
| The proposed system | 144.13 | 133.06 | 127.53 |
| Monica AN24 | 144.18 | 132.95 | 128.27 |
| Difference | 0.05 | 0.11 | 0.74 |